The SavvyWire™ is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures, while measuring the pressure within the heart allowing calculation of hemodynamic parameters. Additionally, the Savvy Wire™ can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart. The OptoMonitor 3 is intended to measure cardiovascular blood pressure, including in heart chambers, coronary vessels and peripheral vessels, during interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve for the diagnosis and treatment of blood vessels and such as valve gradients during structural heart procedures.
Device Story
SavvyWire™ is a 0.035" structural guidewire integrating pressure sensing and temporary pacing capabilities; used during TAVR procedures. Input: fiberoptic pressure signals from heart chambers; electrical pacing signals from external pulse generator. Transformation: fiberoptic sensor transmits pressure data to OptoMonitor 3; unipolar electrode transmits pacing pulses to heart. Output: hemodynamic parameters (valve gradients, pressure) displayed on OptoMonitor 3; pacing stimulus to myocardium. Used in cath lab/OR by interventional cardiologists. Benefits: enables simultaneous valve positioning, hemodynamic monitoring, and rapid ventricular pacing with a single device, reducing procedural complexity and vascular access requirements.
Clinical Evidence
Prospective, single-arm, two-center study (n=20) in patients ≥18 years undergoing TAVR for severe symptomatic aortic stenosis. Primary endpoints: device functionality for rapid pacing, valve delivery, and pressure measurement. Results confirmed successful rapid ventricular pacing and THV system positioning with both balloon-expandable and self-expandable valves. No new safety concerns identified.
Technological Characteristics
0.035" stainless steel guidewire; fiberoptic pressure sensor; unipolar pacing electrode. PTFE shaft coating. System includes OptoMonitor 3 (LCD, touchscreen, fiberoptic input). Sterilized via EtO. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-34. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients ≥18 years undergoing TAVR procedures due to severe symptomatic aortic stenosis. Used to introduce/position interventional devices, measure intracardiac pressure, and provide temporary intracardiac pacing. Contraindicated for use in cerebral vasculature and in patients with absence of anti-coagulation therapy.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 14, 2022
Opsens Inc. % Christina Henza Consultant Ultra Lifescience Solutions, Inc. 2811 Milton Ave, #409 Janesville, Wisconsin 53545
Re: K213854
Trade/Device Name: SavvyWire™ and OptoMonitor 3 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, DXO, LDF Dated: August 9, 2022 Received: August 9, 2022
Dear Christina Henza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K213854
Device Name SavvyWire™
Indications for Use (Describe)
The SavvyWire™ is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures, while measuring the pressure within the heart allowing calculation of hemodynamic parameters.
Additionally, the Savvy Wire™ can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K213854
Device Name OptoMonitor 3
Indications for Use (Describe)
The OptoMonitor 3 is intended to measure cardiovascular blood pressure, including in heart chambers,
coronary vessels and peripheral vessels, during interventional procedures.
Blood pressure measurements provide hemodynamic information, such as fractional flow reserve for the diagnosis and treatment of blood vessels and such as valve gradients during structural heart procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family:Wingdings;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family:Wingdings;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary SavvyWire™
#### SUBMITTER । .
Address: Opsens, Inc.
750, Boulevard du Parc Technologique
Quebec (Quebec) G1P 4S3
Phone: 418.781.0333 ext 3408
Fax Number: 418-781-0024
Contact Person: Marc Chaunet, Regulatory Affairs and Quality System Director
Email: marc.chaunet@opsens.com
Date Prepared: September 13, 2022
#### 2. DEVICE
Name of Device: SavvyWire™ and OptoMonitor 3
Common or Usual Name: structural wire with pacing and pressure sensing
Classification name: Wire, guide, catheter; Transducer, pressure, catheter tip, electrode, pacemaker, temporary
Regulatory Class: II
Product Code: DQX, DXO, LDF
#### PREDICATE DEVICE ന്
The proposed SavvyWire™ is substantially equivalent to the Opsens existing pressure guidewire, OptoWire III cleared in K191907 on 01/02/2020 (Primary Predicate), and the Wattson Temporary Pacing Guidewire cleared in K192454 on 01/15/2020 [Secondary predicate], with the Lake Region Medical Pre-Formed Guidewire (Boston Scientific Safari² Pre-Shaped TAVI Guidewire) cleared in K151244 on 06/11/2015 used as a primary reference device.
The primary and secondary predicates (Predicate #2) are included to support the pressure sensing and pacing indications / technologies as indicated below in accordance with the multiple predicates example 4 for a multi-parameter device from FDA guidance The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] issued on: July 28, 2014.
The Lake Region Medical Pre-Formed Guidewire (Boston Scientific Safari² Pre-Shaped TAVI Guidewire) cleared in K151244 on 06/11/2015 is used as a reference #1) for the mechanical testing performed on the SavvyWire to demonstrate substantial equivalence. The Pacel™ Bipolar Pacing
{5}------------------------------------------------
Catheters cleared in K152784 on 10/22/2015 is used as a secondary reference #2) device to support the pacing function. Bundled together with this new device is a minor change to the OptoMonitor 3 software included in K202943 cleared on 11/24/2020 (Predicate #3) which expands the software to include a separate TAVI interface for use with the new wire. The predicate and reference devices have not been subject to a design-related recall.
#### DEVICE DESCRIPTION 4.
The proposed SavvyWire™ is a new Catheter guidewire that includes functions for a structural wire, a pressure wire, and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.). Throughout the submission this will be referred to as either SavvyWire™ or Opsens Structural Wire (OSW). These two names are interchangeable and represent branding name vs. internal descriptive name.
SavvyWire™ is intended to be used with the OptoMonitor 3 (K202943), and ideally with the OpM3-DU TAVI, which is based on the approved OptoMonitor 3 (K202943), with a fully integrated TAVI software update.
#### ട. INDICATIONS FOR USE
SawyWire™: The SavvyWire™ is intended for use to introduce and position interventional devices within the chambers of the heart, including those used for transcatheter aortic valve procedures, while measuring the pressure within the heart allowing calculation of hemodynamic parameters.
Additionally, the SavyWire™ can be used for temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart.
OptoMonitor 3: The OptoMonitor 3 is intended to measure cardiovascular blood pressure, including in heart chambers, coronary vessels and peripheral vessels, during interventional procedures.
Blood pressure measurements provide hemodynamic information, such as fractional flow reserve for the diagnosis and treatment of blood vessels and such as valve gradients during transcatheter aortic valve procedures.
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed SavvyWire™ is substantially equivalent to the Lake Region Medical Pre-Formed Guidewire (Boston Scientific Safari² Pre-Shaped TAVI Guidewire) cleared in K151244 on on 06/11/2015 [primary predicate], Opsens existing pressure guidewire, OptoWire III cleared in K191907 on 01/02/2020 [Secondary Predicate], and the Wattson Temporary Pacing Guidewire cleared in K192454 on 01/15/2020 [Tertiary predicate]. The Pacel™ Bipolar Pacing Catheters cleared in K152784 on 10/22/2015 is used as a reference device to support the pacing function. The secondary and tertiary predicates are included to support the pressure sensing and pacing indications / technologies as indicated below in accordance with the multiple predicates example 4 for a multi-parameter device from FDA guidance The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] issued on: July 28, 2014. Bundled together with this new device is a minor change to the OptoMonitor 3 software included in K202943 cleared on 11/24/2020 (Predicate 4) which expands the software to include a separate TAVI
{6}------------------------------------------------
interface for use with the SavvyWire™. It is appropriate to bundle the two devices because they are intended to be used together as a system during a therapeutic or diagnostic procedure.
The proposed SavvyWire™ indications for use are a blend of the three predicate devices since the proposed device includes functionalities supported by each of these. The OptoMonitor 3 indications were updated slightly to encompass the new modality including addition of pressure measurements in heart chambers and addition of measurements such as aortic valve gradients during transcatheter aortic valve replacement procedures.
The technological characteristics of the proposed SawyWire™ encompass the functionalities of the predicate and reference devices including functions for a structural wire, a pressure wire, and a pacing wire (temporary intracardiac pacing by transmitting an electrical signal from an external pulse generator to the heart). Each of these functions is fully supported by the relevant predicate/reference devices. Beyond the major difference of combining these functionalities, there is a slight difference in the pacing wire. The SavvyWire™ is unipolar where the reference device and predicate device 3 are bipolar. This necessitates the use of a separate patient electrode for the positive connection to the external pulse generator. The pacing of the SavvyWire™ is tested to be equivalent to that of the reference device, Pacel™ Bipolar Pacing Catheters in an animal study comparing the two devices and clinical performance is also confirmed via a clinical study. No new questions of safety and effectiveness were identified during the execution of Verification and Validation activities.
Therefore, the proposed devices, SavvyWire™ and OptoMonitor 3 with TAVI meet substantial equivalence requirements with regards to the legally marketed predicate 1) the Lake Region Medical Pre-Formed Guidewire (Boston Scientific Safari² Pre-Shaped TAVI Guidewire) cleared in K151244, (Predicate 2) OptoWire III cleared in K191907, (Predicate 3) Watson Temporary Pacing Guidewire cleared in K192454 on 01/15/2020, (Predicate OptoMonitor 3 cleared in K202943 and the reference device, Pacel™ Bipolar Pacing Catheters cleared in K152784.
For detailed comparison, refer to the Substantial Equivalence tables on the following pages. Note that the comparison of the SavvyWire™ is included in Table 1: SavvyWire™ while the comparison for the OptoMonitor 3 is included in Table 2:OptoMonitor 3 with TAVI.
| SavvyWire ™ | | | | | | |
|-------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proposed<br>Device | Predicate 1<br>(Primary<br>Predicate) | Predicate 2 | Reference Device | Predicate 3 | |
| Regulatory Information | Name | SavvyWire™ | Pre-Formed<br>Guidewire<br>(BSC Safari2) | OptoWire III | Pacel Bipolar<br>Pacing Catheters | Wattson Temporary<br>Pacing Guidewire |
| | 510(k)# | K213854 | K151244 | K191907 | K152784 | K192454 |
| | Predicates | K151244,<br>K191907 | K130798 | K152991 | K875059 | K181001, K800298 |
| | Product Code | DQX, DXO,<br>LDF | DQX | DQX, DXO | LDF | DQX, LDF |
| | Class | 2 | 2 | 2 | 2 | 2 |
| SavvyWire™ | | | | | | |
| | Proposed<br>Device | Predicate 1<br>(Primary Predicate) | Predicate 2 | Reference Device | Predicate 3 | |
| | Regulation<br>Number | 870.1330,<br>870.2870,<br>870.3680 | 870.2870 | 870.1330,<br>870.2870 | 870.3680 | 870.1330, 870.3680 |
| | Regulation<br>Generic Name | Wire, guide,<br>catheter;<br>Transducer,<br>pressure,<br>catheter tip,<br>electrode,<br>pacemaker,<br>temporary | Wire, guide,<br>catheter; | Wire, guide,<br>catheter;<br>Transducer,<br>pressure,<br>catheter tip | Electrode,<br>pacemaker,<br>temporary | Wire, guide, catheter;<br>Electrode, pacemaker,<br>temporary |
| | Regulation<br>Intended Use | Coiled wire<br>fits inside a<br>percutaneous<br>catheter for<br>the purpose<br>of directing<br>the catheter<br>through a<br>blood vessel.<br><br>Catheter tip<br>transmits<br>mechanical or<br>electrical<br>property<br>changes in<br>relation to<br>changes in<br>blood<br>pressure to<br>accessory<br>equipment for<br>processing.<br><br>The device is<br>used to<br>transmit a<br>pacing<br>electrical<br>stimulus from<br>the pulse<br>generator to<br>the heart<br>and/or to<br>transmit the<br>electrical<br>signal of the<br>heart to the<br>pulse | Coiled wire<br>fits inside a<br>percutaneous<br>catheter for<br>the purpose<br>of directing<br>the catheter<br>through a<br>blood vessel. | Coiled wire fits<br>inside a<br>percutaneous<br>catheter for<br>the purpose of<br>directing the<br>catheter<br>through a<br>blood vessel.<br><br>Catheter tip<br>transmits<br>mechanical or<br>electrical<br>property<br>changes in<br>relation to<br>changes in<br>blood pressure<br>to accessory<br>equipment for<br>processing. | The device is used<br>to transmit a<br>pacing electrical<br>stimulus from the<br>pulse generator to<br>the heart and/or<br>to transmit the<br>electrical signal of<br>the heart to the<br>pulse generator. | Coiled wire fits inside a<br>percutaneous catheter<br>for the purpose of<br>directing the catheter<br>through a blood vessel.<br><br>The device is used to<br>transmit a pacing<br>electrical stimulus from<br>the pulse generator to<br>the heart and/or to<br>transmit the electrical<br>signal of the heart to<br>the pulse generator. |
| SavvyWire ™ | | | | | | |
| | Proposed<br>Device | Predicate 1<br>(Primary<br>Predicate) | Predicate 2 | Reference Device | Predicate 3 | |
| | Indications | The<br>SavvyWire™<br>is intended<br>for use to<br>introduce and<br>position<br>interventional<br>devices<br>within the<br>chambers of<br>the heart,<br>including<br>those used<br>for<br>transcatheter<br>aortic valve<br>procedures,<br>while<br>measuring<br>the pressure<br>within the<br>heart<br>allowing<br>calculation of<br>hemodynamic<br>parameters.<br>Additionally,<br>the<br>SavvyWire™<br>can be used<br>for temporary<br>intracardiac<br>pacing by<br>transmitting<br>an electrical<br>signal from an<br>external pulse<br>generator to<br>the heart. | The Pre-<br>Formed<br>guidewires<br>are intended<br>to facilitate<br>the<br>introduction<br>and<br>placement of<br>interventional<br>devices<br>within the<br>chambers of<br>the heart<br>including<br>those used<br>within<br>transcatheter<br>aortic valve<br>procedures. | To measure<br>pressure in<br>blood vessels<br>including both<br>coronary and<br>peripheral<br>vessels, during<br>diagnostic<br>angiography<br>and/or other<br>any<br>interventional<br>procedures.<br>Blood<br>pressure<br>measurements<br>provide<br>hemodynamic<br>information,<br>such as<br>fractional flow<br>reserve, for<br>the diagnosis<br>and treatment<br>of blood vessel<br>disease. | The St. Jude<br>Medical Pacel™<br>Bipolar Pacing<br>Catheters are<br>intended for<br>use in the<br>intracardiac<br>pacing and/or<br>ECG recording | The Wattson temporary<br>pacing guidewire is<br>intended to introduce<br>and position catheters<br>and other<br>interventional<br>devices within the<br>chambers of the heart<br>, including those used<br>within transcatheter<br>aortic valve<br>replacement (TAVR)<br>procedures and balloon<br>aortic valvuloplasty<br>(BAV), while<br>transmitting an<br>electrical signal from an<br>external pulse<br>generator to the heart.<br>The temporary pacing<br>guidewire is not<br>intended to remain in<br>place following the<br>clinical procedure. |
| SavvyWire™ | | | | | | |
| | | Proposed<br>Device | Predicate 1<br>(Primary<br>Predicate) | Predicate 2 | Reference Device | Predicate 3 |
| | Contraindications | Cerebral<br>vasculature &<br>absence of<br>anti-<br>coagulation<br>therapy | Cerebral<br>vasculature | Cerebral<br>vasculature | Patients with<br>recurrent sepsis<br>or with a<br>hypercoagulable<br>state should not<br>be considered<br>candidates for<br>transvenous<br>catheters since<br>the catheter could<br>serve as a focal<br>point for septic or<br>bland thrombus<br>formation. In<br>addition, patients<br>with a mechanical<br>tricuspid valve<br>should not be<br>considered for<br>ventricular pacing. | The guidewire is<br>contraindicated for use<br>in the coronary arteries<br>and in the<br>cerebrovascular. |
| | Prescription Use | Rx Only | Rx Only | Rx Only | Rx Only | Rx Only |
| | System<br>Components | Sterile,<br>disposable<br>guidewire | Sterile,<br>disposable<br>guidewire | Sterile,<br>disposable<br>guidewire | Sterile, disposable<br>catheter | Sterile, disposable<br>catheter |
| | Pressure Sensing<br>& Signal<br>Transmission<br>Technology | Fiberoptic<br>sensor & fiber<br>bundle<br>embedded in<br>guidewire. | None. | Fiberoptic<br>sensor & fiber<br>bundle<br>embedded in<br>guidewire. | None | None |
| | Sterile, Single<br>Use Patient<br>Contact<br>Component? | Yes | Yes | Yes | Yes | Yes |
| | Guidewire OD | 0.035" | 0.035" | 0.014" | Multiple catheter<br>sizes available. | 0.035" |
| | Guidewire<br>Length | 280 cm | 275 cm | 180 cm | 110 cm | 280 cm |
| | Guidewire Shaft<br>Material | Stainless steel | Stainless steel | Stainless Steel,<br>Nitinol | Stainless steel | Stainless Steel |
| Technological Characteristics | Tip Coating | None<br>(Stainless<br>steel) | Teflon spray<br>coating<br>(PTFE) | None<br>(Platinum-<br>Tungsten<br>alloy) | None (Stainless<br>steel) | None (Stainless steel) |
| | Shaft Coating | Teflon (PTFE) | Teflon (PTFE) | Teflon (PTFE) | Proprietary blend | Fluropolymer outer |
| | Intermediate<br>Section Coating | Teflon (PTFE) | Teflon (PTFE) | PET +<br>Hydrophilic<br>coating…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.