CrossTorq 14 Guidewire

K193194 · Scientia Cardio Access, LLC · DQX · Dec 13, 2019 · Cardiovascular

Device Facts

Record IDK193194
Device NameCrossTorq 14 Guidewire
ApplicantScientia Cardio Access, LLC
Product CodeDQX · Cardiovascular
Decision DateDec 13, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The CrossTorqTM 14 Guidewire is intended for general vascular use within the coronary and peripheral vasculatures to introduce and position catheters and other interventional devices. The CrossTorqTM 14 Guidewire is not intended for use in the neurovasculature.

Device Story

CrossTorq 14 Guidewire is a 0.014" diameter steerable guidewire with a shapeable tip; used to access vasculature and position interventional devices (catheters, balloons). Features a stainless steel core wire, radiopaque platinum marker coil at the distal tip, hydrophilic coating on the distal portion, and PTFE coating on the proximal portion. Available in 200cm and 300cm lengths with four tip stiffness profiles (Soft, Standard, Support, Extra Support). Supplied sterile for single use. Includes accessories: shaping mandrel, torque device, and introducer. Operated by physicians in clinical settings (e.g., cath lab) under fluoroscopic guidance. Provides mechanical support and navigation through tortuous vasculature; facilitates catheter placement. Benefits patient by enabling minimally invasive access to coronary and peripheral vessels.

Clinical Evidence

Bench testing only. No clinical data. Performance evaluated via ISO 11070:2014 and FDA Coronary and Cerebrovascular Guidewire Guidance. Tests included visual inspection, dimensional verification, flexing, tensile strength, tip shape retention, torqueability, torque strength, tip flexibility, fracture resistance, coating lubricity/durability, particulate analysis, and column buckling. Biocompatibility and sterilization validation leveraged from reference devices (Aristotle 14/18).

Technological Characteristics

0.014" diameter guidewire; stainless steel core wire; distal tip radiopaque platinum marker coil; hydrophilic distal coating; proximal PTFE coating. Sterilized via 100% Ethylene Oxide (EO). Complies with ISO 11070:2014. Accessories: shaping mandrel, torque device, introducer. Labeled MRI Unsafe.

Indications for Use

Indicated for general vascular use in coronary and peripheral vasculatures to introduce and position catheters and interventional devices. Not for neurovascular use.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is in blue and consists of the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the next line. To the left of the FDA logo is the Department of Health & Human Services logo. The Department of Health & Human Services logo is a symbol of a stylized human figure. September 16, 2020 Scientia Cardio Access LLC % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k, Saint Paul, Minnesota 55114 Re: K193194 Trade/Device Name: CrossTorq 14 Guidewire Regulation Number: 21 CFR 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dear Prithul Bom: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 13, 2019. Specifically, FDA is updating this SE Letter as an administrative correction. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Lydia Glaw, OHT2: Office of Cardiovascular Devices, 301-796-1456, Lydia.glaw@fda.hhs.gov. | Sincerely, | | |----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | Lydia S.<br>Glaw -S | Digitally signed by<br>Lydia S. Glaw -S<br>Date: 2020.09.16<br>10:45:14 -04'00' | | Lydia Glaw<br>Assistant Director | | | DHT2C: Division of Coronary<br>and Peripheral Interventional Devices | | | OHT2: Office of Cardiovascular Devices | | | Office of Product Evaluation and Quality<br>Center for Devices and Radiological Health | | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size. December 13, 2019 Scientia Cardio Access LLC % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k, Saint Paul, Minnesota 55114 Re: K193194 Trade/Device Name: CrossTorq 14 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: November 18, 2019 Received: November 19, 2019 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/ training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/6 description: The image shows the name "Lydia S. Glaw -S" in large, bold font on the left side of the image. On the right side, it says "Digitally signed by Lydia S. Glaw -S" followed by the date "Date: 2019.12.13" and the time "09:46:48 -05'00'". The text on the right side is smaller and less bold than the name on the left. Lydia Glaw, Ph.D. Assistant Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K193194 Device Name CrossTorq 14 Guidewire Indications for Use (Describe) The CrossTorqTM 14 Guidewire is intended for general vascular use within the coronary and peripheral vasculatures to introduce and position catheters and other interventional devices. The CrossTorqTM 14 Guidewire is not intended for use in the neurovasculature. | Type of Use (Select one or both, as applicable) | |----------------------------------------------------------------------------------| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/2 description: The image shows the logo for Scientia. The logo consists of a stylized blue "S" shape above the word "SCIENTIA" in a futuristic-looking font. The "S" shape is made up of three parallel lines that curve around each other. # 510(K) SUMMARY (Per 21 CFR 807.92) # SCIENTIA CARDIO ACCESS LLC | 510(k) Sponsor: | Scientia Cardio Access LLC<br>3487 West 2100 South Suite 100<br>West Valley City, UT 84119<br>Tel: (775) 657-6330 | |---------------------|---------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Amy McManus, Regulatory Affairs Manager<br>Phone: 1 (888) 385-9016<br>Email: amcmanus@scientiavascular.com) | | Date Prepared: | December 6, 2019 | | Prepared by: | Ryan O'Callaghan, MS, RAC<br>Phil Triolo and Associates LC<br>Tel: 801.573.2651<br>Fax: 801.328.2399<br>E-mail: ryano@philt.com | | Trade Name: | CrossTorqTM 14 Guidewire | | Common Name: | Guidewire | | Classification Name | Catheter Guide Wire per 21 CFR 870.1330 | | Product Code: | DQX | | Predicate Devices: | KinetixTM Guidewire (K081021) | | | HornetTM Guidewire, HornetTM 10 Guidewire, and HornetTM 14 Guidewire<br>(K152231) | | Reference Devices: | Aristotle 14 Guidewire (K173235) | | | Aristotle 18 Guidewire (K183608) | {5}------------------------------------------------ #### DEVICE DESCRIPTION Scientia Cardio Access's CrossTorg™ 14 Guidewire is a 0.014" diameter steerable guidewire with a shapeable tip to aid in accessing vasculature. The guidewire is supplied sterile and is for single use only. It is provided with the 4 different tip loads, or tip stiffnesses: 1 gram. 10 gram, and 15 gram. These tip stiffnesses are referred to as Soft, Standard, Support, and Extra Support. respectively. The product is offered in lengths of 200cm and 300cm. The distal portion of the guidewire tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. The guidewire has a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels. The guidewire is provided with a shaping mandrel (to shape the flexible tip of the guidewire), an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve) and a torque device (to attach to the proximal portion to facilitate gripping and manipulation of the guidewire during use). The shaping mandrel, introducer and torque accessory devices are included to facilitate use of the guidewire and are not intended to contact the patient's body. The CrossTorg 14 Guidewire is substantially equivalent with respect to technological characteristics, design and materials to Boston Scientific's currently marketed Kinetix™ Guidewire cleared under K081021 and Boston Scientific's Hornet™ Guidewire, Hornet™ 10 Guidewire, and Hornet™ 14 Guidewire (also referred to as the Hornet Guidewire Series in this summary) cleared under K152231. ### INDICATIONS FOR USE The CrossTorq™ 14 Guidewire is intended for general vascular use within the coronary and peripheral vasculatures to introduce and position catheters and other interventional devices. The CrossTorg™ 14 Guidewire is not intended for use in the neurovasculature. ### TECHNOLOGICAL CHARACTERISTICS As shown in the table below, the technological characteristics of the CrossTorg 14 Guidewire are equivalent to those of the predicate devices, the Kinetix Guidewire and Boston Scientific Hornet Guidewire Series. Both of the predicate devices have intended use; however, the Kinetix Guidewire is only offered in soft tip stiffness profiles, while the Hornet Guidewire Series and the CrossTorq 14 Guidewire are offered in a wider range of stiffness profiles. The CrossTorq 14 Guidewire combines the stiffness profiles offered by the Kinetix and Hornet Guidewire Series guidewires. {6}------------------------------------------------ | Comparison of Technological Characteristics between Subject Device and Predicate Devices | | | | | |------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Characteristic | Subject Device<br>CrossTorqTM 14<br>Guidewire | Primary Predicate<br>Device<br>KinetixTM Guidewire<br>(K081021) | Predicate Device<br>Boston Scientific Hornet<br>Guidewire Series<br>(K152231) | Comparison | | Anatomical<br>Location | Coronary and<br>peripheral vasculature | Coronary and<br>peripheral vasculature | Coronary and peripheral<br>vasculature | Same | | Dimensions | <i>O.D.</i> : 0.014"<br>(0.36mm)<br><i>Length</i> : 200cm or<br>300cm | <i>O.D.</i> : 0.014"<br>(0.36mm)<br><i>Length</i> : 180cm or<br>300cm | <i>O.D.</i> : 0.014" (0.36mm)<br><i>Length</i> : 190cm or<br>300cm | Equivalent | | Core Wire | Stainless Steel | Nitinol | Stainless Steel | Equivalent | | Distal Tip | Shapeable straight tip<br><i>Length</i> : 35 cm<br><i>Material</i> : Nitinol | Shapeable straight tip<br>or preshaped "J" tip<br><i>Length</i> : 18 cm<br><i>Material</i> : Nitinol | Shapeable straight tip<br><i>Length</i> : 15 cm<br><i>Material</i> : Stainless Steel | Equivalent | | Stiffness<br>Profiles | Range from<br>Soft(1gram) to Extra<br>Support (15gram) | Range from Moderate<br>Support(0.8 gram) to<br>Plus(1.3gram) | Range from 1gram to 14<br>gram tip stiffness | Equivalent | | Coatings | <i>Distal End</i> :<br>Hydrophilic<br><i>Proximal End</i> :<br>PTFE | <i>Distal End</i> :<br>Hydrophilic<br><i>Proximal End</i> : PTFE | <i>Distal End</i> : Hydrophilic<br><i>Proximal End</i> : PTFE | Same | | Radiopaque<br>Marker | Radiopaque marker at<br>distal tip | Radiopaque marker at<br>distal tip | Radiopaque marker at<br>distal tip | Same | | Accessories | The following<br>accessories are<br>provided with the<br>guidewire:<br>• Shaping Mandrel<br>• Torque Device<br>• Plastic Introducer | No accessories are<br>provided with the<br>predicate device.<br>The predicate device's<br>IFU references use of<br>the guidewire with the<br>following accessories:<br>• Shaping instrument<br>• Torque device<br>• Insertion tool | No accessories are<br>provided with the<br>predicate device.<br>The predicate device's<br>IFU references use of the<br>guidewire with the<br>following accessories:<br>• Shaping instrument<br>• Torque device<br>• Insertion tool | Equivalent | | Comparison of Technological Characteristics between Subject Device and Predicate Devices | | | | | | Characteristic | Subject Device<br>CrossTorqTM 14<br>Guidewire | Primary Predicate<br>Device<br>KinetixTM Guidewire<br>(K081021) | Predicate Device<br>Boston Scientific Hornet<br>Guidewire Series<br>(K152231) | Comparison | | Sterilization<br>Method | 100% Ethylene Oxide<br>(EO) | 100% Ethylene Oxide<br>(EO) | 100% Ethylene Oxide<br>(EO) | Same | {7}------------------------------------------------ As shown in the Table above, the CrossTorq 14 Guidewire differs from the predicate guidewire in the following ways: The CrossTorq 14 core wire is constructed of stainless steel, while the Kinetix core wire is constructed of Nitinol. The Hornet Guidewire Series guidewires are all built using the same stainless steel material as the CrossTorq 14. Stainless steel is a commonly used core wire material and the functionality and performance characteristics of the stainless steel core wire are similar to those of the Nitinol core wire used in the construction of the Kinetix Guidewire. An accessory kit (consisting of a shaping mandrel, a torque device, and an introducer) is provided as part of the subject device, whereas the predicate devices do not provide an accessory kit as part of their finished product. Although the accessories are not provided in the packaging, the Kinetix and Hornet Guidewire Instructions For Use mentions the use of a shaping instrument, insertion tool, and torque device, which are equivalent to the accessories provided with the CrossTorq 14. The tip stiffness profiles available for the CrossTorq 14 Guidewire are: 1gram, 6 gram, and 15 gram. These tip stiffnesses are referred to in reports and this document as Soft. Standard, Support, and Extra Support, respectively. The predicate Kinetix Guidewire is only available in 0.8 gram and 1.3 gram tips, however, the predicate Hornet Guidewire series is available in 1 gram, 10 gram, and 14 gram tips, and thus can be used to determine values for comparative testing of Tip Stiffness and Column Buckling. Results of tests performed on the new CrossTorq 14 Guidewire demonstrate that the new guidewire performs as well as the predicate devices and/or meets requirements of relevant standards. Further, any differences in technological characteristics of the CrossTorq 14 Guidewires when compared with predicate device characteristics do not raise different questions of safety and effectiveness. {8}------------------------------------------------ Scientia Cardio Access LLC Traditional_510(k) CrossTorq™ 14 Guidewire Additionally, the indications for use statement of the subject device is equivalent to that of the predicate devices and the differences in wording are not critical to the intended therapeutic, diagnostic, prosthetic, surgical or other use of the device. The differences do not affect the safety and effectiveness of the device when used as labeled, as the same intent is documented for the predicate devices and the new device. #### NON-CLINICAL PERFORMANCE TESTS #### Biocompatibility The materials and processing of the CrossTorq 14 Guidewires are identical to those of the reference Aristotle 18 Guidewire (manufactured by Scientia Cardio Access LLC's sister company, Scientia Vascular LLC, and cleared under K183608); in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents) with the exception that the outer diameter of the CrossTorg 14 Guidewire is smaller than that of the Aristotle 18 Guidewire. The dimensional change does not alter the chemical or physical properties of the medical device in its final finished form, and therefore, results from the Aristotle 18 Guidewire can be applied to the medical device in its final finished form. There are no additional biologic interactions that need to be considered in order to use the cleared guidewires in the coronary vasculature. Consequently, no additional evaluations are needed to determine that the subject guidewires present a low and acceptable biological and toxicological risk when used in accordance with their intended and indicated uses, and no additional biological safety information is provided in this 510(k). {9}------------------------------------------------ Г #### Test Results Leveraged from Premarket Notifications K173235 (Aristotle 14 Guidewire) and K183608 (Aristotle 18 Guidewire) Results for following tests, performed on the reference Aristotle 14 Guidewires and reference Aristotle 18 Guidewires, were included in Scientia Vascular LLC's premarket notifications K173235 and K183608 for the Aristotle 14 Guidewire and Aristotle 18 Guidewire, respectively. The results from these tests are also applicable to CrossTorq 14 Guidewire, as the materials and processing of the CrossTorq 14 Guidewires are identical to those of the Aristotle 14 and Aristotle 18 Guidewires. | Summaries of Leveraged Functional Tests Conducted to Support this Premarket<br>Notification | | | |---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Test | Test Method Summary | Results | | Radiopacity | Guidewires evaluated by<br>physicians in human cadaver | Guidewires exhibited acceptable<br>radiopacity. | | Corrosion<br>Resistance | Test for corrosion resistance per<br>ISO 11070 | There were no signs of corrosion on<br>guidewires after soaking in typical<br>end-use solutions. | | Chemical<br>Compatibility | Guidewires were exposed to saline<br>and contrast agent/saline solutions<br>and examined for degradation. | All guidewires showed no signs of<br>degradation, corrosion or physical<br>decomposition after exposure. | | Latex | Tested for trace latex proteins per<br>ASTM D6499-07 | No detectable traces of latex were<br>found. | | Accessories<br>Testing | Compatibility of Torque Device<br>and Shaping Mandrel with<br>Aristotle Guidewires was<br>evaluated in cadaver testing and<br>simulated use evaluations.<br>Various tests on Shaping Mandrel<br>per ISO 11070: biocompatibility,<br>visual inspection, corrosion<br>resistance, tensile testing, luer<br>taper dimensions | Acceptance criteria of all tests were<br>met | | Package<br>Integrity | Simulated transportation test per<br>ASTM D 4169:16. Pouch<br>evaluated for seal strength per<br>ASTM F 88-15 and leak tests<br>(bubble test) per ASTM F 2096-11 | Following exposure to typical<br>storage and transportation<br>conditions, all sterile barrier<br>pouches maintained their integrity<br>and labeling remained affixed and<br>legible. | | Summaries of Leveraged Functional Tests Conducted to Support this Premarket<br>Notification | | | | Test | Test Method Summary | Results | | Sterilization<br>Validation | 100% EO is used to sterilize the<br>device to achieve a SAL of at least<br>10-6. The device was adopted into<br>an EO sterilization processing<br>group in accordance with AAMI<br>TIR 28:2009. Validation of the EO<br>sterilization cycle was performed<br>using the Half-cycle, overkill<br>approach described in Section<br>B.1.2 of ISO11135:2007 Annex B. | Results justified adoption into the<br>EO sterilization processing group:<br>Comparative and Bioburden<br>Resistance study results<br>demonstrated that PCDs are more<br>difficult to sterilize than devices;<br>Bioburden Enumeration and<br>Extraction Efficiency tests were<br>used to enumerate the CFUs present<br>on devices; and Bacteriostasis/<br>Fungistasis test results demonstrated<br>that the product does not inhibit the<br>growth of organisms. | | Sterilization<br>Validation:<br>EO and ECH<br>Residuals | Measured EO and ECH residuals<br>per AAMI/ANSI/ISO 10993-7 | The residual traces of EO and ECH<br>remaining in the CrossTorq 14<br>Guidewire after exposure to the EO<br>sterilization process are well below<br>the limits specified in ISO 10993-7. | | Sterilization<br>Validation:<br>Bacterial<br>Endotoxin<br>Levels | LAL testing was conducted in<br>accordance with ANSI/AAMI<br>ST72:2011/(R)2016, USP <161>,<br>and USP <85>, using the kinetic<br>chromogenic method. | < 2.15 EU/device | {10}------------------------------------------------ {11}------------------------------------------------ #### Functional Testing performed on CrossTorq 14 Guidewires Functional testing was performed in accordance with ISO 11070:2014 Sterile single-use intravascular introducers, dilators and guidewires and the FDA Guidance Document Coronary and Cerebrovascular Guidewire Guidance (January 1995). The following table summarizes the functional tests performed and test results obtained to demonstrate substantial equivalence to the predicate device. | Summaries of Functional Tests Conducted to Support this Premarket Notification | | | | | |--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Test | Test Method Summary | Results | | | | Visual<br>Inspection | Tests per ISO 11070:<br>Visual inspection for extraneous<br>matter, process and surface defects<br>or defects that may cause trauma<br>to vessels during use | No extraneous matter, surface<br>defects or visible droplets of coating<br>were present on the CrossTorq 14<br>Guidewires. | | | | Dimensional<br>Verification | Tests per ISO 11070:<br>Dimensional inspection per<br>engineering drawings | All guidewires met dimensional<br>specifications. | | | | Flexing Test | Tests per ISO 11070:<br>Inspection for defects and damage<br>or flaking of the coating after<br>flexing | No defects or damage / flaking of<br>the coating were observed after<br>flexing. | | | | Tensile<br>Strength | Tensile testing per ISO 11070 | All guidewires met minimum<br>clinically relevant force breakage<br>requirements when tested according<br>to ISO 11070. | | | | Tip Shape,<br>Retention | Guidewires must be shapeable and<br>must retain shaped angle after<br>simulated use | All tips met shaping and shape<br>retention requirements after<br>simulated use. | | | | Torqueability | Measurement of torque response<br>(average input to output lag) in an<br>anatomical model | All guidewires demonstrated<br>acceptable torque responses. The<br>torque response of the subject<br>device was comparable to that of the<br>predicate device. | | | | Torque<br>Strength | Torque turns to failure in an<br>anatomical model | All guidewires demonstrated<br>acceptable torque strength. The<br>torque strength of the subject device<br>was comparable to that of the<br>predicate device. | | | | Summaries of Functional Tests Conducted to Support this Premarket Notification | | | | | | Test | Test Method Summary | Results | | | | Tip<br>Flexibility | Measure force to deflect guidewire<br>tips to 45 and 90 degrees at 5mm,<br>10mm, and 20mm test lengths | The forces required to deflect the<br>guidewire tips of the softer<br>CrossTorq 14 tip profiles were<br>comparable to the forces required to<br>deflect the Kinetix, while the forces<br>required to deflect the tips of the<br>stiffer CrossTorq 14 tip profiles<br>were comparable to the forces<br>required for the Boston Scientific<br>Hornet 14 guidewire tips. | | | | Fracture | Tests per ISO 11070:<br>Inspection for fracture, loosening,<br>or failure after wrapping around<br>mandrel | No guidewires showed signs of<br>fracture, loosening, or failure after<br>wrapping them 8 times around a<br>mandrel. | | | | Coating<br>Lubricity and<br>Durability | Frictional force of coated<br>guidewires was determined after<br>simulated use in a tortuous path | All guidewires met specified<br>frictional force requirements. | | | | Coating<br>Integrity | Coating uniformity and integrity<br>visually examined on dyed<br>samples after simulated use in a<br>tortuous path | All samples showed acceptable<br>coating coverage after simulated<br>use. | | | | Particulates | Particulates of various size ranges<br>counted after simulated use in a<br>tortuous path | All guidewires met specified<br>number/size recovered particulate<br>requirements. | | | | Simulated<br>Use Model<br>Testing and<br>Product<br>Compatibility | Tortuous path model defined by<br>ASTM F2394 Figure X2.4 used<br>for simulated use testing<br>Compatibility of the accessories<br>with the CrossTorq 14 Guidewire<br>was evaluated in the simulated<br>tortuous path model evaluation. | Guidewires and predicate devices<br>were found to perform acceptably in<br>evaluations of: Torqueability in<br>tortuous vasculature, Lubricity,<br>Microcatheter and Balloon Catheter<br>Support & Tracking, Compatibility<br>with Introducer, Compatibility with<br>Torque Device, and Compatibility<br>with Microcatheter and Balloon<br>Catheter. | | | | Summaries of Functional Tests Conducted to Support this Premarket Notification | | | | | | Test | Test Method Summary | Results | | | | Column<br>Buckling | Measure force to buckle the<br>guidewire 1cm from the distal tip. | The forces required to buckle the<br>guidewire tips of the softer<br>CrossTorq 14 tip profiles were<br>comparable to the forces required to<br>deflect the Kinetix, while the forces<br>required to buckle the tips of the<br>stiffer CrossTorq 14 tip profiles<br>were comparable to the forces<br>required for the Boston Scientific<br>Hornet 14 guidewire tips. | | | | MRI<br>Compatibility | Guidewires are constructed of<br>metallic materials and should not<br>be exposed to MRI procedures. | No testing performed. CrossTorq 14<br>Guidewires are labeled "MRI<br>Unsafe." | | | | Shelf Life | Device performance attributes that<br>can be affected by storage<br>conditions were evaluated after<br>exposure to accelerated aging<br>conditions per ASTM F1980. | After exposure to accelerated aging<br>conditions simulating real-time<br>storage under ambient conditions<br>for 13 months, all device<br>performance acceptance criteria<br>were met, justifying labeling the<br>devices with a 1-year shelf life. | | | {12}------------------------------------------------ {13}------------------------------------------------ #### CONCLUSION Scientia Cardio Access, LLC has presented information in this premarket notification supporting its contention that the CrossTorq 14 Guidewire is substantially equivalent with respect to technological characteristics and indications for use to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...