B. BRAUN GUIDEWIRES
Device Facts
| Record ID | K961107 |
|---|---|
| Device Name | B. BRAUN GUIDEWIRES |
| Applicant | B.Braun Medical, Inc. |
| Product Code | DQX · Cardiovascular |
| Decision Date | Oct 1, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1330 |
| Device Class | Class 2 |
Intended Use
The Braun Guidewire is intended to fit inside a catheter for the purpose of directing the catheter through a blood vessel. These catheters may be used for PTCA and PTA procedures. These guidewires are not intended for use for cerebral, reproductive, or neurological procedures.
Device Story
Steerable, shapeable guidewire; used to direct catheters through blood vessels during PTCA/PTA procedures. Available in various diameters, stiffnesses, and lengths. Extendible versions feature proximal attachment for extension wire; allows intravascular device exchange while maintaining distal position. Extension wire is removable; restores original length/functionality. Used by clinicians in interventional settings.
Clinical Evidence
Bench testing only.
Technological Characteristics
Steerable, shapeable guidewire; available in multiple diameters, lengths, and stiffnesses. Features proximal attachment mechanism for extension wire compatibility. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients undergoing PTCA and PTA procedures requiring catheter navigation through blood vessels. Contraindicated for cerebral, reproductive, or neurological procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- ACS Guidewires (K852123, K852354, K854830, K861574, K861765, K862360, K864585, K872220, K874202, K874776, K881787, K881897, K884262, K884895, K902755, K905502, K913328, K913631, K915723, K925381, K934602, K942066, K950156)
- Target Guidewires (K884198, K922912, K914032)
- USCI Guidewires (K924009, K933466)
Related Devices
- K021228 — HI-TORQUE BALANCE, BALANCE MIDDLEWEIGHT, BALANCE HEAVYWEIGHT, BALANCE TREK AND THE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE · Guidant Corp. · May 15, 2002
- K173795 — HI-TORQUE TurnTrac Guide Wire Family · Abbott Vascular · Jan 12, 2018
- K250552 — Hi-Torque Command 14 ST Guide Wire and Hi-Torque Command 14 MT Guide Wire · ABBOTT MEDICAL · Jul 25, 2025
- K980360 — SCIMED QUEST FLOPPY AND MODERATE SUPPORT GUIDE WIRES · Scimed Life Systems, Inc. · Apr 29, 1998
- K101011 — HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL · Abbott Vascular, Inc. · May 24, 2010