iSLEEVE Introducer Set

K190184 · Boston Scientific · DYB · Mar 7, 2019 · Cardiovascular

Device Facts

Record IDK190184
Device NameiSLEEVE Introducer Set
ApplicantBoston Scientific
Product CodeDYB · Cardiovascular
Decision DateMar 7, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system.

Device Story

The iSLEEVE Introducer Set is a sterile, single-use catheter system used to provide percutaneous femoral artery access. It consists of an introducer sheath with a three-way stopcock, a dilator, a product mandrel, and a flushing tube. The sheath features a passively expandable distal end that is radially compliant, allowing it to expand and collapse during the delivery and removal of transcatheter heart valve systems; this design minimizes the duration of vessel expansion. A hydrophilic coating is applied to the sheath to enhance lubricity upon activation. The device is operated by clinicians in a clinical setting to facilitate the delivery of compatible valve systems. It does not involve electronic processing or software.

Clinical Evidence

No clinical data. Substantial equivalence was demonstrated through non-clinical bench testing only.

Technological Characteristics

Sterile, single-use catheter introducer. Components: introducer sheath with three-way stopcock, dilator, mandrel, flushing tube. Features: passively expandable, radially compliant distal tip; hydrophilic coating for lubricity. Mechanical device; no energy source or software.

Indications for Use

Indicated for patients requiring femoral vascular access for the introduction and removal of the ACURATE TF Valve System and ancillary devices. Suitable for use in vessels with a diameter of ≥ 5.5 mm.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 7, 2019 Boston Scientific Kripa Pandya Regulatory Affairs Specialist I Three Scimed Place Maple Grove, Minnesota 55311 Re: K190184 Trade/Device Name: iSLEEVE™ Introducer Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: January 31, 2019 Received: February 1, 2019 ### Dear Kripa Pandya: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/7 description: The image shows the name "Jaime Raben" in a large, sans-serif font. The text is black and stands out against a background that appears to be a light blue or white. The name is presented in a simple, clean format, with the first name and last name clearly separated. for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K190184 Device Name iSLEEVE™ Introducer Set Indications for Use (Describe) 14F iSLEEVE Introducer Set: The 14F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of the ACURATE TF Valve System and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.5 mm diameter. Type of Use (Select one or both, as applicable) | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |----------------------------------------------------------------------------------| | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary per 21 CFR §807.92 | Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Contact Name and<br>Information | Kripa Pandya<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2870<br>Fax: 763-494-2222<br>e-mail: Kripa.Pandya@bsci.com | | Prepared by | Kripa Pandya<br>31 January 2019 | | Proprietary Name | iSLEEVE™ Introducer Set | | Common Name | Catheter Introducer | | Product Code | DYB | | Classification | Class II, 21 CFR Part 870.1340 | | Predicate Device | iSLEEVE™ Introducer Set K180785 22 June 2018 | #### Device Description The iSLEEVE Introducer Set is a sterile, single-use introducer catheter that provides percutaneous access to the femoral artery for introduction and removal of compatible transcatheter heart valve systems and ancillary devices into the vascular system. The iSLEEVE Introducer Set is composed of a dilator and an introducer sheath with a threeway stopcock. The distal end of the introducer sheath is passively expandable which allows for transient sheath expansion when the compatible valve system is introduced through it. Once expanded, this region of the introducer sheath is radially compliant, which allows it to expand and collapse as needed during device delivery and removal. This reduces the time the access vessel is expanded. A hydrophilic coating is applied to the entire iSLEEVE introducer sheath which increases the lubricity to aid in delivery when activated. A product mandrel and flushing tube are also supplied with the iSLEEVE Introducer Set. {4}------------------------------------------------ #### Indications for Use / Intended Use #### 14F iSLEEVE Introducer Set: INTENDED USE: The iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system. INDICATIONS FOR USE: The 14F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular svstem for introduction and removal of the ACURATE TF Valve Svstem and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.5 mm diameter. #### Comparison of Technological Characteristics The iSLEEVE Introducer Set incorporates substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate, iSLEEVE Introducer Set, K180785. The proposed 14F iSLEEVE Introducer Set has a modified distal tip design to enhance robustness and ease of vessel insertion compared to the predicate 14F iSLEEVE Introducer Set. #### Non-clinical Performance Data Determination of substantial equivalence is based on an assessment of non-clinical performance bench testing data. #### Bench Testing: Bench testing was performed to evaluate physical integrity, functionality, and performance of the iSLEEVE Introducer Sheath Set. Performance criteria includes dimensional requirements, visibility under fluoroscopy, and interface with compatible devices. #### Clinical Testing Performance testing from clinical studies is not required to demonstrate substantial equivalence of the iSLEEVE Introducer Set. #### Conclusion Based on the indications for use, technological characteristics, and performance testing the iSLEEVE Introducer Set has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the iSLEEVE Introducer Set, K180785.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...