LIAISON N-TACT PTH CALIBRATION VERIFIERS MODEL 310913

K093498 · DiaSorin, Inc. · JJX · Dec 14, 2009 · Clinical Chemistry

Device Facts

Record IDK093498
Device NameLIAISON N-TACT PTH CALIBRATION VERIFIERS MODEL 310913
ApplicantDiaSorin, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateDec 14, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

The DiaSorin LIAISON® N-TACT® PTH Calibration Verifiers are assayed quality control materials intended for use in the quantitative verification of calibration and reportable range of the LIAISON® N-TACT® PTH Assay when performed on the LIAISON® Analyzer.

Device Story

LIAISON® N-TACT® PTH Calibration Verifiers are quality control materials used to verify calibration and reportable range of LIAISON® N-TACT® PTH Assay. Product consists of four levels of lyophilized pooled human plasma spiked with PTH and preservatives. Used in clinical laboratory settings on LIAISON Analyzer. Healthcare providers use output to verify assay performance and ensure accuracy of patient PTH measurements. Benefits include improved confidence in clinical decision-making regarding parathyroid hormone levels.

Clinical Evidence

No clinical data. Bench testing only. Precision/reproducibility studies conducted over 20 days and 5 days using multiple LIAISON analyzers and assay lots. Results demonstrated acceptable SD and %CV across four concentration levels (17.4 to 1303.5 pg/mL). Stability studies confirmed shelf-life, open-vial, and freeze-thaw performance.

Technological Characteristics

Lyophilized pooled human plasma spiked with PTH; contains stabilizers and 0.2% Proclin® 300. Four levels (20, 150, 350, 1500 pg/mL). Reconstituted with 2.0 mL deionized/distilled water. Analyzed via LIAISON® Analyzer. Storage: 2-8°C (pre-reconstitution), -20°C (post-reconstitution).

Indications for Use

Indicated for use as assayed quality control materials to verify calibration and reportable range of the LIAISON® N-TACT® PTH Assay on the LIAISON® Analyzer. No specific patient population, age, or gender is defined as this is an in vitro diagnostic control material.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k093498 B. Purpose for Submission: New device C. Measurand: Calibration verification materials for the LIASION® N-TACT® PTH Assay D. Type of Test: Quality control materials E. Applicant: DiaSorin Inc. F. Proprietary and Established Names: LIAISON® N-TACT® PTH Calibration Verifiers G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJX | Class I, reserved | 862.1660 | Chemistry 75 | H. Intended Use: 1. Intended use(s): Refer to indications for use below 2. Indication(s) for use: {1} The DiaSorin LIAISON® N-TACT® Calibration Verifiers are assayed quality control materials intended for use in the quantitative verification of calibration and reportable range of the LIAISON® N-TACT® PTH Assay when performed on the LIAISON Analyzer. 3. Special conditions for use statement(s): For Prescription Use 4. Special instrument requirements: DiaSorin LIASION® Analyzer. Included in cleared assays: k033426, k081685, k082050. I. Device Description: The DiaSorin LIAISON® N-TACT® PTH calibration Verifiers consist of four levels (20, 150, 350 and 1500 pg/mL). Each vial contains lyophilized pooled human plasma with preservatives, stabilizer and spiked with PTH. All human source materials were tested and found to be non-reactive for HIV ½, HBsAg, and HCV by FDA- approved methods. J. Substantial Equivalence Information: 1. Predicate device name(s): LIAISON® N-TACT® PTH Control Set 2. Predicate 510(k) number(s) k033426 3. Comparison with Predicate | Similarities | | | | --- | --- | --- | | Item | Candidate Device | Predicate Device | | Intended Use | Assayed quality control samples for use in the quantitative verification of calibration and reportable range of the LIASION® N-TACT® PTH Assay | Same | | Analyte | Parathyroid Hormone | Same | | Matrix | Pooled human plasma with stabilizers and Proclin® 300 | Same | {2} | Format | Lyophilized | Same | | --- | --- | --- | | Product Storage | 2- 8°C before reconstitution -20°C after reconstitution | Same | | Volume | 2.0 mL after reconstitution | Same | | Required Reagent | LIAISON® N-TACT® PTH Assay | Same | | Differences | | | | --- | --- | --- | | Item | Candidate Device | Predicate Device | | Levels | Four | Two | # K. Standard/Guidance Document Referenced (if applicable): CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods CLSI EP15-A2, User Verification of Performance for Precision and Trueness Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material # L. Test Principle: Not applicable # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: A twenty day reproducibility/precision study was performed. One lot of LIAISON® N-TACT® PTH Controls (KC 1 and KC 2) and one lot of Calibration Verifiers A-D were tested in two replicates per day, for 20 operating days on two LIAISON® analyzers for a total of 80 replicate results per sample. One kit lot of LIAISON® N-TACT® PTH Assay was used throughout the study. The CLSI document EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods, was consulted for this study design. A five day reproducibility/precision study was conducted to verify the results of the twenty day study. The CLSI document EP15-A2, User Verification of Performance for Precision and Trueness, was consulted in the preparation of the testing protocol. This study included one lot of LIAISON® N-TACT® PTH Assay and LIAISON® N-TACT® PTH Control Set, and 3 lots of LIAISON® N-TACT® PTH Calibration {3} Verifiers. Testing was performed on three LIAISON® analyzers; one lot of calibration verifiers per LIAISON® analyzer. The calibration verifier samples were analyzed daily in quadruplicate, the assay controls were analyzed daily in duplicate for five days. The results of the five day precision study were provided and found to be sufficient. Reproducibility/Precision Results – 20 day | Sample ID | N | Mean conc. pg/mL | within run | | between run (day) | | between instrument | | overall | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | KC 1 | 80 | 47.1 | 1.29 | 2.8 | 2.87 | 6.1 | 0.29 | 0.6 | 3.15 | 6.7 | | KC 2 | 80 | 491.1 | 4.61 | 0.9 | 17.06 | 3.5 | 3.73 | 0.8 | 17.85 | 3.6 | | cal ver A | 80 | 17.4 | 0.77 | 4.5 | 1.42 | 8.3 | 0.29 | 1.7 | 1.86 | 10.7 | | cal ver B | 80 | 137.0 | 2.26 | 1.7 | 6.79 | 5.0 | 1.43 | 1.0 | 7.17 | 5.2 | | cal ver C | 80 | 310.8 | 5.25 | 1.7 | 15.27 | 4.9 | 1.93 | 0.6 | 15.94 | 5.1 | | cal ver D | 80 | 1303.5 | 22.96 | 1.8 | 85.36 | 6.6 | 1.84 | 0.1 | 92.50 | 7.1 | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability The LIAISON® N-TACT® PTH Calibration Verifier concentrations are referenced to standard in-house materials containing a pure preparation of synthetic human PTH. The Calibration Verifiers are composed of the reference material added to a human plasma-based matrix. Value-Assignment A minimum of 10 vials of each level of calibration verifier are used in the final range assignment. The LIAISON® N-TACT® PTH Calibration Verifiers are tested on multiple LIAISON® Analyzers, using multiple LIAISON® N-TACT® PTH assay kit lots, on multiple assay runs. A range of two standard deviations is assigned for each level of calibration verifier. Stability Stability characteristics of the LIAISON® N-TACT® PTH Calibration Verifiers were determined using real time stability, open vial, and freeze thaw stability studies. An unopened shelf-life of 3 months is expected at the recommended storage temperature of 2-8°C. Open vial and freeze thaw studies determined that LIAISON® N-TACT® {4} PTH Calibration Verifiers are stable through 1 freeze/thaw cycle at -20°C and for at least 8 hours at room temperature (18° to 22°C). d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The expected mean and ranges are provided in the labeling. The sponsor recommends in {5} the labeling that each laboratory should determine the ranges based on their own test system and tolerance limits. **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...