K974713 · C.R. Bard, Inc. · DQX · Oct 14, 1998 · Cardiovascular
Device Facts
Record ID
K974713
Device Name
BARD HYDROPHILLIC COATED GUIDE WIRE
Applicant
C.R. Bard, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Oct 14, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Intended Use
The Bard Hydrophilic Coated Guide Wire is a guide wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel. They may be used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion. Hydrophilic Coated Angiography Guide Wires are indicated for use for percutaneous entry into a vessel using the Seldinger technique. Hydrophilic Coated Angioplasty (PTCA) Guide Wires are steerable guidewires that are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
Device Story
Bard Hydrophilic Coated Guide Wire; steerable wire designed for percutaneous catheter navigation. Used in coronary and peripheral vasculature to facilitate placement of diagnostic or interventional devices and cross target lesions. Operated by clinicians in clinical settings. Device functions as mechanical guide; hydrophilic coating reduces friction during vessel navigation. Substantially equivalent to existing Bard and Terumo guide wires.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance testing performed to demonstrate substantial equivalence to predicate devices.
Technological Characteristics
Hydrophilic coated guide wire; materials and construction similar to predicate Bard and Terumo guide wires. Mechanical device; no energy source or software. Sterilization method not specified.
Indications for Use
Indicated for percutaneous vessel entry via Seldinger technique (Angiography wires) and introduction/placement of diagnostic or interventional devices in coronary and peripheral vasculature to reach/cross target lesions (PTCA wires). No specific age/gender restrictions; no contraindications listed.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Bard® Preamendment Angiographic Guide Wires
Bard® PTCA Steerable Standard Guide Wire
Bard® Silk™ Guide Wire
Terumo Glidewire™
Related Devices
K993000 — BARD HYDROPHILIC COATED GUIDE WIRES · C.R. Bard, Inc. · Nov 2, 1999
K141295 — MERIT HYDROPHILIC GUIDEWIRE · Merit Medical Systems, Inc. · Dec 16, 2014
K140695 — COUGAR GUIDEWIRE, ZINGER GUIDWIRE, THUNDER GUIDEWIRE, PROVIA GUIDEWIRE, INTUITION GUIDEWIRE · Medtronic, Inc. · Jun 12, 2014
K964611 — TRANSEND GUIDEWIRE · Boston Scientific Corp · Jan 28, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date, "OCT 1 4 1998", in a simple, sans-serif font. The letters and numbers are printed in black ink against a white background. The date is formatted with the month abbreviated as "OCT", followed by the day "14", and the year "1998".
C.R. Bard, Inc. Requlatory Affairs 129 Concord Road P.O. Box 566 Billerica, MA 01821 508-667-2511 FAX: (508) 667-8594
K974713
Image /page/0/Picture/3 description: The image shows the letters "USCI" in bold, black font. The letters are enclosed in a thin, black oval. The letters are evenly spaced within the oval.
# SECTION 6.0 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
As required under Section 12, part (a)(I)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information respecting safety and effectiveness follows.
# General Information
| Submitter Information: | |
|------------------------------|-----------------------------------------------------------------------------------------|
| Name: | C.R. Bard, Inc. |
| Address: | 129 Concord Road, Billerica, MA 01821 |
| Phone: | (978) 667-2511 extension 4131 |
| Fax: | (978) 667-8594 |
| Contact: | |
| Fred Boucher | |
| Regulatory Affairs Manager | |
| Date of Summary: | |
| December 16, 1997 | |
| Name of Device: | |
| | Bard® Hydrophilic Coated Guide Wire |
| Common/Usual Name of Device: | |
| Catheter Guide Wire | |
| Device Classification: | |
| | 21 CFR 870.1330 |
| Predicate Device(s): | |
| | Bard® Preamendment Angiographic Guide Wires<br>Bard® PTCA Steerable Standard Guide Wire |
| | Bard® Silk™ Guide Wire |
| | Terumo Glidewire™ |
Image /page/0/Picture/8 description: The image shows the word "BARD" in a stylized, blocky font. The letters are thick and outlined in black, with the interior of the letters being white. The font style gives the word a bold and somewhat retro appearance.
{1}------------------------------------------------
## Description and Intended Use of Device: �
The Bard Hydrophilic Coated Guide Wire is a guide wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel. They may be used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
Hydrophilic Coated Angiography Guide Wires are indicated for use for percutaneous entry into a vessel using the Seldinger technique. Hydrophilic Coated Angioplasty (PTCA) Guide Wires are steerable guidewires that are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
#### Technological Characteristics Summary: �
The Bard Hydrophilic Coated Guide Wire is similar to the Terumo Glidewire regarding materials and construction, and is similar to Bard Silk Guide Wires, Bard Angiography Guide Wires and Bard PTCA Steerable Standard Guide Wire regarding materials and construction, packaging and sterilization. The indications for use are similar to both Terumo Glidewire and the Bard Guide Wires. They are all indicated for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion.
#### � Performance Data:
Safety and performance testing was performed to demonstrate that the Bard Hydrophilic Coated Guide Wire is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three heads. The eagle is facing to the right.
Public Health Service
OCT 1 4 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Fred L. Boucher Requlatory Affairs Manager Bard Cardiology Division C.R. Bard, Inc. 129 Concord Road P.O. Box 566 Billerica, MA 01821
Re: K974713 Bard® Hydrophilic Coated Guide Wires Trade Name: Requlatory Class: II Product Code: DQX Dated: July 16, 1998 Received: July 16, 1998
Dear Mr. Boucher:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this
{3}------------------------------------------------
Paqe 2 - Mr. Fred Boucher
response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use D.
Device Name: Hydrophilic Coated Guide Wire Indications for Use: Hydrophilic Coated Angiography Guide Wires are indicated for use for percutaneous entry into a vessel using the Seldinger technique. Hydrophilic Coated Angioplasty (PTCA) Guide Wires are steerable guidewires that are used for the introduction and placement of diagnostic or interventional devices in the coronary and peripheral vasculature and may be used to reach and cross a target lesion. None Contraindications:
Concurrence of CDRH, Office of Device Evaluation (ODE)
Tn A. D
(Division Sign-Off) Division of Cardiovascula and Neurological Devic 510(k) Number
Prescription Use X (Per 21 CFR 801.109)
Over-the Counter Use OR
1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.