ASAHI PTCA GUIDE WIRE, FIELDER J, MODEL AGP140000J, AGP140300J
Applicant
Asahi Intecc Co., Ltd.
Product Code
DQX · Cardiovascular
Decision Date
Nov 9, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The ASAHI PTCA Guide Wire, Fielder J is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The Asahi PTCA Guide Wires are not to be used in the cerebral blood vessel.
Device Story
Steerable guide wire (0.014 inch diameter; 180 cm or 300 cm length) used to facilitate placement of balloon dilatation catheters during PTCA and PTA procedures. Device features stainless steel core wire with soldered coil, radiopaque distal tip, and pre-shaped 'J' configuration for vessel navigation. Distal section includes polyurethane and hydrophilic coating; proximal section includes PTFE coating. Operated by physicians in clinical settings (e.g., cath lab) to navigate coronary or peripheral vasculature. Provides tactile feedback and trackability to assist in positioning interventional devices. Benefits include improved access to target lesions during angioplasty.
Clinical Evidence
Bench testing only. Mechanical and functional testing performed to demonstrate device performance and substantial equivalence to predicates.
Technological Characteristics
Stainless steel core wire; soldered coil; radiopaque distal tip; polyurethane and hydrophilic coating (distal); PTFE coating (proximal). 0.014 inch diameter. Manual operation; no energy source or software.
Indications for Use
Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA) or percutaneous transluminal angioplasty (PTA) to facilitate balloon catheter placement. Contraindicated for use in cerebral blood vessels.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K063819 — ASAHI PTCA GUIDE WIRE, FIELDER FC · Asahi Intecc Co., Ltd. · Jan 26, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
### SUMMARY OF SAFETY AND EFFECTIVENESS 9.0
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Asahi Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024<br>Japan | NOV - 9 2006 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| OFFICIAL<br>CORRESPONDENT | Yoshi Terai<br>President, CEO<br>Asahi Intecc USA, Inc.<br>1301 Dove Street, Suite 350<br>Newport Beach, CA 92660<br>Tel: (949) 756-8252<br>FAX (949) 756-8165<br>e-mail: yoshi@asahi-intecc.com | |
| TRADE NAME: | ASAHI PTCA Guide Wire, Fielder J | |
| COMMON NAME: | Guide Wire | |
| CLASSIFICATION<br>NAME: | Catheter Guide Wire | K062186 |
| DEVICE<br>CLASSIFICATION: | Class 2 per 21 CFR §870.1330 | |
| PRODUCT CODE | DQX | |
| PREDICATE DEVICE: | Asahi PTCA Guide Wire, Fielder K052022<br>Asahi PTCA Guide Wire, J Shape Series K043422<br>Guidant HI-Torque Floppy Guide Wire K974773<br>Guidant ACS HI-Torque Cross IT K990639 | |
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Asahi PTCA Guide Wire, Fielder J shape is steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 180 cm and 300 cm length. The extension wire is connected to the end of the guide wire outside the body for 180cm wire. The wire is constructed from a stainless steel core wire. The core wire and coil are soldered. The distal tip of the guide wire has a radiopaque tip to achieve visibility, and is available as a pre shape "J" and made flexible to bend easily at the vessel curve. There is polyurethane coating covered with hydrophilic coating applied to the distal section of the guide wire. The proximal section of this guide wire is coated with PTFE.
- 9.
{1}------------------------------------------------
# INDICATION FOR USE:
The ASAHI PTCA Guide Wire, Fielder J is intended to facilitate the placement of balloon The ROATH I TOA Oulde viro, Holdovinal coronary angioplasty (PTCA) and dilation catherers during poloatanover (PTA). The Asahi PTCA Guide Wires are not to be used in the cerebral blood vessel. 11/2017 11:12:14 PM 11/2 11/2 11:11 11/2 11/2 11:11 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11/2 11
# TECHNICAL CHARACTERISTICS:
The ASAHI PTCA Guide Wire, Fielder J is of the same materials as the predicate devices. The dimensional specifications and design of the device ensures compatibility for the intended use.
### Performance Data:
This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI ASAHI PTCA Guide Wire, Fielder J performs as intended.
# SUMMARY/CONCLUSION:
The ASAHI PTCA Guide Wire, Fielder J characteristics are substantially equivalent to the specified predicate devices and other currently marketed devices for the same indication for use.
Bench testing demonstrates that the device functions as intended.
Kc62186
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The caduceus is depicted in black, and the text is also in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV -- 9 2006
Asahi Intecc Co., Ltd. c/o Mr. Yoshi Terai President, CEO 1301 Dove Street, Suite 350 Newport Beach, CA 92660
Rc: K062186
ASAHI PTCA Guide Wire, Fielder J Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (Two) Product Code: DQX Dated: July 10, 2006 Received: July 31, 2006
Dear Mr. Terai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Yoshi Terai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Brimmer for
Brain D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT 2.0
510(k) Number (if known): 1002 Bb
Device Name: ASAHI PTCA Guide Wire, Fielder J
Indications for Use:
The ASAHI PTCA Guide Wire, Fielder J is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The Asahi PTCA Guide Wires are not to be used in the cerebral blood vessel.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blumming
of Sign-Off I wision of Cardiovaso 510(k) Number
Page of
000015
- 2.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.