CINCH QR SGW EXTENSION

K963171 · Cordis Corp. · DQX · Nov 7, 1996 · Cardiovascular

Device Facts

Record IDK963171
Device NameCINCH QR SGW EXTENSION
ApplicantCordis Corp.
Product CodeDQX · Cardiovascular
Decision DateNov 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Intended Use

The Cordis CINCH® QR Steerable Guidewire Extension is designed to extend Cordis Steerable Guidewires so that a catheter can be exchanged for another catheter.

Device Story

Flexible stainless steel wire; 145 cm length; 0.014 inch shaft diameter; DURAGLIDE™ coated. Used as accessory to Cordis steerable guidewires; facilitates catheter exchange during interventional procedures. Operated by physicians in clinical settings. Device extends existing guidewire length to allow safe removal and replacement of catheters over the wire.

Clinical Evidence

Bench testing only; no clinical data provided. Equivalence based on material and design similarity to predicate.

Technological Characteristics

Stainless steel wire; DURAGLIDE™ coating; 145 cm length; 0.014 inch diameter. Mechanical device; no energy source or software.

Indications for Use

Indicated for use as an accessory to Cordis Steerable Guidewires to facilitate catheter exchange in patients requiring guidewire extension.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} NOV - 7 1996 K963171 # PREMARKET NOTIFICATION 510(k) ## Cordis Corporation ### Cinch® QR Steerable Guidewire Extension ## SUMMARY OF SAFETY AND EFFECTIVENESS ### I. General Provisions - Common or Usual Name: Catheter Guide Wire (Accessory) - Proprietary Name: Cordis CINCH® QR Steerable Guidewire Extension ### II. Name of Predicate Devices - CINCH® 3 Steerable Guidewire Extension ### III. Classification - Catheter Guide Wire Accessory, Class II - 21 CFR 870.1330 ### IV. Performance Standards Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act. ### V. Intended Use and Device Description The Cordis CINCH® QR Steerable Guidewire Extension is designed to extend Cordis Steerable Guidewires so that a catheter can be exchanged for another catheter. The CINCH® QR Steerable Guidewire Extension is a flexible stainless steel wire coated with DURAGLIDE™. The extension is 145 cm long and has a shaft diameter of 0.014 inches. ### VI. Biocompatibility Biocompatibility testing was not repeated since the CINCH® QR and the currently marketed CINCH® 3 (K914863) Steerable Guidewire Extensions are made of the same materials. The DURAGLIDE™ coating and method of application have also been cleared for use with the WIZDOM™ Steerable Guidewire (K953760).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...