VESSIX GUIDE SHEATH 7F, 45 CM, RDC, VESSIX GUIDE SHEATH 7F, 45CM, LIMA

K140641 · Boston Scientific Corporation · DYB · Jul 3, 2014 · Cardiovascular

Device Facts

Record IDK140641
Device NameVESSIX GUIDE SHEATH 7F, 45 CM, RDC, VESSIX GUIDE SHEATH 7F, 45CM, LIMA
ApplicantBoston Scientific Corporation
Product CodeDYB · Cardiovascular
Decision DateJul 3, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The Vessix™ Guide Sheath is designed to be used for the introduction of interventional and diagnostic devices through the vasculature to the renal arteries.

Device Story

Vessix™ Guide Sheath is a 7F introducer sheath and guide catheter; facilitates delivery of interventional/diagnostic devices to renal arteries. Features braided shaft for torque response; pre-formed RDC or LIMA tip shapes; Tuohy-Borst hemostatic valve; sidearm with three-way stopcock for flushing/contrast media. Includes dilator for guidewire delivery (up to 0.038 in). Distal 15 cm has hydrophilic coating; distal tip is radiopaque. Used by physicians in clinical settings. Provides access to renal vasculature; enables device placement; prevents bleeding via valve. Benefits patient by facilitating minimally invasive renal procedures.

Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, reactivity, toxicity, pyrogenicity, hemolysis, hemocompatibility, complement activation, thromboresistance) and in-vitro performance (length, diameter, flow rate, burst pressure, visual appearance, radiopacity, kink resistance, tensile strength, torque strength, coating integrity, valve leakage).

Technological Characteristics

7F guide sheath; braided shaft; polycarbonate hub; Tuohy-Borst valve; hydrophilic coating (Bioslide™). Materials: polymer construction, PTFE inner liner, stainless steel braid. Dimensions: 45 cm effective length; accommodates ≤0.038 in guidewires. Sterilization: Ethylene Oxide. Radiopaque distal tip.

Indications for Use

Indicated for the introduction of interventional and diagnostic devices through the vasculature to the renal arteries.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 、「 K140641 p. 1 of 3 ﺎ ﺮ ン # 510(k) Summary per 21 CFR §807.92 : · ) l ( · । | Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311<br>Phone: 763-494-1700<br>Fax: 763-494-2222 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Name<br>and<br>Information | Kurtis Hunsberger<br>Senior Regulatory Affairs Specialist<br>Phone: 763-255-0303<br>Fax: 763-494-2222<br>e-mail: kurtis.hunsberger@bsci.com | | Date Prepared | June 26, 2014 | | Proprietary<br>Name | Vessix™ Guide Sheath | | Common Name | Guide Sheath | | Product Code | DYB | | Classification | Class II, 21 CFR Part 870.1340 | | Predicate<br>Device | Terumo Destination Renal Guiding<br>Sheath K081045 May 28, 2008 | | Device<br>Description | Vessix™ Guide Sheath is a 7F guide sheath designed to perform as an<br>introducer sheath and guide catheter for the introduction of interventional<br>and diagnostic devices through the vasculature to the renal arteries. The<br>Vessix Guide Sheath has been optimized for torque to allow access to<br>both renal arteries during procedures. The guide sheath has an effective<br>length of 45 cm and is available in two curve shapes: RDC (renal double<br>curve) and LIMA (left internal mammary artery).<br><br>The guide sheath has a braided shaft design to maximize torque<br>response, and a pre-formed tip shape (either RDC or LIMA) to enable<br>access. It is equipped with a Tuohy-Borst valve (hemostatic valve) to<br>prevent bleeding and a sidearm with a three-way stopcock to allow for<br>flushing and introduction of contrast media. It is also packaged with a<br>dilator to help with delivery over a guidewire. The guide sheath can<br>accommodate guidewires with diameters less than or equal to 0.038 in<br>(0.97 mm). The distal 15 cm of the outer surface of the guide sheath has a<br>hydrophilic coating (Bioslide™). The distal tip is radiopaque to help with<br>placement.<br><br>A polycarbonate hub is adhesively bonded to the proximal section of the<br>guide sheath. It incorporates a luer fitting which serves as a junction to the<br>hemostatic valve. | 1 {1}------------------------------------------------ ### Comparison of Technological Characteristics The Vessix ™ Guide Sheath incorporates substantially equivalent device materials and design, packaging materials and design, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the Terumo Destination Renal Guiding Sheath K081045 cleared May 28, 2008. Similarities and differences in technological characteristics between the predicate and subject device are listed below. Similarities: - Polymer material construction . - PTFE inner liner material . - Hydrophilic coating . - Sheath dimensions and shapes t - Dilator and Tuohy-Borst valve accessories ● - Ethylene Oxide sterilization . - Packaging design with same function . #### Differences: - Radiopaque extrusion (Vessix) versus radiopaque marker . (predicate) for visibility - Stainless steel braid (Vessix) versus stainless steel coil . (predicate) to provide wall support Performance Data The Vessix™ Guide Sheath was subjected to testing according to the requirements of Guidance for Industry and FDA Staff - Class il Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010. Bench testing and biocompatibility testing were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing and, therefore, this device may be considered substantially equivalent to the predicate device. The following biocompatibility and chemical characterization tests were completed on the Vessix™ Guide Sheath and its accessories: | Cytotoxicity | Hemolysis (Extract Method) | |---------------------------------|-----------------------------| | Sensitization | Partial Thromboplastin Time | | Intracutaneous Reactivity | In Vitro Hemocompatibility | | Acute Systemic Toxicity | Complement Activation | | Materials Mediated Pyrogenicity | In Vivo Thromboresistance* | | Hemolysis (Direct Contact) | USP <661> Physicochemical | *This test was performed only on the Ditator accessory. {2}------------------------------------------------ - ﻣﻌﺴﺴﺴﺴﺴ The following in-vitro performance tests were completed on the Vessix™ Guide Sheath: | Sheath Effective Length | Dye Flow Rate | |--------------------------------------|--------------------------| | Sheath Inner and Outer Diameter | Sheath Burst Pressure | | Dilator Inner Diameter | Device Visual Appearance | | Dilator Length | Radiopacity | | Dilator to Sheath Compatibility (OD) | Sheath Kink Resistance | | Sheath Tensile | Particulates | | Sheath to Hub Tensile | Torque Strength | | Dilator to Hub Tensile | Coating Integrity | Tuohy-Borst Valve Leakage Conclusion Based on the indications for use, technological characteristics, and safety and performance testing, the Vessix™ Guide Sheath has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Terumo Destination Renal Guiding Sheath. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird. July 3, 2014 Boston Scientific Corporation Attention: Mr. Kurtis Hunsberger Senior Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311 Re: K140641 Trade/Device Name: Vessix™ Guide Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: June 5, 2014 Received: June 6, 2014 Dear Mr. Hunsberger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 {4}------------------------------------------------ ## Page 2 - Mr. Kurtis Hunsberger If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hillemann for Bram D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Vessix™ Guide Sheath Indications for Use: The Vessix™ Guide Sheath is designed to be used for the introduction of interventional and diagnostic devices through the vasculature to the renal arteries. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.A. Hillemann
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...