HYDROPHILIC COATED GUIDEWIRE

K133155 · Lake Region Medical · DQX · Apr 7, 2014 · Cardiovascular

Device Facts

Record IDK133155
Device NameHYDROPHILIC COATED GUIDEWIRE
ApplicantLake Region Medical
Product CodeDQX · Cardiovascular
Decision DateApr 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Intended Use

To facilitate the placement of devices during diagnostic and interventional procedures.

Device Story

Steerable metallic guidewire; Nitinol core; radiopaque polyurethane jacket; hydrophilic coating. Available in lengths 80-260cm; diameters 0.018"-0.038"; straight or angled distal tip. Used by clinicians during diagnostic and interventional procedures to navigate vasculature and facilitate placement of other medical devices. Device provides mechanical support and guidance; no electronic or software components.

Clinical Evidence

Bench testing only. Testing included dimensional, tensile strength, tip flexibility, coating adherence, torque control, kink resistance, stiffness, and radiopacity per ISO 11070 and FDA guidance. Biocompatibility testing performed per ISO 10993, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, and thrombogenicity.

Technological Characteristics

Nitinol core; polyurethane jacket; hydrophilic coating. Dimensions: 0.018"-0.038" diameter; 80-260cm length. Standards: ISO 10993 (biocompatibility), ISO 11070 (guidewires), ISO 15223, EN 980, ISO 594. Mechanical device; no energy source or software.

Indications for Use

Indicated for use in patients undergoing diagnostic and interventional procedures to facilitate the placement of medical devices.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Lake Region Medical. The logo consists of a circular symbol on the left, followed by the words "Lake Region" in a bold, serif font. Below "Region" is the word "Medical" in a smaller font size. PRODUCT: Hydrophilic Coated Guidewires SUBMISSION DATE: September 30, 2013 SUBMISSION TYPE: Traditional 510(k) K133155, 510(k) Summary Page 1 of 3 SECTION 1.0: 510(k) SUMMARY #### 1.1 MANUFACTURER / REGISTRATION INFORMATION | Lake Region Medical | Contact Person: | Tracy Meyer | |----------------------------------|-----------------|-----------------------| | 340 Lake Hazeltine Drive | Title: | Regulatory Specialist | | Chaska, MN 55318-1029 USA | Telephone: | 952-641-8510 | | FDA REGISTRATION NUMBER: 2126666 | Fax: | 952-448-3441 | #### 1.2 TRADE NAME (PROPRIETATY NAME) Hydrophilic Coated Guidewires #### 1.3 DEVICE COMMON NAMES/USUAL NAMES/CLASSIFICATION NAMES These devices are commonly known as guidewires, or spring guidewires. The current classification names and product codes are Wire, Guide, Catheter (DQX). #### CLASS OF DEVICE 1.4 The classification names listed above were originally classified as Class II devices by the Cardiovascular (74DQX) Review Panel. #### IDENTIFICATION OF PREDICATE DEVICE(s) 1.5 | 510(k)<br>NUMBER | MANUFACTURER | DEVICE NAME | |------------------|---------------------|------------------------------| | K000011 | Lake Region Medical | Hydrophilic Coated Guidewire | #### PROPOSED DEVICE DESCRIPTION 1.6 Utilizing proprietary processes, these guides are constructed from a steerable, metallic core (Nitinol) with a radiopaque polymer (polyurethane) jacket and a hydrophilic coating is applied over the jacket. Guidewires are available up to 260cm length and in diameters from 0.018" to 0.038" depending on specific design requirements. Guidewires may have a straight or angled distal tip. ### OUTSIDE DIAMETER: 0.018"-0.038" LENGTHS: 80cm - 260cm TIP SHAPE: Straight or Angled #### 1.7 COMPLIANCE WITH APPLICABLE STANDARDS The proposed Hydrophilic Coated Guidewires is in compliance with ISO 10993, ISO 11070, ISO 15223, EN 980, ISO 594. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a circular symbol to the left of the words "Lake Region" in a bold, sans-serif font. Below "Region" is the word "Medical" in a smaller font size. PRODUCT: Hydrophilic Coated Guidewires SUBMISSION DATE: September 30, 2013 SUBMISSION TYPE: Traditional 510(k) SECTION 1.0: 510(k) SUMMARY #### INTENDED USE STATEMENT 1.8 To facilitate the placement of devices during diagnostic and interventional procedures. #### 1.9 CONTRAINDICATIONS None Known #### 1.10 COMPARISON The proposed Hydrophilic Coated Guidewire is substantially equivalent to the predicate Hydrophilic Coated Guidewire with 510(k) number K000011. #### QUALIFICATION TESTING 1.11 The conclusions drawn from non-clinical and biocompatibility testing demonstrate the device is substantially equivalent to the claimed predicate device. ### BENCH TESTING In order to demonstrate equivalence of the proposed hydrophilic guidewire, Lake Region Medical performed testing to establish requirements. Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional attributes. Test methods were developed using FDA Coronary and Cerebrovascular Guidance and ISO 11070:1998. The following table lists the applicable bench tests performed at baseline and aging: - . Dimensional - FDA Device Compatibility . - FDA Tensile Strength ● - FDA Tip Flexibility . - FDA Coating Adherence/Integrity 0 - FDA Catheter Compatibility . - FDA Torque Control . - . FDA Combined Load - . Packaging Study - Particulate . - Kink Resistance . - Linear Stiffness . - Lateral Stiffness . - ISO Visual ● - ISO Fracture - ISO Flex . - ISO Corrosion . - ISO Strength of Union . - . ISO Radiopacity - Body Stiffness . - Tip Integrity . - Jacket Durability . - Tip Puncture Resistance . - Coating Performance Testing . - 3 Point Bend . {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a circular graphic on the left, followed by the words "Lake Region" in a bold, serif font. The word "Medical" is written in a smaller font below "Region." PRODUCT: Hydrophilic Coated Guidewires SUBMISSION DATE: September 30, 2013 SUBMISSION TYPE: Special 510(k) # K133155, 510(k) Summary Page 3 of 3 8 | P a g e #### 1 , 1 1 QUALIFICATION TESTING Continued ### BIOCOMPATIBILITY TESTING Biocompatibility testing per ISO 10993 series has been performed on the proposed Hydrophilic Guidewires and has been found to be acceptable: - ISO Cytotoxicity . - . ISO Klingman Maximization Test - ISO Irritation / Intracutaneous Reactivity . - ISO Systemic Toxicity . - . ISO Rabbit Pyrogen - . ASTM Hemolysis - ISO Complement Activation Assay - ISO Thrombogenicity - ISO Lee and White Coagulation � - ISO Prothrombin Time Assay ◆ - ISO Unactivated Partial Thromboplastin Time . Assay #### 1.12 SUBSTANTIAL EQUIVALENCE DATA The proposed Hydrophilic Coated Guidewire has the same intended use as predicate Hydrophilic Coated Guidewires legally marketed by Lake Region Medical and cleared by 510(k) K000011. The proposed Hydrophilic Coated Guidewires have the same physical characteristics as the predicate device with the exception of the change to the hydrophilic coating. All required tests results support the claim of substantial equivalence. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ## April 7, 2014 Lake Region Medical % Tracy Meyer, Regulatory Specialist 340 Lake Hazeltine Drive Chaska, MN 55318-1029 Re: K133155 > Trade/Device Name: Hydrophilic Coated Guidewires Regulation Number: 21 CFR 870.1330 Regulation Name: Guidewire Catheter Regulatory Class: Class II Product Code: DQX Dated: March 5, 2014 Received: March 6, 2014 Dear Tracy Meyer, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Tracy Meyer Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Image /page/4/Picture/5 description: The image shows the name "Bram D. Zuckerman" in bold, black font. Below the name is the letter "S" in a similar font. To the right of the name and the letter is the acronym "FDA" in a stylized, outlined font. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ C Lak PRODUCT: Hydrophilic Coated Guidewires SU8MISSION DATE: September 30, 2013 SUBMISSION TYPE: Traditional 510(k) # INDICATIONS FOR USE 510(k) NUMBER (IF KNOWN): K133155 DEVICE NAME: Hydrophilic Coated Guidewire INDICATIONS FOR USE: To facilitate the placement of devices during diagnostic and interventional procedures. PRESCRIPTION × USE OVER-THE-.4ND/OR COUNTER USE (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) PAGE 1 of Tanya X. Shortus Kenya Brothers -S 2014.04.07 17:30:11 -04'00' 23 | Pag C
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...