The Vigil™ HbA1c Controls are designed for monitoring the overall reliability of HbA1c assays on SYNCHRON CX® and LX™ Systems.
Device Story
Vigil HbA1c Controls consist of two-level lyophilized control sets derived from hemolyzed human and ovine blood; intended for use in clinical laboratories to monitor performance of Beckman SYNCHRON CX and LX HbA1c test systems. Product requires reconstitution with distilled water; stored at 2°C to 8°C. Laboratorians use controls to assess analytical error and imprecision; ensuring reliability of patient test results. Device is manufactured by Boehringer Mannheim (marketed as Precinorm/Precipath HbA1c) and labeled for Beckman Coulter, which assigns system-specific values. Stability is 24 months when stored as specified.
Clinical Evidence
Bench testing only. Stability studies performed at 25°C, 32°C, and 37°C over 65, 30, and 18 days respectively support a 24-month expiration claim.
Technological Characteristics
Lyophilized control set; materials: hemolyzed human and ovine blood; form factor: 1 mL bottles; storage: 2°C to 8°C; reconstitution: 1.0 mL distilled water; classification: Quality Control Material (21 CFR 862.1660).
Indications for Use
Indicated for monitoring the reliability and overall performance of Beckman Hemoglobin A1c (HbA1c) test systems in the clinical laboratory. Used by laboratorians to monitor analytical error and imprecision via two levels of control.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
K987072
Beckman Coulter, Inc., Section 510(k) Notification VIGIL™ HbA1c Controls Summary of Safety & Effectiveness
JUL 6 1998
Summary of Safety & Effectiveness Beckman VIGIL™ HbA1c Controls
# 1.0 Submitted By:
Lucinda Stockert Staff Regulatory Specialist, Product Submissions Beckman Coulter, Inc. 200 S. Kraemer Blvd., W-104 Brea, California 92822-8000 Telephone: (714) 961-3777 FAX: (714) 961-4123
# Date Submitted: 2.0
8 June 1998
# Device Name(s): 3.0
# 3.1 Proprietary Names
VIGIL™ Hemoglobin A1c Controls
# Classification Name 3.2
Quality Control Material (assayed and unassayed) (21 CFR § 862.1660)
# Predicate Device(s): 4.0
| BECKMAN Reagent | Predicate | Predicate Company | Docket Number |
|-------------------------------------|------------------------------------------------------------|------------------------|---------------|
| Vigil TM HbA1c Controls | Boehringer Mannheim<br>Tina-quant®<br>Hemoglobin A1c assay | Boehringer<br>Mannheim | K934070 |
{1}------------------------------------------------
Beckman Coulter, Inc., Section 510(k) Notification VIGIL™ HbA1c Controls Summary of Safety & Effectiveness
#### 5.0 Description:
"The Vigil HbA1c Controls are a two level lyophilized control set made from hemolyzed human and ovine blood. Each kit contains 3 X 1 mL bottles of each specific level of control. The levels are identified as Vigil HbA1c Control Level 1, and Vigil HbA1c Control Level 2. The products require reconstitution with exactly 1.0 mL of distilled water prior to use. The lyophilized products must be stored at +2℃ to +8℃ when not in use. The reconstituted controls are stable for 1 week at +2℃ to +8℃, 8 hours at +20℃ to +25℃, and 3 months when stored frozen between -15℃ to -20℃.
#### 6.0 · Intended Use:
The Vigil™ HbA1c Controls are designed for monitoring the overall reliability of HbA1c assays on SYNCHRON CX® and LX™ Systems.
### Comparison to Predicate(s): 7.0
Boehringer Mannheim Corporation manufactures the Vigil™ HbA1c controls for Beckman Coulter, Inc. Boehringer Mannheim has clearance to market these products under the names Precinorm® HbA1c and Precipath® HbA1c. Beckman Coulter, Inc. labels are placed on the product and Beckman assigns SYNCHRON® Systems values to the product.
#### 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence of the Vigil HbA1c Controls to the Boehringer Mannheim Precinorm® and Precipath®. Stress stability studies of the lyophilized Vigil HbA1c Controls support the Beckman stability claim of 24 months.
| Stress Temperature | Duration of Incubation | Beckman Stability Claim* |
|--------------------|------------------------|--------------------------|
| 25°C | 65 Days | 24 Months |
| 32°C | 30 Days | 24 Months |
| 37°C | 18 Days | 24 Months |
# Vigil HbA1c Controls Stability Study Summary
*Expiration dating placed on the package based on date of manufacture
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular fashion around the bird symbol. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
6 1998 . IUL
Lucinda Stockert Staff Requlatory Specialist Beckman Coulter, Inc. 200 S. Kraemer Boulevard, M/S W-104 P.O. Box 800 Brea, California 92822-8000
K982022 Re : VIGIL™ HbAlc Controls Requlatory Class: I Product Code: JJX Dated: June 8, 1998 Received: June 9, 1998
Dear Ms. Stockert:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "dsmo@fdadr.cdrh.fda.gov".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
Attachment 2
Image /page/4/Picture/1 description: The image shows the words "page" and "of" in a simple, sans-serif font. There are two horizontal lines underneath each word, suggesting that there are supposed to be numbers written in the blanks. The words are likely part of a page numbering system, indicating the current page and the total number of pages. The text is black against a white background.
510(k) Number (if known): Not yet assigned
Device Name: VIGIL™ HbA1c Controls
Indications for Use:
The Beckman Vigil Hemoglobin A1c (HbA1c) Controls are designed for monitoring the reliability and overall performance of Beckman Hemoglobin A1c (HbA1c) test systems in the clinical laboratory. The use of two levels of control allows the laboratorian to monitor analytical error and imprecision.
21 CFR 862.1660 Quality Control Material (assayed and unassayed)
(a) Identification. A quality control material (assayed and unassayed) is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
(b) Classification. Class I.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | 1982022 |
Prescription Use
(per 21 CFR 801.109)
OR
Over-the-Counter Use
Optional Format 1-2-96
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.