LIQUICHEK ELEVATED CRP CONTROL

K042837 · Bio-Rad · JJX · Nov 2, 2004 · Clinical Chemistry

Device Facts

Record IDK042837
Device NameLIQUICHEK ELEVATED CRP CONTROL
ApplicantBio-Rad
Product CodeJJX · Clinical Chemistry
Decision DateNov 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Liquichek Elevated CRP Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for C-Reactive Protein (CRP).

Device Story

Liquichek Elevated CRP Control is a liquid, human serum-based quality control material used in clinical laboratories. It contains C-Reactive Protein (CRP) along with stabilizers and preservatives. The device is used by laboratory personnel to monitor the precision of analytical testing procedures for CRP. By measuring the control alongside patient samples, clinicians can verify the accuracy and reliability of their testing equipment and methods, ensuring consistent results for patient diagnosis and monitoring.

Clinical Evidence

No clinical data. Stability studies were performed to validate open vial stability (30 days at 2°C to 8°C) and shelf life (3 years at -10°C to -70°C).

Technological Characteristics

Human serum-based quality control material. Ready-to-use liquid form. Analyte: C-Reactive Protein (CRP). Compatible with specific clinical chemistry analyzers (Beckman Coulter, Dade BEHRING, Roche). Stability: 3 years closed-vial at -10 to -70°C; 30 days open-vial at 2 to 8°C. No electronic or software components.

Indications for Use

Indicated for use as a quality control serum to monitor the precision of laboratory testing procedures for C-Reactive Protein (CRP). Prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k042837 B. Purpose for Submission: Premarket Notification 510(k) of intention to manufacture and market the Bio-Rad Laboratories, Liquichek Elevated C-Reactive Protein (CRP) Control C. Measurand: Quality Control Material (assayed) for C-Reactive Protein (CRP) D. Type of Test: Not applicable E. Applicant: Bio-Rad Laboratories F. Proprietary and Established Names: Liquicheck™ Elevated CRP Control G. Regulatory Information: 1. Regulation section: 21 CFR §862.1660; Quality Control Material (assayed and unassayed) 2. Classification: Class I 3. Product code: JJX; Single (specified) analyte controls (assayed and unassayed) {1} 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): Liquichek Elevated CRP Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for C-Reactive Protein (CRP). 2. Indication(s) for use: Liquichek Elevated CRP Control is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for C-Reactive Protein (CRP). 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Beckman Coulter Array Beckman Coulter IMMAGE Dade BEHRING BN Series Dade BEHRING Dimension Series Roche Cobas INTEGRA Roche Hitachi I. Device Description: Liquichek™ Elevated CRP Control contains three levels of quality control material. The control materials are human serum-based and are provided ready-to-use. Value assignment and stability testing information are below. Human source material was tested and found negative for HIV 1 and 2, HBV and HCV using FDA approved methods. J. Substantial Equivalence Information: 1. Predicate device name(s): Liquichek™ Lipids Control {2} 2. Predicate 510(k) number(s): k012513 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Quality Control Material | Quality Control Material | | Form | Liquid | Liquid | | Matrix | Human serum based | Human serum based | | Differences | | | | Tem | Device | Predicate | | Storage (Unopened) | -10 to -70°C Until expiration date | -10 to -20°C Until expiration date | | Open Vial Claim | 30days at 2 to 8°C | 14 days at 2 to 8°C | | Analytes | C-Reactive Protein (CRP) | CRP,Apolipoprotein A-1, Apolipoprotein B, Cholesterol,Cholesterol HDL, Cholesterol LDL, Lipoprotein(a), triglycerides | K. Standard/Guidance Document Referenced (if applicable): None stated L. Test Principle: Not applicable. This submission is for clearance of control material. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable {3} c. Traceability, Stability, Expected values (controls, calibrators, or methods): C-Reactive Protein (CRP) is a purchased product which is prepared into a stock solution. Several dilutions are made from the stock solution at specific concentrations within the CRP assay range. The dilutions are assayed using a commercially available assay and percent recovery is calculated for the stock solution, verifying the raw materials activity. CRP is then added to human serum based matrix at the appropriate target values. The material is analyzed by the reagent manufacturer and/or independent laboratories using manufacturer supported reagents. Refer to package insert for the assigned values and ranges. Stability: Real time accelerated stability studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The stability is listed below: - Open vial stability is 30 days at 2 to 8°C - Closed vial stability is 3 years at -10 to -70°C d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: {4} Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...