ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature.
Device Story
ASAHI PTCA Guide Wire (Gladius Mongo 14/14 ES) is a steerable guidewire used to facilitate balloon dilatation catheter placement during PTCA and PTA procedures. Device consists of a one-piece stainless steel core wire and a distal coil assembly (inner/outer coils). Distal portion is radiopaque for fluoroscopic visualization. Surface features include a polyurethane undercoating, a hydrophilic outer distal coating, and a PTFE proximal coating. Used in coronary and peripheral vasculature by physicians in clinical settings. Device provides mechanical support and navigation for catheters; radiopacity allows real-time positioning. Benefits include improved access to target lesions during angioplasty.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via in vitro testing: tensile strength, torque strength, tip flexibility, coating adhesion, catheter compatibility, particulate testing, and lubricity. Biocompatibility verified via comparison to predicate/reference devices.
Technological Characteristics
Materials: Stainless steel core/inner coil, Platinum-Nickel/Stainless steel outer coil. Coatings: Polyurethane undercoating, hydrophilic distal coating, PTFE proximal coating. Dimensions: 0.014 inch nominal OD, 190cm/300cm lengths. Sterilization: Ethylene Oxide (SAL 10^-6). Mechanical device; no software or energy source.
Indications for Use
Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA) or percutaneous transluminal angioplasty (PTA) requiring balloon dilatation catheter placement. Contraindicated for use in the neurovasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 18, 2018
Asahi Intecc Co., Ltd. % Candace Cederman Principal Consultant CardioMed Device Consultants, LLC 3168 Braverton Street, Suite 200 Edgewater, Maryland 21037
Re: K180784
Trade/Device Name: ASAHI PTCA Guide Wire ASAHI Gladius Mongo 14, ASAHI Gladius Mongo 14 ES Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: June 20, 2018 Received: June 21, 2018
Dear Candace Cederman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
## Sincerely, Michael John -S 2018.07.18 15:29:55 -04'00'
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K180784
Device Name
ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14, ASAHI Gladius Mongo 14 ES
#### Indications for Use (Describe)
ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <div><span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ASAHI INTECC CO.,LTD.
3-100 Akatsuki-cho, Seto, Aichi 489-0071 Japan Tel. +81-561-48-5551 Fax. +81-561-48-5552 Branch o ffice : Tokyo, Nagoya, Osaka, Hong Kong, EU, S ingapore, Beijin g, Ind ia, Middle Eastern Research Facilitie s and Factorie s: Osaka, Seto, Thail and, Hanoi
#### 510(k) Summary [as required by 21 CFR 807.92(c)]
## ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14 and ASAHI Gladius Mongo 14 ES
## 510(k) 180784
| DATE PREPARED: | July 18, 2018 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | ASAHI INTECC CO., LTD.<br>3-100 Akatsuki-cho<br>Seto, Aichi 489-0071, Japan |
| CONTACT | Yoshi Terai<br>President/CEO<br>ASAHI INTECC USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252, FAX: (949) 756-8165<br>e-mail: ASAHI.ra-fda@ASAH-lintecc.com |
| TRADE NAME: | ASAHI® PTCA Guide Wire:<br>ASAHI Gladius Mongo 14 and ASAHI Gladius Mongo 14 ES |
| DEVICE CLASSIFICATION: | Class 2 per 21 CFR §870.1330 |
| CLASSIFICATION NAME: | Catheter, Guide, Wire |
| PRODUCT CODE | DQX- Catheter Guide Wire |
| PREDICATE DEVICES: | ASAHI® PTCA Guide Wire<br>• ASAHI® Fielder XT-A (K153106) |
| REFERENCE DEVICES: | ASAHI®Guide Wire Family (K163426)<br>ASAHI® PTCA Guide Wire<br>• ASAHI SUOH 03 (K162842)<br>• ASAHI RG3 (K141339)<br>• ASAHI Gaia (K133865)<br>• ASAHI SION (K122469)<br>• ASAHI ULTIMATEbros 3 (K101986)<br>• ASAHI Grand Slam (K043422)<br>ASAHI®Chikai Guide Wires (K141751)<br>ASAHI®Peripheral Guide Wire ASAHI Gladius<br>(K150445) |
## INTENDED USE/INDICATIONS FOR USE
ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14 (ES)
ASAHI PTCA Guide Wires are intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI PTCA Guide Wires are not to be used in the neurovasculature.
{4}------------------------------------------------
#### DEVICE DESCRIPTION:
The ASAHI® PTCA Guide Wire ASAHI Gladius Mongo consists of a one-piece core wire and a distal coil assembly. The coil assembly consists of an inner coil and an outer coil, soldered to the core wire. The distal portion of the coil is radiopaque so as to easily confirm its position under radioscopy. In addition, coatings are applied on the surface of the ASAHI® PTCA Guide Wire ASAHI Mongo. The coil and distal portion of the quide wire are coated with polyurethane and then covered with hydrophilic coating. The proximal portion of the guide wire is coated with PTFE only. The ASAHI® PTCA Guide Wire ASAHI Mongo is available in various lengths. The shorter length device is designed for use with a commercially available ASAHI INTECC extension, wire.
## COMPARISON WITH PREDICATE DEVICE:
Comparisons of the ASAHI® PTCA Guide Wire ASAHI Gladius Mongo and predicate devices show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life, and operating principle are identical or similar to the currently marketed predicate devices. The intended use of the subject device and its predicates are the same.
| Name of Device | ASAHI® PTCA Guide Wire ASAHI<br>Gladius Mongo 14 and ASAHI<br>Gladius Mongo 14 ES | ASAHI® PTCA Guide Wire<br>ASAHI Fielder XT-A |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| | Subject | Predicate |
| 510(k) | TBD | K153106 |
| Intended Use and<br>Indications | ASAHI PTCA Guide Wires are intended to facilitate the placement of<br>balloon dilatation catheters during percutaneous transluminal coronary<br>angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).<br>The ASAHI PTCA Guide Wires are not to be used in the<br>neurovasculature. | |
| Target Location | Coronary and Peripheral vasculature | |
| Nominal OD | 0.36mm (0.014 inch) | |
| Overall Lengths | 190cm and 300cm | |
| Outer Coil Material | ASAHI Gladius Mongo 14:<br>Platinum-Nickel, Stainless Steel<br>ASAHI Gladius Mongo 14 ES:<br>Platinum-Nickel | Platinum-Nickel |
| Inner Coil Material | Stainless Steel | |
| Core Wire Material | Stainless Steel | |
| Undercoating | Polyurethane | |
| Outer Distal Coating | Hydrophilic coating | |
| Proximal Coating | PTFE | |
| Sterilization | Provided sterile via Ethylene Oxide to SAL10-6 | |
| Shelf Life | 3 years | |
#### NON-CLINICAL TESTING/ PERFORMANCE DATA:
Non-clinical laboratory testing was performed on the ASAHI® PTCA Guide Wire ASAHI Gladius Mongo to to determine substantial equivalence. The following testing/assessments were performed:
- Tensile Strength
- . Torque Strength
{5}------------------------------------------------
- Torqueability
- . TipFlexibility
- . Coating Adhesion
- . Catheter Compatibility
- Particulate Testing/ Coating Integrity .
- . Lubricity Testing
The in vitro bench tests demonstrated that the ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14 and ASAHI Gladius Mongo 14 ES met all acceptance criteria and performed similarly to the predicate device. Performance data demonstrate that the devices functions as intended, and have a safety and effectiveness profile that is similar to the predicate devices.
## BIOCOMPATIBLIITY:
The ASAHI® PTCA Guide Wire ASAHI Gladius Mongo were compared to the predicate and reference devices. Based on similarities of the materials used in the subject device to its predicates / reference devices, the biocompatibility of the ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14 and ASAHI Gladius Mongo 14 ES were verified to be the same as those of the predicate / reference devices.
#### CONCLUSION:
The ASAHI® PTCA Guide Wire ASAHI Gladius Mongo have identical intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference devices. Performance data demonstrates that the device functions as intended.
Therefore, the ASAHI® PTCA Guide Wire ASAHI Gladius Mongo 14 and ASAHI Gladius Monqo 14 ES are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.