K240864 · Opsens, Inc. · DQX · May 24, 2024 · Cardiovascular
Device Facts
Record ID
K240864
Device Name
PacePro Wire
Applicant
Opsens, Inc.
Product Code
DQX · Cardiovascular
Decision Date
May 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PacePro Wire is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures. Additionally, the PacePro Wire can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.
Device Story
PacePro Wire is a sterile, single-use, unipolar guidewire used by physicians in clinical settings to navigate interventional devices through blood vessels into heart chambers. It functions as a structural guidewire and a temporary pacing electrode. The device transmits electrical signals from an external pulse generator to the heart to provide temporary pacing. It is physically identical to the predicate SavvyWire but lacks the fiberoptic pressure-sensing capabilities. The physician connects the wire to an external pulse generator via an alligator clamp. The device facilitates transcatheter aortic valve procedures by providing both mechanical support for device delivery and electrical pacing capability, potentially reducing the need for additional pacing leads and improving procedural efficiency.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence was established through bench testing, including mechanical testing, particulate testing, and rapid pacing testing, alongside biocompatibility, sterilization, and packaging validation per ISO standards.
Technological Characteristics
Stainless steel guidewire, 0.035" OD, 280 cm length. Shaft and intermediate section coated with PTFE. Unipolar electrode configuration for temporary pacing. Compatible with standard external pulse generators. Sterilized via ethylene oxide. No fiberoptic sensor or fiber bundle present.
Indications for Use
Indicated for use in patients requiring introduction and positioning of interventional devices within heart chambers, including transcatheter aortic valve procedures, and for temporary intracardiac pacing. Contraindicated for use in cerebral vasculature and in the absence of anti-coagulation therapy.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
May 24, 2024
Opsens Inc. Marc Chaunet Regulatory Affairs Director 750. Boulevard du Parc Technologique Quebec, QC G1P 4S3 Canada
Re: K240864
Trade/Device Name: PacePro Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, LDF Dated: March 27, 2024 Received: March 28, 2024
Dear Marc Chaunet:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/8 description: The image shows a digital signature. The signature is for Sara M. Royce. The date of the signature is 2024.05.24, and the time is 12:10:01 -04'00'.
Sara Royce Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics
{2}------------------------------------------------
and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240864
Device Name PacePro Wire
Indications for Use (Describe)
The PacePro Wire is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures.
Additionally, the PacePro Wire can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div> <span> <b> <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div></td></tr><tr><td><div> <span> <b> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div></td></tr></table> | <div> <span> <b> <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> <b> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div> |
| <div> <span> <b> <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | | |
| <div> <span> <b> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div> | | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "opSens". The "op" is in orange, and the "Sens" is in gray. There are three orange dots above the "e" in "Sens".
# 510(K) Summary for PacePro Wire
## SUBMITTER 1.
Address: Opsens Inc.
750, Boulevard du Parc Technologique
Quebec (Quebec) G1P 4S3
Phone: 418.781.0333 ext 3408
Fax Number: 418-781-0024
Contact Person: Marc Chaunet, Regulatory Affairs Director
Email: marc.chaunet@opsens.com
Date Prepared: March 27, 2024
## DEVICE 2.
Name of Device: PacePro Wire
Common or Usual Name: Structural guidewire with temporary pacing capabilities
Classification name: Wire, guide, catheter; electrode, pacemaker, temporary
Regulatory Class: II Product
Code: DQX, LDF
### PREDICATE DEVICE 3.
The proposed PacePro Wire™ is substantially equivalent to the Opsens existing guidewire, SavyWire™ cleared in K213854 on 09/14/2022.
The subject device is physically the same product as the SavvyWire, without the pressure sensing capabilities. The predicate device has not been subject to a design-related recall.
### 4. DEVICE DESCRIPTION
The proposed PacePro Wire is a guidewire that includes functions for a structural wire and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.). Throughout the submission this will be referred to as either PacePro Wire or Opsens Pacing Wire (OPW). These two names are interchangeable and represent branding name vs. internal descriptive name.
{5}------------------------------------------------
## 5. INDICATIONS FOR USE
The PacePro Wire is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures.
Additionally, the PacePro Wire can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE 6. DEVICE
The proposed PacePro Wire is substantially equivalent to the SavvyWire™ cleared in K213854 on 09/14/2022.
The proposed PacePro Wire indications for use are essentially the same as for the SawyWire, minus the following portion "while measuring the pressure within the heart allowing calculation of hemodynamic parameters" as the subject device does not feature the pressurement capabilities of the cleared predicate SavvyWire.
The technological characteristics of the proposed PacePro Wire encompass the functionalities of the predicate device including functions for a structural wire, and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart). Each of these functions is fully supported by the relevant predicate device. Identical to the SawyWire, the PacePro Wire is unipolar. This necessitates the use of a separate patient electrode for the positive connection to the external pulse generator. The pacing capabilities of the PacePro Wire is tested to be equivalent to that of the predicate device. No new questions of safety and effectiveness were identified during the execution of Verification and Validation activities.
Therefore, the proposed devices, PacePro Wire meets substantial equivalence requirements with regards to the legally marketed predicate SavvyWire cleared in K213854.
For detailed comparison, refer to the Substantial Equivalence table on the following pages.
| | | Proposed Device | Predicate Device | Differences |
|------------------------|----------------------------|--------------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------|
| Regulatory Information | Name | Pace Pro Wire | SavvyWire™ | N/A |
| | 510(k)# | Pending | K213854 | N/A |
| | Predicates | K213854 | K151244, K191907 | N/A |
| | Product Code | DQX, LDF | DQX, DXO, LDF | Same except for DXO |
| | Class | 2 | 2 | Same |
| | Regulation<br>Number | 870.1330, 870.3680 | 870.1330, 870.2870, 870.3680 | Same except for 870.2870 |
| | Regulation<br>Generic Name | Wire, guide, catheter; electrode, pacemaker, temporary | Wire, guide, catheter; Transducer, pressure, catheter tip, electrode, pacemaker, temporary | Same except for "Transducer, |
# T ABLE 1: PACEPRO WIRE COMPARISON TABLE
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OpSens. The "op" part of the name is in orange, while the "Sens" part is in gray. There are three orange dots above the "Sens" part of the name. The logo is simple and modern.
| | | pressure, catheter<br>i tip" |
|--|--|------------------------------|
| | | |
| | | Proposed Device | Predicate Device | Differences |
|---------------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Regulation<br>Intended Use | Coiled wire fits inside a<br>percutaneous catheter for the<br>purpose of directing the<br>catheter through a blood<br>vessel.<br>The device is used to transmit<br>a pacing electrical stimulus<br>from the pulse generator to<br>the heart and/or to transmit<br>the electrical signal of the<br>heart to the pulse generator. | Coiled wire fits inside a percutaneous<br>catheter for the purpose of directing the<br>catheter through a blood vessel.<br>Catheter tip transmits mechanical or<br>electrical property changes in relation to<br>changes in blood pressure to accessory<br>equipment for processing.<br>The device is used to transmit a pacing<br>electrical stimulus from the pulse<br>generator to the heart and/or to transmit<br>the electrical signal of the heart to the<br>pulse generator. | Same except for<br>"Catheter tip<br>transmits<br>mechanical or<br>electrical property<br>changes in relation<br>to changes in blood<br>pressure to<br>accessory<br>equipment for<br>processing." |
| | Indications | The PacePro Wire™ is<br>intended for use to introduce<br>and position interventional<br>devices within the chambers<br>of the heart, including those<br>used for transcatheter aortic<br>valve procedures.<br>Additionally, the PacePro<br>Wire™ can be used for<br>temporary intracardiac<br>pacing by transmitting an<br>electrical signal from an<br>external pulse generator to<br>the heart. | The SavvyWire™ is intended for use to<br>introduce and position interventional<br>devices within the chambers of the<br>heart, including those used for<br>transcatheter aortic valve procedures,<br>while measuring the pressure within the<br>heart allowing calculation of<br>hemodynamic parameters.<br>Additionally, the SavvyWire™ can be<br>used for temporary intracardiac pacing<br>by transmitting an electrical signal from<br>an external pulse generator to the heart. | Same except for<br>"while measuring<br>the pressure within<br>the heart allowing<br>calculation of<br>hemodynamic<br>parameters" not<br>relevant for the<br>subject device<br>PacePro Wire. |
| | Contraindications | Cerebral vasculature &<br>absence of anti-coagulation<br>therapy | Cerebral vasculature & absence of<br>anticoagulation therapy | Same |
| | Prescription Use | Rx Only | Rx Only | Same |
| | System<br>Components | Sterile, disposable guidewire | Sterile, disposable guidewire | Same |
| Technological Characteristi<br> | Pressure Sensing<br>& Signal<br>Transmission<br>Technology | N/A | Fiberoptic sensor & fiber bundle<br>embedded in guidewire. | There are no<br>fiberoptic sensor &<br>fiber bundle<br>embedded in the<br>PacePro Wire<br>guidewire |
| | Sterile, Single<br>Use Patient<br>Contact<br>Component? | Yes | Yes | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "opSens". The "op" is in orange, while the "Sens" is in gray. There are three orange dots above the "ens" in the logo.
| Guidewire OD | 0.035" | 0.035" | Same |
|---------------------------------|---------------------------------------|---------------------------------------|------|
| Guidewire<br>Length | 280 cm | 280 cm | Same |
| Guidewire Shaft<br>Material | Stainless steel | Stainless steel | Same |
| Tip Coating | None (Stainless steel) | None (Stainless steel) | Same |
| Shaft Coating | Teflon (PTFE) | Teflon (PTFE) | Same |
| Intermediate<br>Section Coating | Teflon (PTFE) | Teflon (PTFE) | |
| Tip Configuration | Spiral (XS OD 3.2 cm,<br>S OD 4.2 cm) | Spiral (XS OD 3.2 cm,<br>S OD 4.2 cm) | Same |
| | Proposed Device | Predicate Device | Differences |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------|
| Guidewire Tip<br>Design | Coiled | Coiled | Same |
| Guidewire Tip<br>Length | XS: 2.9 cm<br>S: 3.8 cm | XS: 2.9 cm<br>S: 3.8 cm | Same |
| Radiopaque Tip? | Yes | Yes | Same |
| Pressure Sensor<br>Location | XS: 2.9 cm<br>S: 3.8 cm | XS: 2.9 cm<br>S: 3.8 cm | Same |
| Electrode | Unipolar | Unipolar | Same |
| External Pulse<br>Generator<br>Connection | Alligator clamp (negative<br>lead) is connected to one of<br>the PacePro Wire pacing<br>connection zones | Alligator clamp (negative lead) is<br>connected to one of the SavvyWire<br>pacing connection zones | Same |
| Pulse Generator<br>compatibility | Compatible with standard<br>external pulse generators | Compatible with standard external pulse<br>generators | Same |
#### PERFORMANCE DATA 7.
# Sterilization, Packaging, Shelf-life Testing:
The following standards are utilized to show equivalence of Sterilization, Packaging, and Shelf life:
- ISO 11135-1 (2014) Sterilization of health-care products - ethylene oxide - requirements for the development, validation and routine control of a sterilization process for medical devices. (Sterility)
- . ISO 11607-1 (2019) – Amendment 1:2023 Packaging for terminally sterilized medical devices - part 1: requirements for materials, sterile barrier systems and packaging systems
# Biocompatibility testing:
The biocompatibility evaluation was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (Attachment A) published June 16, 2016. This device is categorized in ISO
{8}------------------------------------------------
10993-1:2018 as "External communicating device – circulating blood" per section 5.2.3 c). The device will have limited exposure, a cumulative sum of single, multiple or repeated duration of contact up to 24 h. Testing completed on this device includes:
- Cytotoxicity
- . Sensitization
- Intracutaneous reactivity
- . Systemic toxicity (acute)
- . Hemocompatibility
- . Pyrogenicity
- SC5b-9 complement
# Design Verification:
Testing was conducted based on requirements of the latest FDA guidance document on guidewires: FDA Guidance Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff October 2019. All Design Requirements were successfully verified for the PacePro Wire device.
Design verification testing included:
- Mechanical Testing
- Particulate Testing
- . Rapid Pacing Testing
No new questions of safety and effectiveness were identified during execution of Verification and Validation activities.
# Animal studies:
No animal studies were necessary in support of the substantial equivalence of the PacePro Wire as the intended use/indications for use and technological characteristics are equivalent to the predicate device.
# Clinical Studies
No clinical studies were necessary in support of the substantial equivalence of the PacePro Wire as the intended use/indications for use and technological characteristics are equivalent to the predicate device.
## 8. CONCLUSIONS
The results from these tests mentioned above demonstrate that the technological and performance characteristics of the PacePro Wire is comparable to the predicate device, supports the substantial equivalence of the device that is the subject of this 510(k), and ensures the subject device can perform in a manner equivalent to the predicate device for the shared intended use.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "opSens". The "op" is in orange, and the "Sens" is in gray. There are three orange dots above the "s" in "Sens".
The results of the verification/validation tests have demonstrated that the PacePro Wire does not raise any new questions of safety and efficacy and is therefore substantially equivalent to the legally marketed predicate device, the SavvyWire cleared in K213854.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.