K081045 · Terumo Medical Corp. · DYB · May 28, 2008 · Cardiovascular
Device Facts
Record ID
K081045
Device Name
DESTINATION RENAL GUIDING SHEATH
Applicant
Terumo Medical Corp.
Product Code
DYB · Cardiovascular
Decision Date
May 28, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries.
Device Story
The Destination® Renal Guiding Sheath functions as both a guiding catheter and an introducer sheath. It is a manual device used by clinicians to facilitate the introduction of diagnostic or interventional instruments into the human vasculature, specifically targeting the renal arteries. The system includes a sheath, dilator, hemostatic valve, and dilator retaining clip. It is operated manually by a physician in a clinical setting. The device provides a pathway for other medical instruments, aiding in minimally invasive vascular procedures. Its design allows for navigation through the vasculature to reach target sites, potentially improving the efficiency and safety of renal artery interventions.
Clinical Evidence
No clinical data. Equivalence demonstrated through bench testing and vessel model testing.
Technological Characteristics
Catheter introducer; 5Fr size; 45-55 cm length; hydrophilic coating on distal 5 cm; distal shape configurations include straight, HS, RDC, LIMA, MP. Materials are biocompatible per ISO 10993. Sterilization via ethylene oxide per EN ISO 11135-1.
Indications for Use
Indicated for the introduction of interventional and diagnostic devices into the human vasculature, including the renal arteries, in patients requiring such procedures.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
K08/045 100% Lot 4
Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary
MAY 2 8 2008
# SECTION II. 510(k) SUMMARY
#### A. DEVICE NAME
Proprietary Name: Destination® Renal Guiding Sheath Classification Name: Catheter Introducer Common Name: Guiding Sheath
#### B. PREDICATE DEVICE
The predicate device is the Renal Guiding Sheath, which is manufactured by Terumo Medical Corporation. The Renal Guiding Sheath is cleared through the premarket notification process (K012854).
#### C. INTENDED USE
The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries.
#### D. DESCRIPTION
The Destination® Renal Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The Guiding Sheath is packaged with the following components: a Sheath, a Dilator, a Hemostatic Valve, and a Dilator Retaining Clip.
{1}------------------------------------------------
### E. PRINCIPLE OF OPERATION / TECHNOLOGY
The Destination® Renal Guiding Sheath is operated manually or by a manual process.
### F. DESIGN / MATERIALS
The Destination® Renal Guiding Sheath uses similar materials as the predicate device. Differences in materials between the two devices do not raise any new issues of safety and effectiveness.
#### G. SPECIFICATIONS
| Sheath Size: | 5Fr. |
|------------------------------|-----------------------------|
| Sheath Length: | 45-55 cm |
| Hydrophilic Coating: | Distal 5 cm |
| Distal Shape Configurations: | Straight, HS, RDC, LIMA, MP |
#### H. PERFORMANCE
The performance of the Destination® Renal Guiding Sheath is substantially equivalent to the performance of the unmodified Renal Guiding Sheath. The equivalence was shown through bench and vessel model testing.
{2}------------------------------------------------
Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary
#### I. ADDITIONAL SAFETY INFORMATION
Sterilization conditions have been validated in accordance with EN ISO 11135-1 "Sterilization of health care products – Ethylene Oxide – Part 1 : requirements for development, validation and routine control of sterilization process for medical devices." to provide a Sterility Assurance Level of 10-6.
Blood contacting materials were tested in accordance with the test recommendations in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices – Part I: Evaluation and Testing." The Destination® Renal Guiding Sheath is categorized as "Externally Communicating Device, Circulating Blood, Limited Contact (≤24hrs)". The blood contacting materials were found to be biocompatible.
Expiration dating for the Destination® Renal Guiding Sheath will be 30 months.
### J. SUBSTANTIAL EQUIVALENCE
The Destination® Renal Guiding Sheath submitted in this 510(k) is substantially equivalent in intended use, design, principle of operation / technology, materials and performance to the Renal Guiding Sheath (K012854), which is manufactured by Terumo Medical Corporation. Differences between the devices do not raise any issues of safety or effectiveness.
{3}------------------------------------------------
KOE 1045 page let y
Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary
### K. SUBMITTER INFORMATION
## Name and Address
Terumo Medical Corporation 950 Elkton Blvd. Elkton, MD 21921
## Contact Person
Mr. Mark Unterreiner Sr. Regulatory Affairs Specialist Ph: 410-392-7213 Fax: 410-398-6079 Email: mark.unterreiner@terumomedical.com
. "A
. . .
Date Prepared
April 11, 2008
.
{4}------------------------------------------------
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with its wings forming a flowing, abstract shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 8 2008
Terumo Medical Corporation c/o Mr. Mark Unterreiner Sr. Regulatory Affairs Specialist 950 Elkton Blvd. Elkton, MD 21921
Re: K081045
Trade/Device Name: Destination Renal Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class; Class II (two) Product Code: DYB Dated: May 19, 2008 Received: May 20, 2008
Dear Mr. Unterreiner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Mark Unterreiner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
uma R.v.luhner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K081045 |
|---------------------------|---------|
|---------------------------|---------|
Destination® Renal Guiding Sheath Device Name:
Indications For Use:
The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anna R. Kirchner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_ Ko81045