R2P Radifocus Glidewire Advantage

K240818 · Terumo Corporation · DQX · Nov 26, 2024 · Cardiovascular

Device Facts

Record IDK240818
Device NameR2P Radifocus Glidewire Advantage
ApplicantTerumo Corporation
Product CodeDQX · Cardiovascular
Decision DateNov 26, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Intended Use

The R2P Radifocus Glidewire Advantage is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.

Device Story

Manual guide wire; used to direct catheters to specific anatomical locations in peripheral vasculature. Input: manual manipulation by clinician. Transformation: mechanical guidance through vasculature. Output: catheter positioning. Used in diagnostic or interventional procedures; operated by physicians. Benefits: facilitates navigation of peripheral vasculature for clinical procedures.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing including dimensional verification, simulated use, tensile strength, torque strength, torqueability, coating integrity, particulate evaluation, lubricity, corrosion resistance, kink resistance, tip flexibility, radiopacity, and ISO 10993 biocompatibility testing.

Technological Characteristics

Guide wire with nickel-titanium alloy or stainless steel core; polyurethane and hydrophilic coating; PTFE spiral coating. Dimensions: 0.018" or 0.035" diameter; lengths 350-500cm. Sterilization: Ethylene oxide (ISO 11135). Manual operation; no software or energy source.

Indications for Use

Indicated for directing catheters to desired anatomical locations in peripheral vasculature during diagnostic or interventional procedures. Not for neurovascular or coronary interventions.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the date November 26, 2024. The text is written in a clear, sans-serif font. The date is written in the standard US format, with the month first, followed by the day and year. The text is centered and takes up most of the image. Image /page/0/Picture/9 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the emblem of the Department of Health on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" being larger and bolder than the rest of the text. Terumo Medical Corporation Qing Liu Regulatory Affairs Specialist 265 Davidson Avenue, Suite 320 Somerset, New Jersey 08873 Re: K240818 Trade/Device Name: R2P Radifocus Glidewire Advantage Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 25, 2024 Received: November 1, 2024 Dear Qing Liu: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Samuel G. Raben -S for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K240818 Device Name R2P Radifocus Glidewire Advantage Indications for Use (Describe) The R2P Radifocus Glidewire Advantage is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(K) SUMMARY ## A. SUBMITTER INFORMATION (807.92(a)(1)) | Prepared by: | Qing Liu | |--------------|-------------------------------| | | Regulatory Affairs Specialist | | | Terumo Medical Corporation | | | Tel. (908) 842-9016 | | | Fax (410) 398-6079 | | Prepared for: | Owner/Operator | |---------------|--------------------------------| | | Terumo Corporation | | | 44-1, 2-Chome, Hatagaya | | | Shibuya-Ku, Tokyo | | | Japan 151-0072 | | | Owner/Operator Number: 8010026 | #### Manufacturer and Sterilization Facility (Applicant) Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya, Shizuoka 418-0015, Japan Registration Number: 9681834 ### Contact Person: Qing Liu Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, NJ 08873 Tel. (908) 842-9016 Fax (410) 398-6079 E-mail: kyo.ryu@terumomedical.com March 25, 2024 Date prepared: {5}------------------------------------------------ #### B. DEVICE NAME (807.92(a)(2)) | Proprietary Name: | R2P Radifocus Glidewire Advantage | |-----------------------|-----------------------------------| | Common Name: | Guide Wire | | Classification Name: | Catheter Guide Wire | | Classification Panel: | Cardiovascular | | Regulation: | 21 CFR 870.1330 | | Product Code: | DQX | | Classification: | Class II | ## C. PREDICATE DEVICE (807.92(a)(3)) The legally marketed device to which substantial equivalence is claimed is: Predicate Device: K152740 - Radifocus Glidewire, manufactured by Ashitaka Factory of Terumo Corporation. Reference Devices: - K063372 Radifocus Glidewire Advantage, Ashitaka Factory of Terumo Corporation - K163004 Radifocus Glidewire Advantage Track, Ashitaka Factory of Terumo Corporation - K033742 BOSTON SCIENTIFIC V-18 CONTROL WIRE, BOSTON SCIENTIFIC CORP - K172073 Hi-Torque Command 18 Guide Wire, Abbott Vascular - K150445 ASAHI Peripheral Guide Wires (ASAHI Gladius, ASAHI Halberd, and ASAHI Gaia PV), ASAHI Intecc Co., Ltd. # D. REASON FOR 510(k) SUBMISSION This traditional 510(k) for R2P Radifocus Glidewire Advantage is being submitted for a new device for the purposes of establishing substantial equivalence to a legally marketed predicate device. {6}------------------------------------------------ # E. DEVICE DESCRIPTION (807.92(a)(4)) #### Principle of Operation Technology The subject device, R2P Radifocus Glidewire Advantage, and the predicate device, Radifocus Glidewire (K152740), are both operated through a manual process. #### Design/Construction The subject device, R2P Radifocus Glidewire Advantage, and the predicate device, Radifocus Glidewire (K152740), exhibit some differences in design and construction. Terumo has confirmed that these differences don't introduce any new concerns in safety and performance compared to the predicate device. Table 1 provides details on the materials used in the construction and Table 2 provides specifications of R2P Radifocus Glidewire Advantage. | Raw material | | Patient contact | |--------------------------------------------------|-----------------------|-----------------| | 0.018" | 0.035" | | | Stainless steel | Nickel-Titanium alloy | Direct | | Nickel-Titanium alloy | | Direct | | Polyurethane containing tungsten | | Direct | | Hydrophilic polymer | | Direct | | Polytetrafluoroethylene (PTFE) | | Non-contact | | Polytetrafluoroethylene (PTFE)/ Colorant | | Direct | | Platinum/ Iridium (Pt/Ir) | | Direct | | Tin/Silver (Sn/Ag) | | Direct | | Gold (Au) | None | Non-contact | | Modified acrylate | None | Non-contact | | Silicone oil | | Indirect | | Polyethylene | | Indirect | | Polypropylene | | Non-contact | | Polypropylene/ Colorant | | Indirect | | Styrene-ethylenebutylene-styrene block copolymer | | Indirect | {7}------------------------------------------------ | Table 2: Specifications | | |-------------------------|------------------------------------------------------------------------| | Part | Specification | | Diameter of Wire | 0.018"<br>0.035" | | Length of Wire | 400, 450 and 500cm<br>350, 400, and 450cm | | Distal Tip Shape | Angled<br>Angled<br>J Shaped (Available only in 400 and 450cm lengths) | | Accessory Device | Guide wire inserter, Guide wire clip | #### Specifications #### F. INDICATIONS FOR USE (807.92(a)(5)) The R2P Radifocus Glidewire Advantage is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions. #### G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6)) R2P Radifocus Glidewire Advantage, the subject device of this Traditional 510(k), is substantially equivalent in its Intended Use/Indication for Use and Technological Characteristics to the predicate device, Radifocus Glidewire (K152740). In addition to the above-listed primary predicate, Terumo has identified the following reference devices. These are market leading devices with the similar intended use and basic design as the subject device. - K063372 Radifocus Glidewire Advantage, Ashitaka Factory of Terumo Corporation - K163004 Radifocus Glidewire Advantage Track, Ashitaka Factory of Terumo Corporation The comparison of the technological characteristics is summarized in Table 3.A and 3.B. {8}------------------------------------------------ | Comparison of<br>Device<br>Characteristics | Subject device:<br>R2P Radifocus Glidewire Advantage | Predicate Device: Radifocus<br>Glidewire(K152740) | Reference Device 1: Radifocus<br>Glidewire Advantage (K063372) | Comparison of<br>Device<br>Characteristics | Subject device:<br>R2P Radifocus Glidewire Advantage | Predicate Device: Radifocus<br>Glidewire(K152740) | Reference Device 2: Radifocus<br>Glidewire Advantage Track<br>(K163004) | |--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Ashitaka Factory of Terumo Corp. | Same | Same | Manufacturer | Ashitaka Factory of Terumo Corp. | Same | Same | | Intended<br>Use/Indication<br>for Use | The R2P Radifocus Glidewire<br>Advantage is designed to direct a<br>catheter to the desired anatomical<br>location in the peripheral vasculature<br>during diagnostic or interventional<br>procedures. This device is not intended<br>for neurovascular or coronary<br>interventions. | Same | The GLIDEWIRE<br>ADVANTAGE is designed to<br>direct a catheter to the desired<br>anatomical location in the<br>peripheral vasculature during<br>diagnostic or interventional<br>procedures. | Intended<br>Use/Indication for<br>Use | The R2P Radifocus Glidewire<br>Advantage is designed to direct a<br>catheter to the desired anatomical<br>location in the peripheral vasculature<br>during diagnostic or interventional<br>procedures. This device is not intended<br>for neurovascular or coronary<br>interventions. | Same | Same | | Operation<br>Principle | Manual | Same | Same | Operation<br>Principle | Manual | Same | Same | | Design/<br>Construction | The R2P Radifocus Glidewire<br>Advantage (0.035") consists of a<br>Nickel Titanium alloy core wire. A<br>polyurethane and hydrophilic coating is<br>applied to the distal portion of the wire.<br>• Core wire (proximal portion)<br>• Core wire (intermediate portion)<br>• Core wire (distal portion)<br>• First coating (distal portion)<br>• Second coating (distal portion)<br>Under coat<br>• PTFE spiral coating (intermediate<br>portion)<br>Under coating<br>Top coating<br>Spiral coating:<br>• Edge protection part<br>Metal part<br>Solder<br>• Silicone coating (proximal portion) | Fully hydrophilic guide Wire<br>• Core wire<br>• First Coating<br>• Second Coating<br>Under coat | Same<br>• Core wire (proximal and distal<br>portion)<br>• First coating (distal portion)<br>• Second coating (distal portion)<br>Under coat<br>• PTFE spiral coating proximal<br>portion)<br>Under coating<br>Top coating<br>Spiral coating<br>• Edge protection part<br>Metal part<br>Solder<br>Third coating on edge protection | Design/<br>Construction | The R2P Radifocus Glidewire<br>Advantage (0.018") consists of a<br>Nickel Titanium alloy and stainless<br>steel core wire. A polyurethane and<br>hydrophilic coating is applied to the<br>distal portion of the wire.<br>• Core wire (proximal portion)<br>• Core wire (intermediate portion)<br>• Core wire (distal portion)<br>• First coating (distal portion)<br>• Second coating (distal portion)<br>Under coat<br>• PTFE spiral coating (intermediate<br>portion)<br>Under coating<br>Spiral coating<br>• Edge protection part<br>Metal part<br>Solder<br>• Tip coil marker<br>• Tip coil marker adhesive<br>• Silicone coating (proximal portion) | Fully hydrophilic guide Wire. | Same | | Package | Individual package on which the<br>product label and the peel-off labels are<br>attached. 1 unit per package | Same | Same | | | • Core wire<br><br>• First Coating<br>• Second Coating<br>Under coat | • Core wire (proximal portion)<br>• Core wire (distal portion)<br><br>• First coating (distal portion)<br>• Second coating (distal portion)<br>Under coat<br>• PTFE spiral coating (proximal<br>portion)<br>Under coating<br>Top coating<br>Spiral coating<br>• Edge protection part<br>Metal part<br>Solder<br>Third coating on edge protection<br>• Tip coil marker<br>• Tip coil marker adhesive | | Specifications | • Diameter of Wire:<br>0.035" | • Diameter of Wire:<br>0.035" | • Diameter of Wire:<br>0.018 - 0.038" | Package | Individual package on which the<br>product label and the peel-off labels<br>are attached. 1 unit per package | Same | Same | | | • Length of Wire:<br>350,400,450 cm | • Length of Wire:<br>260, 300, 350, 400, 450cm | • Length of Wire:<br>150 - 300 cm | Specifications | • Diameter of Wire:<br>0.018"<br><br>• Length of Wire:<br>400,450,500 cm<br><br>• Shapes of Wire (distal tip)<br>Angled<br><br>• Accessory Devices:<br>Guide wire inserter<br>Guide wire clip | • Diameter of Wire:<br>0.035"<br><br>• Length of Wire:<br>260, 300, 350, 400, 450cm<br><br>• Shapes of Wire (distal tip)<br>Angled, Straight<br><br>• Accessory Devices:<br>Guide wire inserter | • Diameter of Wire:<br>0.014 and 0.018"<br><br>• Length of Wire:<br>180 and 300 cm<br><br>• Shapes of Wire (distal tip):<br>Angled<br><br>• Accessory Devices:<br>Guide wire inserter | | | • Shapes of Wire (distal tip)<br>350,400,450 cm: Angled<br>400,450 cm: J Shaped | • Shapes of Wire (distal tip)<br>Angled, Straight | • Shapes of Wire (distal tip):<br>Angled, straight, J Shaped | Sterilization | Ethylene oxide | Same | Same | | | • Accessory Devices:<br>Guide wire inserter<br>Guide wire clip | • Accessory Devices:<br>Guide wire inserter | • Accessory Devices:<br>Guide wire inserter | Shelf Life | 24 months | Same | Same | | Sterilization | Ethylene oxide | Same | Same | | | | | # Table 3.A: 0.035" vs. Predicate Device and Reference Device 1 {9}------------------------------------------------ # Table 3.B: 0.018" vs. Predicate Device and Reference Device 2 {10}------------------------------------------------ #### H. NON CLINICAL TESTS (807.92(b)(1)) #### Performance Testing Performance testing was conducted to ensure the safety and effectiveness of R2P Radifocus Glidewire Advantage throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device.Table 4 provides a list of performance tests that were performed on R2P Radifocus Glidewire Advantage. | Test Item | |-------------------------------| | Dimensional Verification | | Visual Inspection | | Simulated Use | | Tensile Strength and Tip Pull | | Torque Strength | | Torqueability | | Coating integrity | | Particulate evaluation | | Lubricity | | Corrosion resistance | | Kink Resistance | | Tip flexibility | | Radiopacity | Table 4: Summary of Performance Testing Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life. #### Biocompatibility In accordance with ISO 10993-1, R2P Radifocus Glidewire Advantage is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." {11}------------------------------------------------ Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. The Table 5 below provides a list of biocompatibility tests conducted on R2P Radifocus Glidewire Advantage. | Non-aged, sterile, whole device | |---------------------------------------------| | Cytotoxicity | | Sensitization | | Intracutaneous Reactivity | | Systemic Toxicity (Acute) | | Pyrogenicity | | Hemolysis (Direct and Indirect) | | Complement Activation (SC5b-9) | | <i>In vivo</i> Thrombogenicity | | <i>In vivo</i> Thrombogenicity(HEPARINIZED) | | Accelerated-aged (2 years), whole device | | Cytotoxicity | | Hemolysis (Direct and Indirect) | | Physicochemical | | FT-IR tests | #### Table 5: Summary of ISO 10993 Biocompatibility Testing #### Sterilization The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014/Amd 1:2018, Sterilization of Health Care Products – Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6. {12}------------------------------------------------ # I. CLINICAL TESTS (807.92(b)(2)) This 510(k) does not include data from clinical tests. # J. CONCLUSION (807.92(b)(3)) In summary, R2P Radifocus Glidewire Advantage, subject of this 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, Radifocus Glidewire (K152740), manufactured by Ashitaka Factory of Terumo Corporation.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...