The Wattson temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
Device Story
Dual-purpose 0.035" guidewire; facilitates delivery of interventional devices; provides temporary cardiac pacing. Distal pigtail shape; bipolar electrode configuration; stainless steel electrodes. Proprietary adapter converts proximal end to positive/negative terminal pins for connection to external pulse generators. Used in clinical settings (e.g., cath lab) by physicians during TAVR/BAV procedures. Device transmits electrical signals from external pulse generator to heart; enables rapid pacing during valve deployment. Not intended for permanent implantation. Benefits include combined guidewire functionality and pacing capability, reducing need for separate pacing leads.
Clinical Evidence
No human clinical data. Evidence includes biocompatibility testing per ISO 10993-1, bench testing for structural integrity and performance, and a GLP porcine animal study. Animal study compared subject to predicate; demonstrated equivalent safety, no sustained arrhythmias post-pacing, and no adverse histological findings.
Technological Characteristics
0.035" guidewire; stainless steel wire with FEP outer jacket and stainless-steel electrodes. 280 cm length. Bipolar electrode configuration (distal positive, proximal negative). Hydrophilic/silicone oil coating. Ethylene oxide sterilization. Single-use. Connects to external pulse generator via proprietary adapter. No software/AI components.
Indications for Use
Indicated for patients undergoing cardiac interventional procedures, including TAVR and BAV, requiring temporary cardiac pacing. Contraindicated for use in coronary arteries and cerebrovasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K252674 — Solo Pace Fusion System (SOLOFUSE1) · Solo Pace, Inc. · Jan 9, 2026
K160260 — Tempo Temporary Pacing Lead · Biotrace Medical, Inc. · Oct 25, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 5, 2023
Vascular Solutions Vite Beka Regulatory Manager 6464 Sycamore Court North Maple Grove, Minnesota 55369
Re: K230637
Trade/Device Name: Wattson Temporary Pacing Guidewire (2250) Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX, LDF Dated: March 7, 2023 Received: March 7, 2023
Dear Vite Beka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
ce determination does not mean that FDA
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed
Sara M. by Sara M. Royce
-S
Royce -S Date: 2023.05.05
19:34:49 -04'00'
for Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K230637
Device Name
Wattson temporary pacing guidewire (2250)
Indications for Use (Describe)
The Wattson temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: May 5, 2023
510(k) Number: K230637
### Submitter's Name / Contact Person
Manufacturer Vascular Solutions LLC 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812
Contact Person Beka Vite Regulatory Manager Tel: 612.441.0914 (direct)
#### General Information
| Trade Name | Wattson™ temporary pacing guidewire |
|---------------------|-----------------------------------------------------------------------------------|
| Common / Usual Name | Catheter Guidewire |
| Classification Name | 21 CFR 870.1330, Catheter Guidewire, DQX, Class II<br>Secondary Product Code: LDF |
| Predicate Device | K192454, Wattson temporary pacing guidewire (Vascular Solutions LLC |
#### Device Description
The Wattson temporary pacing guidewire is a dual-purpose 0.035" guidewire designed for the delivery of devices and for temporary pacing of the heart. The distal end of the device has an atraumatic pigtail shape. The shaft of the device has a hydrophilic coating.
Pacing is achieved through a bipolar electrode configuration integrated in the guidewire. The electrodes terminate in a single positive electrode at the distal tip of the device and three discrete negative electrodes located proximal to the tip within the pigtail section. The entire device is visible using standard fluoroscopic methods.
The device is packaged with a proprietary adapter that locks on to the proximal end of the guidewire and converts it to discrete positive and negative terminal pins. The terminal pins of the adapter are compatible with standard external pulse generators. Unshrouded adaptor pins are provided to ensure compatibility with all connecter cable terminals.
The Wattson temporary pacing guidewire is sterilized with ethylene oxide.
#### Indications for Use
The Wattson temporary pacing guidewire is intended to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve replacement (TAVR) procedures and balloon aortic valvuloplasty (BAV), while transmitting an electrical signal from an external pulse generator to the heart. The temporary pacing guidewire is not intended to remain in place following the clinical procedure.
{4}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device
A comparison of the technological characteristics between the Wattson temporary pacing guidewire and the predicate device are provided in the following table.
| Comparison of Technological Characteristics | | | |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristic | Subject Device | Primary Predicate Device | |
| | Wattson temporary pacing<br>guidewire | Wattson temporary pacing<br>guidewire | |
| Indications for Use | The Wattson™ temporary pacing<br>guidewire is intended to<br>introduce and position catheters<br>and other interventional devices<br>within the chambers of the<br>heart, including those used<br>within transcatheter aortic valve<br>replacement (TAVR) procedures<br>and balloon aortic valvuloplasty<br>(BAV), while transmitting an<br>electrical signal from an external<br>pulse generator to the heart. The<br>temporary pacing guidewire is<br>not intended to remain in place<br>following the clinical procedure. | The Wattson™ temporary pacing<br>guidewire is intended to introduce<br>and position catheters and other<br>interventional devices within the<br>chambers of the heart, including<br>those used within transcatheter<br>aortic valve replacement (TAVR)<br>procedures and balloon aortic<br>valvuloplasty (BAV), while<br>transmitting an electrical signal<br>from an external pulse generator<br>to the heart. The temporary pacing<br>guidewire is not intended to<br>remain in place following the<br>clinical procedure. | |
| Contraindication | The guidewire is contraindicated<br>for use in the coronary arteries<br>and in the cerebrovasculature. | The guidewire is contraindicated<br>for use in the coronary arteries<br>and in the cerebrovasculature. | |
| Anatomical sites | Left Ventricle | Left Ventricle | |
| Materials of construction | Stainless Steel Wire with<br>fluoropolymer outer jacket (FEP)<br>and stainless-steel electrodes | Stainless Steel Wire with<br>Polyetheretherketone outer jacket<br>(PEEK) and stainless-steel<br>electrodes | |
| Length | 280 cm | 280 cm | |
| Wire OD | 0.035"/0.89 mm | 0.035"/0.89 mm | |
| Method of Sterilization | Ethylene Oxide | Ethylene Oxide | |
| Single Use or Reusable | Single Use | Single Use | |
| Coating | Silicone Oil Lubricant | Hydrophilic coating | |
| Radiopacity | Yes | Yes | |
| Comparison of Technological Characteristics | | | |
| Characteristic | Subject Device | Primary Predicate Device | |
| | Wattson temporary pacing<br>guidewire | Wattson temporary pacing<br>guidewire | |
| Tip Configuration | Pigtail | Pigtail | |
| External Pacemaker<br>Connections | Removable guidewire adapter<br>terminating in two shrouded<br>positive/negative connectors | Removable guidewire adapter<br>terminating in two shrouded<br>positive/negative connectors and<br>two unshrouded adapter pins | |
| Pulse Generator<br>compatibility | Compatible with standard<br>external pulse generators | Compatible with standard external<br>pulse generators | |
{5}------------------------------------------------
# Performance Data
### Biocompatibility
The biocompatibility evaluation for the Wattson temporary pacing guidewire was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Wattson temporary pacing guidewire is considered an externally communicating device in contact with circulating blood and tissue for a limited period of time (<24 hours) during use. The battery of tests included the following:
- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity
- Acute Systemic Toxicity ●
- Pyrogenicity
- Hemolysis
- Complement Activation
- Thrombogenicity
# Performance Data - Bench
The device design was verified through performance, patient safety, structural integrity, infection, microbial and particulate contamination, and device interface tests.
# Performance Data - Animal:
A Good Laboratory Practice (GLP) Safety study was performed to demonstrate the substantial equivalence of the Wattson temporary pacing guidewire in comparison to the predicate Wattson temporary pacing guidewire in a porcine model. The testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies. The subject Wattson temporary pacing guidewire shares the same fundamental scientific technology as the predicate device.
The GLP Safety study demonstrated that the Wattson temporary pacing guidewire performed similarly to the predicate with regard to adverse events and other animal health concerns. No animals
{6}------------------------------------------------
experienced sustained arrhythmia in the post-rapid pacing monitoring period, all animals returned to normal sinus rhythm after rapid pacing was turned off, and there were no histological findings that indicate any differences between the subject device and its predicate.
### Conclusion
Results of biocompatibility, benchtop, and animal studies performed on the Wattson temporary pacing guidewire did not raise any new questions of safety or effectiveness compared to the predicate device. The Wattson temporary pacing guidewire is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.