REPROCESSED STEERABLE INTRODUCER

K123334 · Stryker Sustainability Solutions, Inc. · DYB · Jan 24, 2013 · Cardiovascular

Device Facts

Record IDK123334
Device NameREPROCESSED STEERABLE INTRODUCER
ApplicantStryker Sustainability Solutions, Inc.
Product CodeDYB · Cardiovascular
Decision DateJan 24, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The Reprocessed Steerable Introducer is indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

Device Story

Reprocessed steerable introducer set includes dilator, guidewire, and steerable sheath; provides flexible catheter positioning in cardiac anatomy. Features hemostasis valve to minimize blood loss; sideport with three-way stopcock for aspiration, fluid infusion, blood sampling, and pressure monitoring. Handle includes rotating collar for tip deflection (up to 90 degrees); distal vent holes facilitate aspiration; radiopaque tip marker aids fluoroscopic visualization. Used by physicians in clinical settings for cardiovascular catheterization. Device is reprocessed from predicate devices via cleaning and decontamination; individual units undergo functional testing prior to packaging. Benefits include safe, reliable access for cardiac procedures.

Clinical Evidence

Bench testing only. Testing included biocompatibility, validation of reprocessing, sterilization validation, functional testing, and packaging validation. No clinical data provided.

Technological Characteristics

Mechanical steerable introducer; materials identical to predicate devices. Features: hemostasis valve, sideport with three-way stopcock, rotating handle collar for tip deflection, distal vent holes, radiopaque tip marker. Non-powered, manual operation. Sterilization validated for reprocessed units.

Indications for Use

Indicated for patients requiring introduction of cardiovascular catheters into the heart, including left-sided access via the interatrial septum.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SECTION 5: 510(k) SUMMARY # JAN 2 4 2013 גן ל #### Submitter: Stryker Sustainability Solutions 1810 W Drake Dr Tempe, Arizona 85283 #### Contact: Amanda Babcock Regulatory Affairs Lead (480) 763-5300 (o) (863) 904-2312 (f) amanda.babcock@stryker.com Date of preparation: 24 October 2012 Name of device: Trade/Proprietary Name: Reprocessed Steerable Introducer Classification Name: Catheter introducer | Predicate Device | 510(k) Title | Manufacturer | |------------------|----------------------------------------------------|------------------| | K081645 | AGILIS NxT Steerable Introducer (Model<br>G408324) | St. Jude Medical | | K061363 | AGILIS NxT Steerable Introducer | St. Jude Medical | #### Device Description: The Reprocessed Steerable Introducer Set consists of a dilator, guidewire, and steerable sheath, which is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer is filtered with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling and pressure monitoring. A handle equipped with a rotating collar to deflect the tip clockwise less than or equal to 90 degrees. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization. ## Indications for Use: The Reprocessed Steerable Introducer is indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. # Technological Characteristics: The design, materials, and intended use of Reprocessed Steerable Introducer are identical to the predicate devices. The mechanism of Reprocessed Steerable Introducer is identical to the predicate devices in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of Steerable Introducer includes removal of adherent visible soil and decontamination. Each > Stryker Sustainability Solutions Reprocessed Steerable Introducer Traditional 510(k) **Page 13** {1}------------------------------------------------ individual Introducer is tested for appropriate function of its components prior to packaging and labeling operations # Performance data: Bench and laboratory testing was conducted to demonstrate performance of Reprocessed Steerable Introducer. This included the following tests: - � Biocompatibility - Validation of reprocessing ● - Sterilization Validation ● - Function test(s) � - Packaging Validation . Performance testing demonstrates that Reprocessed Steerable Introducer perform as originally intended. ## Conclusion: Stryker Sustainability Solutions concludes that the modified devices (Reprocessed Steerable Introducer) are substantially equivalent to the predicate devices as described herein. K12333 {2}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 JAN 2 4 2013 Stryker Sustainability Solutions, Inc. c/o Amanda Babcock Regulatory Affairs Lead 1810 W Drake Dr Tempe, Arizona 85283 Re: K123334 Trade Name: Reprocessed Steerable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: October 24, 2012 Received: October 26, 2012 Dear Ms. Babcock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of ICFR 807); as this (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you active spec.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Matthew G. Hillebrenner for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION 4: INDICATIONS FOR USE STATEMENT 510(k) Number (if known): )く1233334 Device Name: Reprocessed Steerable Introducer Indications For Use: The Reprocessed Steerable Introducer is indicated when introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.g. Willh (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K123334 Stryker Sustainability Solutions Reprocessed Steerable Introducer Traditional 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...