K012854 · Terumo Medical Corp. · DYB · Nov 21, 2001 · Cardiovascular
Device Facts
Record ID
K012854
Device Name
RENAL GUIDING SHEATH
Applicant
Terumo Medical Corp.
Product Code
DYB · Cardiovascular
Decision Date
Nov 21, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Intended Use
The Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the renal arteries.
Device Story
Renal Guiding Sheath functions as both guiding catheter and introducer sheath; facilitates introduction of interventional/diagnostic devices into human vasculature (e.g., renal arteries). Device features coil-reinforced construction, radiopaque marker, and distal hydrophilic coating (5 cm). Packaged with dilator, hemostatic valve, and dilator retaining clip. Operated manually by clinicians in clinical settings. Performance verified via bench testing including penetration, kink resistance, leakage, tensile strength, and tortuous path durability. Biocompatibility confirmed per ISO 10993 for limited contact (≤24hrs) with circulating blood.
Clinical Evidence
Bench testing only. Tests included penetration, kink resistance, leakage, tensile strength, and tortuous path/durability. Biocompatibility testing performed per ISO 10993 for externally communicating, blood-contacting devices (≤24hrs).
Technological Characteristics
Coil-reinforced sheath; distal 5 cm hydrophilic coating; radiopaque marker. Sizes: 6-7Fr. Lengths: 45-55cm. Configurations: straight, hockey stick, multipurpose. Materials biocompatible per ISO 10993. Sterilization: ANSI/AAMI/ISO 11135-1994 (SAL 10^-6). Manual operation.
Indications for Use
Indicated for patients requiring introduction of interventional and diagnostic devices into the human vasculature, including renal arteries.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Super Arrow-Flex Percutaneous Sheath Introducer Set (K924607)
Related Devices
K140641 — VESSIX GUIDE SHEATH 7F, 45 CM, RDC, VESSIX GUIDE SHEATH 7F, 45CM, LIMA · Boston Scientific Corporation · Jul 3, 2014
K081045 — DESTINATION RENAL GUIDING SHEATH · Terumo Medical Corp. · May 28, 2008
K012812 — CAROTID GUIDING SHEATH · Terumo Medical Corp. · Nov 14, 2001
K193125 — R2P Destination Slender Guiding Sheath · Terumo Medical Corporation · Jan 6, 2020
{0}------------------------------------------------
## NOV 21 2001
## SECTION II. SUMMARY OF SAFETY & EFFECTIVENESS
#### DEVICE NAME A.
Proprietary Name: Has not been established yet Classification Name: Catheter Introducer Common Name: Guiding Sheath
#### B. PREDICATE DEVICE
The predicate device is the Super Arrow-Flex Percutaneous Sheath Introducer Set, which is manufactured by Arrow International, Inc. The Super Arrow-flex is cleared through the premarket notification process (K924607).
#### C. INTENDED USE
The Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the renal arteries.
#### D. DESCRIPTION
The Renal Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The sheath is coil reinforced, has a radiopaque marker and is hydrophilically coated on the distal 5 cm. It comes packaged with a dilator, a rotating hemostatic valve (or a detachable hemostatic valve), and a dilator retaining clip.
{1}------------------------------------------------
#### PRINCIPLE OF OPERATION / TECHNOLOGY E.
The Renal Guiding Sheath and the Super Arrow-Flex Percutaneous Sheath Introducer Set are operated manually or by a manual process.
#### DESIGN / MATERIALS F.
The Renal Guiding Sheath use similar materials as the predicate device. Differences in materials between the two devices do not raise any new issues of safety and effectiveness.
#### G. SPECIFICATIONS
| Sheath Size: | 6-7Fr. |
|------------------------------|--------------------------------------|
| Nominal ID / OD: | 6Fr.: .087" / .109" |
| | 7Fr.: .100" / .122" |
| Sheath Length: | 45-55cm |
| Hydrophilic Coating: | Distal 5 cm |
| Distal Shape Configurations: | Straight, Hockey Stick, Multipurpose |
#### H. PERFORMANCE
The performance of the Renal Guiding Sheath is substantially equivalent to the performance of the Super Arrow-Flex Percutaneous Sheath Introducer. The following tests were performed to demonstrate the substantial equivalence of the devices.
- > Penetration
- Kink Resistance A
- > Leakage
- > Tensile Strength
- Tortuous Path / Durability A
{2}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
#### ADDITIONAL SAFETY INFORMATION I.
Sterilization conditions have been validated in accordance with ANSI / AAMI / ISO 11135-1994 to provide a Sterility Assurance Level of 10-6.
Blood contacting materials were tested in accordance with the test recommendations in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part I: Evaluation and Testing." The Renal Guiding Sheath is categorized as "Externally Communicating Device, Circulating Blood, Limited Contact (≤24hrs)". The blood contacting materials were found to be biocompatible.
Expiration dating for the Renal Guiding Sheath will be 30 months.
#### SUBSTANTIAL EQUIVALENCE J.
The Renal Guiding Sheath submitted in this 510(k) is substantially equivalent in intended use, design, principle of operation / technology, materials and performance to the Super Arrow-Flex Percutaneous Sheath Introducer Set (K924607), which is manufactured by Arrow International, Inc. Differences between the devices do not raise any issues of safety or effectiveness.
Terumo's statement of substantial equivalence is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
{3}------------------------------------------------
#### SUBMITTER INFORMATION K.
Name and Address
Terumo Medical Corporation 125 Blue Ball Rd. Elkton, MD 21921
## Contact Person
Mrs. Yuk-Ting Lewis Senior Regulatory Specialist Ph: 410-392-7213 Fax: 410-398-6079 Email: yukting.lewis@terumomedical.com
## Date Prepared
August 1, 2001
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three parallel lines curving upwards, resembling a human figure in profile.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 21 2001
Ms. Yuk-Ting Lewis Senior Regulatory Specialist TERUMO Medical Corporation 125 Blue Ball Road Elkton, MD 21921
Re: K012854
> Renal Guiding Sheath Regulation Number: 870.1340 Regulation Name: Catheter introducer. Regulatory Class: II (two) Product Code: 74 DYB Dated: August 23, 2001 Received: August 24, 2001
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Yuk-Ting Lewis
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Pcoelho
James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Number (if known):
Device Name: Renal Guiding Sheath
## Indications For Use:
The Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the renal arteries.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-------------------------------------------------------------------|----------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | <span style="font-size: 2em;">✓</span> |
| OR | Over-The-Counter Use ______ |
| Division of Cardiovascular & Respiratory Devices<br>510(k) Number | K017854 |
| | (Optional Format 1-2-96) |

Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.