K250305 · Innovative Health · PNE · Jan 7, 2026 · Cardiovascular
Device Facts
Record ID
K250305
Device Name
Reprocessed Agilis NxT Steerable Introducer
Applicant
Innovative Health
Product Code
PNE · Cardiovascular
Decision Date
Jan 7, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Reprocessed Agilis NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.
Device Story
Reprocessed steerable introducer system; includes dilator, guidewire, and bi-directional steerable sheath. Sheath features hemostasis valve, sideport with three-way stopcock, and distal vent holes. Sheath handle includes rotating collar for tip deflection (clockwise ≥180°, counterclockwise ≥90°). Used by physicians in clinical settings to facilitate flexible catheter positioning within cardiac anatomy. Radiopaque tip marker and barium sulfate-filled sheath allow fluoroscopic visualization. Device enables fluid infusion, blood sampling, pressure monitoring, and aspiration. Reprocessed by Innovative Health via cleaning, decontamination, and functional testing; limited to two reprocessing cycles. Benefits include providing access for cardiac arrhythmia treatment while minimizing blood loss.
Clinical Evidence
No clinical data. Substantial equivalence established through bench and laboratory testing, including biocompatibility, cleaning and sterilization validation, functional testing (tensile, deflection, leak, friction), visual inspection, dimensional verification, simulated use, radiopacity, and particulate testing.
Technological Characteristics
Materials: braided stainless-steel wire, Pebax, Nylon, barium sulfate, platinum/iridium marker. Design: bi-directional steerable sheath with hemostasis valve and sideport. Dimensions: 8.5F ID, 11.5F OD, 61-71cm length. Energy: mechanical. Connectivity: none. Sterilization: validated process. Reprocessing: maximum two cycles, serialized and tracked.
Indications for Use
Indicated for patients undergoing treatment for cardiac arrhythmias requiring the introduction of cardiovascular catheters into the heart, including the left side of the heart.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 7, 2026
Innovative Health
Meerna Muradvich
Regulatory Affairs Engineer
1435 N. Hayden Rd.
Suite 100
Scottsdale, Arizona 85257
Re: K250305
Trade/Device Name: Reprocessed Agilis NxT Steerable Introducer
Regulation Number: 21 CFR 870.1340
Regulation Name: Catheter Introducer
Regulatory Class: Class II
Product Code: PNE
Dated: October 3, 2025
Received: October 6, 2025
Dear Meerna Muradvich:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250305 - Meerna Muradvich
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K250305 - Meerna Muradvich
Page 3
Sincerely,
FINN E.
DONALDSON - S
Digitally signed by FINN
E. DONALDSON - S
Date: 2026.01.07
09:57:57 -05'00'
For
Misti Malone, PhD
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
K250305 - Meerna Muradvich
Page 4
The following devices are included in the scope of this 510(k) premarket notification:
| Description | Item Number | French Size | | Useable Length (cm) | Curve Type | Curve Reach (mm) |
| --- | --- | --- | --- | --- | --- | --- |
| | | ID | OD | | | |
| Reprocessed Agilis NxT Steerable Introducer | 408309 | 8.5F | 11.5F | 71 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | 408310 | 8.5F | 11.5F | 71 | Medium Curl Dual Reach Bi-Directional | 22.4 |
| | G408324 | 8.5F | 11.5F | 71 | Large Curl Dual Reach Bi-Directional | 50.0 |
| | G408318 | 8.5F | 11.5F | 61 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | G408319 | 8.5F | 11.5F | 61 | Medium Curl Dual Reach Bi-Directional | 22.4 |
| | G408320 | 8.5F | 11.5F | 71 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | G408321 | 8.5F | 11.5F | 71 | Medium Curl Dual Reach Bi-Directional | 22.4 |
{5}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250305 | |
| Device Name Reprocessed Agilis NxT Steerable Introducer | |
| Indications for Use (Describe) The Reprocessed Agilis NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{6}
K250305
Page 1
# 510(k) Summary
## Submitter's Name and Address:
Innovative Health, LLC.
1435 N. Hayden Road, Suite 100
Scottsdale, AZ 85257
Contact Name and Information:
Meerna Muradvich
Regulatory Affairs Engineer
Innovative Health, LLC.
(480) 692-7176 (office)
(888) 965-7706 (fax)
mmuradvich@innovative-health.com
Amanda Babcock
Sr. Regulatory and Compliance Manager
Innovative Health, LLC.
(480) 525-5911 (office)
(888) 965-7706 (fax)
ababcock@innovative-health.com
## Date prepared:
January 31, 2025
## Device Information:
Trade/Proprietary Name: Reprocessed Agilis NxT Steerable Introducer
Common Name: Steerable Introducer
Classification Name: Reprocessed Catheter Introducer
Classification Number: Class II, 21 CFR 870.1340
Product Code: PNE
## Predicate Device:
| 510(k) Number | 510(k) Device | Manufacturer |
| --- | --- | --- |
| K251211 | Agilis NxT Steerable Introducer | St. Jude Medical/Abbott |
## Reference Device:
| 510(k) Number | 510(k) Device | Manufacturer |
| --- | --- | --- |
| K170311 | Reprocessed Agilis NxT Steerable Introducer | Innovative Health, LLC. |
| K230376 | Reprocessed Agilis NxT Steerable Introducer | Innovative Health, LLC. |
Innovative Health
Reprocessed Agilis NxT Steerable Introducer
Traditional 510(k)
{7}
K250305
Page 2
# Device Description:
The Reprocessed Agilis NxT Steerable Introducer consists of a dilator, guidewire, and bi-directional steerable introducer, which is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with a three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling and pressure monitoring. The device has either a small, medium, or large curl at the distal tip. The sheath handle is equipped with a rotating collar to deflect the tip clockwise $\geq 180^{\circ}$ and counterclockwise $\geq 90^{\circ}$ . The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization. The sheath material consists of braided stainless-steel wire covered with Pebax (polyether block amide) and Nylon. The sheath is filled with barium sulfate and the distal tip has a platinum/iridium marker for visualization under fluoroscopy. A plastic dilator and stainless-steel guidewire are packaged with the introducer and are designed to facilitate the introduction and passage of the introducer through the vasculature.
Note: Only the steerable sheath and dilator are the subject (reprocessed devices) of this submission. The guidewire is purchased off-the shelf (K935170) and packaged with the reprocessed devices.
# Indications for Use:
The Reprocessed Agilis NxT Steerable Introducer is indicated for the introduction of various cardiovascular catheters into the heart, including the left side of the heart, during the treatment of cardiac arrhythmias.
The item numbers included in the scope of this submission are as follows:
| Description | Item Number | French Size | | Useable Length (cm) | Curve Type | Curve Reach (mm) |
| --- | --- | --- | --- | --- | --- | --- |
| | | ID | OD | | | |
| Reprocessed Agilis NxT Steerable Introducer | 408309 | 8.5F | 11.5F | 71 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | 408310 | 8.5F | 11.5F | 71 | Medium Curl Dual Reach Bi-Directional | 22.4 |
| | G408324 | 8.5F | 11.5F | 71 | Large Curl Dual Reach Bi-Directional | 50.0 |
| | G408318 | 8.5F | 11.5F | 61 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | G408319 | 8.5F | 11.5F | 61 | Medium Curl Dual Reach Bi-Directional | 22.4 |
| | G408320 | 8.5F | 11.5F | 71 | Small Curl Dual Reach Bi-Directional | 16.8 |
| | G408321 | 8.5F | 11.5F | 71 | Medium Curl Dual Reach Bi-Directional | 22.4 |
Table 1: Item Numbers
Innovative Health
Reprocessed Agilis NxT Steerable Introducer
Traditional 510(k)
{8}
K250305
Page 3
# Technological Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Agilis NxT Steerable Introducer is identical to the predicate device. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of this device includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
# Functional and Safety Testing:
Bench and laboratory testing evaluated substantial equivalence of the Reprocessed Agilis NxT Steerable Introducer to the predicate device.
This included the following:
- Biocompatibility
- Cleaning Validation
- Sterilization Validation
- Functional Testing
- Visual Inspection
- Dimensional Verification
- Tensile Testing
- Deflection Testing
- Simulated Use Testing
- Leak Testing
- Radiopacity Testing
- Particulate Testing
- Friction Testing
- Packaging Validation
The Reprocessed Agilis NxT Steerable Introducer is reprocessed no more than two (2) times. Each device is marked, serialized, and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
# Conclusion:
Innovative Health concludes that the testing evaluated substantial equivalence to the predicate devices described herein.
Innovative Health
Reprocessed Agilis NxT Steerable Introducer
Traditional 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.