PathBuilder Steerable Introducer

K211530 · Shanghai Microport EP Medtech Co., Ltd. · DYB · Feb 8, 2022 · Cardiovascular

Device Facts

Record IDK211530
Device NamePathBuilder Steerable Introducer
ApplicantShanghai Microport EP Medtech Co., Ltd.
Product CodeDYB · Cardiovascular
Decision DateFeb 8, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The PathBuilder™ Steerable Introducer is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

Device Story

PathBuilder Steerable Introducer consists of a steerable sheath, dilator, and guidewire. Used by clinicians in professional healthcare environments to introduce cardiovascular catheters into heart chambers, including left-sided access via interatrial septum. Sheath features distal marker band, vent holes, and proximal hemostasis valve with sideport/three-way stopcock for aspiration, infusion, or pressure monitoring. Handle includes rotating collar for bidirectional tip deflection (clockwise ≥180°, counter-clockwise ≥90°). Dilator provides support and ensures smooth advancement; guidewire facilitates percutaneous placement. Radiopaque tip marker aids fluoroscopic visualization. Device minimizes blood loss during catheter exchange and facilitates navigation to complex heart structures. Benefits include improved access to cardiac anatomy and procedural efficiency.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-1), visual inspection, dimensional verification, packaging/transit testing, and mechanical performance (tensile force, flexion, fracture testing). Compatibility of components, bending load, curve force, insertion/withdrawal force, and simulated use fatigue testing were performed. All samples passed.

Technological Characteristics

Components: steerable sheath, dilator, guidewire. Materials: biocompatible per ISO 10993-1. Features: bidirectional deflection handle, hemostasis valve, sideport, radiopaque marker band. Sterilization: Ethylene Oxide (ISO 11135). Compliance: ISO 11070. Connectivity: None. Software: None.

Indications for Use

Indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. February 8, 2022 Shanghai MicroPort EP MedTech Co., Ltd. Tian Xia RA Engineer Building 23&28, Lane 588, Tianxiong Rd. Shanghai, Shanghai 201318 China Re: K211530 Trade/Device Name: PathBuilder Steerable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: November 29, 2021 Received: January 10, 2022 Dear Tian Xia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211530 Device Name PathBuilder Steerable Introducer Indications for Use (Describe) The PathBuilder Steerable Introducer is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for EverPace. The logo consists of a stylized "E" made up of a blue horizontal bar and a red circle. To the right of the "E" is the word "EverPace" in blue font. A small "TM" symbol is located in the upper right corner of the word "EverPace". ### 510(k) # K211530 510(k) Summary #### Submitter Information - Shanqhai MicroPort EP MedTech Co., Ltd. A. Company Name: - B. Company Address: Building 23&28, Lane 588, Tianxiong Rd., 201318 Shanghai, PEOPLE'S REPUBLIC OF CHINA - C. Company Phone: +86 21 38954600*3613 - D. Company Facsimile: +86 21 20903925 - E. Email: txia@everpace.com - F. Contact Person: Xia, Tian - G. Summary Prepared on: October 20, 2021 #### Identification of Proposed Device - PathBuilder™ Steerable Introducer A. Trade Name: - B. Common Name: Catheter Introducer - Introducer, Catheter C. Classification Name: - D. Regulation Number 21 CFR 870.1340 - E. Product Code: DYB - F. Device Class: Class II - G. Review Panel Cardiovascular - H. 510(k) Number K211530 ## Device Description PathBuilder™ Steerable Introducer is composed of a steerable sheath, a dilator and a guidewire. It is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. The steerable sheath is introduced into a body vessel, along with the dilator, over the guidewire, it is available in a variety of lengths, sizes and curve configurations; Dilator is assembled with the sheath and introduced into a body vessel over the guidewire. It is used to provide support to the sheath and ensure smooth advancement; the guidewire is percutaneously placed into the body vessel to function as a guide for the introduction into the chambers of the heart. The distal end of steerable sheath is embedded with a marker band and vent holes. The steerable sheath is finished at the proximal end with a hemostasis valve and sideport with a three-way stopcock. By turning the rotating collar on handle, the steerable sheath can be bidirectionally bended to facilitate the steerable sheath to reach complex heart structures more conveniently. The distal end of dilator has a tip, while its proximal end is connected with luer fitting. The inner diameter of dilator distal end is smaller than the main body for the purpose of locking the position of a puncture needle. The distal end of {4}------------------------------------------------ dilator has a fixed curve to adapt to the heart structure. The steerable introducer is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling and pressure monitoring. A handle equipped with a rotating collar to deflect the tip clockwise ≥180° and counter-clockwise ≥90°. The steerable introducer features distal vent holes to facilitate aspiration and minimize cavitation and a radiopaque tip marker to improve fluoroscopic visualization. # Intended Use Statement The PathBuilder™ Steerable Introducer is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. # Identification of Predicate Device - Agilis™ NxT Steerable Introducer A. Product Name: - B. Manufacturer: St. Jude Medical C. 510(k) Number: K081645 D. Regulation Number 21 CFR 870.1340 E. Product Code: DYB F. Device Class: Class II # Non-Clinical Performance Testing Non-clinical performance testing was completed for the PathBuilder™ Steerable Introducer to support its substantial equivalence to the predicate device. The test results demonstrated that the proposed device complies with the following standards and guidance: - (1) Biocompatibility Verification: The biological safety of the introducer was verified as per the requirements of ISO 10993-1:2018 Biological evaluation of medical devices- Part 1:Evaluation and testing within a risk management process and FDA's modified ISO guidelines in accordance with the FDA quidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". - Cytocompatiblity a) - Sensitization b) - Irritation C) - Acute Systemic Toxicity d) - Hemocompatibility e) {5}------------------------------------------------ - Complement Activation f) - Thrombogenicity q) - (2) Bench Validation: Validation testing of the introducer was performed to validate the design of the device . This testing included visual inspection, dimensional verification, packaging testing (visual inspection, sealing strength, dye penetration), transit testing and mechanical failure/performance (Peak tensile force of joints, flexion testing and fracture testing of quidewire). Additionally, the compatibly of the of the introducer, guidewire and dilator was validated with compatibility, bending load, controlling curve force, insertion withdrawal force, pushing force, simulated use fatigue testing (put the device through repeated physiologically relevant turns and observed for damage), radio detectability, freedom from leakage, dilator hub and 3way stopcock Luer testing. All samples were sterilized and aged prior to testing. All tested samples passed bench testing. - (3) Cleaning and Sterilization Validation: Validation testing was performed to demonstrate that the introducer could be sterilized in accordance with ISO 11135:2014 Sterilization of health care products-Ethylene Oxide: Requirements for development, validation and routine control of a sterilization process for medical device, AAMI TIR 28:2009 Product Adoption and Process Equivalency for Ethylene Oxide sterilization and the FDA guidance document. - (4) Shelf Life Validation: Validation testing was performed to demonstrate the shelf life of the introducer is three years. ## Clinical Tests Conclusion No clinical study was used to support this submission. | Description | Proposed Device | Predicate Device<br>(K081645) | Remark | | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------| | Product Code | DYB | DYB | SE | | | Regulation No. | 21 CFR 870.1340 | 21 CFR 870.1340 | SE | | | Class | Class II | Class II | SE | | | Intended Use | The PathBuilderTM<br>Steerable Introducer is<br>indicated for introducing<br>various cardiovascular<br>catheters into the heart,<br>including the left side of | The AgilisTM NxT<br>Steerable Introducer is<br>indicated for introducing<br>various cardiovascular<br>catheters into the heart,<br>including the left side of | SE | | | | the heart through the | the heart through the | | | | | interatrial septum. | interatrial septum. | | | | | The PathBuilder™<br>Steerable Introducer is | The Agilis™ NxT<br>Steerable Introducer is | | | | Indications for Use | indicated for introducing<br>various cardiovascular<br>catheters into the heart,<br>including the left side of<br>the heart through the<br>interatrial septum. | indicated for introducing<br>various cardiovascular<br>catheters into the heart,<br>including the left side of<br>the heart through the<br>interatrial septum. | SE | | | | Configuration | Steerable sheath | Steerable sheath | SE | | | | Dilator | Dilator | SE | | | | Guidewire | Guidewire | SE | | Functional<br>performance | Comply with ISO 11070:<br>2014 | Comply with ISO 11070:<br>2014 | SE | | | Biological<br>characteristics | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SE | | | | Total length of steerable<br>sheath | Total length of steerable<br>sheath | SE | | | | Outer diameter of<br>steerable sheath | Outer diameter of<br>steerable sheath | SE | | | | Inner diameter of<br>steerable sheath tip | Inner diameter of<br>steerable sheath tip | SE | | | | Curve shape of steerable<br>sheath tip | Curve shape of steerable<br>sheath tip | SE | | | Dimensional<br>Comparision | Effective length of dilator | Effective length of dilator | SE | | | | Outer diameter of dilator | Outer diameter of dilator | SE | | | | Inner diameter of dilator<br>tip | Inner diameter of dilator<br>tip | SE | | | | Curve shape of dilator tip | Curve shape of dilator tip | SE | | | | Effective length of<br>guidewire | Effective length of<br>guidewire | SE | | | | Outer diameter of<br>guidewire | Outer diameter of<br>guidewire | SE | | | | Inner diameter of J-curve<br>tip | Inner diameter of J-curve<br>tip | SE | | | | Intended<br>healthcare<br>environment | Professional healthcare<br>environment | Professional healthcare<br>environment | SE | | | | Sterilization<br>method | EO Sterilized | EO Sterilized | | Single use | | Yes | Yes | SE | | Shelf Life | 3 years | 3 years | SE | | #### Comparison to Predicate Device {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for EverPace. The logo consists of two stylized circles, one in blue and one in red, followed by the word "EverPace" in blue font. A small "TM" symbol is located to the right of the word "EverPace". Shanghai MicroPort EP MedTech Co., Ltd. Building 23&28, Lane 588, Tianxiong Rd., 201318 Shanghai, PEOPLE'S REPUBLIC OF CHINA 4 / 5 {7}------------------------------------------------ ## Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device. The proposed and predicate devices share the same intended use and fundamental scientific technology, including principles of operation and mechanism of action. Design and technological differences between the proposed and predicate devices do not raise any new concerns of safety and effectiveness. The results of verification and validation testing demonstrate that the PathBuilder™ Steerable Introducer is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...