ASAHI Fielder XT-A, ASAHI Fielder XT-R

K153106 · Asahi Intecc Co., Ltd. · DQX · Jun 21, 2016 · Cardiovascular

Device Facts

Record IDK153106
Device NameASAHI Fielder XT-A, ASAHI Fielder XT-R
ApplicantAsahi Intecc Co., Ltd.
Product CodeDQX · Cardiovascular
Decision DateJun 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Intended Use

ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature.

Device Story

ASAHI Fielder XT-A and XT-R are coronary/peripheral guide wires used to facilitate balloon dilation catheter placement during PTCA and PTA procedures. Device consists of a tapered stainless steel core wire, a tapered outer coil, and an inner rope coil, all soldered together. Distal coil and core wire are coated with polyurethane and a hydrophilic polymer; proximal core wire is PTFE-coated. Radiopaque coils allow for visualization under fluoroscopy. Used by physicians in clinical settings (e.g., cath lab) to navigate coronary or peripheral vasculature. Device provides mechanical support and guidance for interventional catheters, aiding in lesion access and treatment. Benefits include improved trackability and navigation through complex vascular anatomy.

Clinical Evidence

Bench testing only. Performance evaluated via dimensional verification, tensile strength, torque strength, torqueability, tip flexibility, coating adhesion, catheter compatibility, and coating integrity/particulate testing. Results met all acceptance criteria and demonstrated performance comparable to predicate devices.

Technological Characteristics

Materials: Stainless steel core, platinum-nickel/stainless steel coils, polyurethane undercoating, hydrophilic distal coating, PTFE proximal coating. Dimensions: 0.36mm nominal OD, 0.26mm distal diameter, 190cm/300cm lengths. Sensing/Actuation: Mechanical guide wire. Sterilization: Ethylene Oxide (SAL 10^-6).

Indications for Use

Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA) or percutaneous transluminal angioplasty (PTA) requiring balloon dilation catheter placement. Contraindicated for use in the neurovasculature.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure composed of three curved lines, representing the department's mission of protecting the health of all Americans and providing essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 21, 2016 Asahi Intecc Co., Ltd. % Candace Cederman Senior Regulatory Affairs Consultant CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228 US Re: K153106 Trade/Device Name: Asahi Fielder XT-A, ASAHI Fielder XT-R Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: May 9, 2016 Received: May 10, 2016 Dear Candace Cederman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements {1}------------------------------------------------ as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K153106 Device Name ASAHI Fielder XT-A, ASAHI Fielder XT-R #### Indications for Use (Describe) ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal any vnancon cance). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." EF {3}------------------------------------------------ # ASAHI INTECC CO.,LTD 1703 Wakita-cho, Moriyama-ku, Nagoya, Aichi 463-0024 Japan Tel. +81-52-768-1211 Fax. +81-52-768-1221 Branch offices: Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore, Beijing Research Facilities and Factories: Osaka, Seto, Thailand, Hanoi > 510(k) Summary [as required by 21 CFR 807.92(c)] #### ASAHI PTCA Guide Wires ASAHI Fielder XT-A ASAHI Fielder XT-R 510(k) | DATE PREPARED: | October 16, 2015 | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | APPLICANT | ASAHI Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024, Japan | | OFFICIAL<br>CORRESPONDENT | Yoshi Terai<br>President, CEO<br>ASAHI Intecc USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252<br>FAX: (949) 756-8165<br>e-mail: ASAHI.ra-fda@ASAHI-intecc.com | | TRADE NAME: | ASAHI PTCA Guide Wires:<br>• ASAHI Fielder XT-A<br>• ASAHI Fielder XT-R | | DEVICE<br>CLASSIFICATION: | Class 2 per 21 CFR §870.1330 | | CLASSIFICATION<br>NAME: | Catheter Guide Wire | | PRODUCT CODE | DQX - Wire, Guide, Catheter | | PREDICATE<br>DEVICES: | ASAHI PTCA Guide Wire ASAHI Gaia First (K133865)<br>ASAHI PTCA Guide Wire ASAHI Fielder FC (K063819, K072705)<br>ASAHI PTCA Guide Wire ASAHI Fielder XT (K072431) | #### Intended Use/Indications for Use ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature. {4}------------------------------------------------ #### DESCRIPTION: As a basic structure, the Fielder XT-A, Fielder XT-R, consists of a taper core wire and 2 types coils (a tapered outer coil and an inner rope coil). The taper core wire and the coils are soldered together. The coil is radiopaque so as to easily confirm its position under radioscopy. The coil and distal part of tapered core wire of the products are coated with polyurethane and then covered with hydrophilic polymer. The proximal end of the tapered core wire is coated with PTFE. The nominal outer diameter of both the Fielder XT-A and Fielder XT-R is 0.36mm. with a distal diameter of 0.26mm. Both the Fielder XT-A and Fielder XT-R are available in 190cm and 300cm, total length. The tip shape is straight. #### COMPARISON WITH PREDICATE DEVICES: Comparisons of the Fielder XT-A and Fielder XT-R and predicate devices show that the technological characteristics of the Subject devices such as the components, design, materials, sterilization method, shelf life and operating principle are identical or similar to the currently marketed predicate devices. The intended use/indications between the Subject Device and its primary predicates are identical. There are specific design features of the Subject device that are the same as the primary predicates. | Name of Device | ASAHI Fielder XT-A<br>ASAHI Fielder XT-R | ASAHI Gaia First | ASAHI Fielder FC<br>ASAHI Fielder XT | |---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------------------------------------------| | 510(k) | Current Application | K133865 | K063819, K072705<br>K072431 | | Intended Use and<br>Indications | ASAHI PTCA Guide Wires are intended to facilitate the placement of<br>balloon dilation catheters during percutaneous transluminal coronary<br>angioplasty (PTCA) and percutaneous transluminal angioplasty<br>(PTA).<br>The ASAHI PTCA Guide Wires are not to be used in the<br>neurovasculature. | | | | Sterilization | Provided sterile via Ethylene Oxide to SAL10-6 | | | | Shelf Life | 3 Years | | | | Target Body Location | Coronary and Peripheral | | | | Inner coil | Yes | | No | | Undercoating | Polyurethane | NA | Polyurethane | | Outer Distal coating | Hydrophilic | | | | Proximal Coating | PTFE | | | | Overall Length | 190, 300 cm | | 180, 300 cm (FC)<br>190, 300 cm (XT) | | Nominal OD | 0.36 mm (0.014 inches) | | | | Outer Coil Material | | Platinum-Nickel | Platinum-Nickel and<br>Stainless Steel (FC),<br>Platinum-Nickel (XT) | | Core Wire Material | Stainless Steel | | | ASAHI Intecc Co., Ltd. {5}------------------------------------------------ ### Non Clinical testing / Performance Data: Non clinical laboratory testing was performed on the ASAHI Fielder XT-A and ASAHI Fielder XT-R to determine substantial equivalence. The following testing/assessments were performed: - Dimensional Verification ● - Tensile Strenath ● - Torque Strength ● - Torqueability ● - . Tip Flexibility - Coating Adhesion - Catheter Compatibility ● - Coating Integrity / Particulate testing ● The in vitro bench tests demonstrated that the ASAHI Fielder XT-A and ASAHI Fielder XT-R met acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended, and has a safety and effectiveness profile that is similar to the predicate devices. #### BIOCOMPATIBILITY: The ASAHI Fielder XT-A and ASAHI Fielder XT-R were compared to the predicate devices. Based on similarities of the materials used in the subject device to its predicates, the biocompatibility of the ASAHI Fielder XT-A and ASAHI Fielder XT-R was verified to be the same as those of the predicates. #### Conclusion: The ASAHI Fielder XT-A and ASAHI Fielder XT-R have identical intended uses and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the devices function as intended. Therefore, the ASAHI Fielder XT-A and ASAHI Fielder XT-R are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...