MAS AND MINI-TROL PTH CONTROL

K092749 · Microgenics Corporation · JJX · Dec 3, 2009 · Clinical Chemistry

Device Facts

Record IDK092749
Device NameMAS AND MINI-TROL PTH CONTROL
ApplicantMicrogenics Corporation
Product CodeJJX · Clinical Chemistry
Decision DateDec 3, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

MAS® PTH Control is intended for use in the clinical laboratory as a consistent test sample of known concentration for monitoring assay conditions in many Parathyroid hormone (PTH) determinations. Include PTH Control with patient serum specimens when assaying for Parathyroid hormones. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument. Moni-Trol® PTH Control is intended for use in the clinical laboratory as a consistent test sample of known concentration for monitoring assay conditions in many Parathyroid hormone (PTH) determinations. Include PTH Control with patient serum specimens when assaying for Parathyroid hormones. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument.

Device Story

MAS® PTH Control and Moni-Trol® PTH Control are liquid stable, human serum-based quality control materials; used in clinical laboratories to monitor assay conditions for Parathyroid hormone (PTH) testing. Controls are provided in three levels, ready-to-use. Laboratory personnel include the control with patient serum specimens during PTH assays. Users compare observed values against provided expected ranges to verify reagent and instrument performance. Human-derived components are tested for HBsAg, HCV, HIV-1, and HIV-2. Product integrity is maintained via added preservatives and stabilizers.

Clinical Evidence

No clinical data. Bench testing only. Stability studies (open vial, closed vial, and shelf-life) were conducted using Roche COBAS PTH reagents on the Roche Elecsys 2010 analyzer across three lots. Shelf-life established via Arrhenius model using stress data. Value assignment performed via replicate assays across 4 instruments.

Technological Characteristics

Liquid stable control material; human serum matrix; additives include chemicals, stabilizers, and preservatives. Storage: -25°C to -15°C (closed), 2-8°C (open). Analyte: PTH.

Indications for Use

Indicated for use in clinical laboratories as a quality control material for monitoring assay performance in Parathyroid hormone (PTH) determinations. No specific patient population or contraindications described.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k092749 B. Purpose for Submission: New Device C. Measurand: Quality control material for parathyroid hormone (PTH) determinations D. Type of Test: Not applicable E. Applicant: Microgenics Corporation, Thermo Fisher Scientific Specialty Diagnostics Division F. Proprietary and Established Names: Liquid Assayed PTH Control G. Regulatory Information: 1. Regulation section: 21 CFR§862.1660 2. Classification: Class I (Reserved) 3. Product code: JJX - Quality control material (assayed and unassayed) 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): MAS® PTH Control is intended for use in the clinical laboratory as a consistent test sample of known concentration for monitoring assay conditions in many Parathyroid hormone (PTH) determinations. Include PTH Control with patient serum specimens when assaying for Parathyroid hormones. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument. Moni-Trol® PTH Control is intended for use in the clinical laboratory as a consistent test sample of known concentration for monitoring assay conditions in many Parathyroid hormone (PTH) determinations. Include PTH Control with patient serum specimens when assaying for Parathyroid hormones. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument. 2. Indication(s) for use: Same as intended use above. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: {1} Roche Cobas 6000; Roche Elecsys; Roche Modular E170; Siemens Immulite I. Device Description: The MAS® PTH Control and Moni-Trol® PTH Control are liquid stable control materials prepared from human serum. Components of the controls which are derived from human source material have been tested using FDA accepted methods and found non-reactive for Hepatitis B Surface antigen (HBsAg), Hepatitis C (HCV), HIV-1 and HIV-2. Analyte levels are adjusted with various pure chemicals and preparations from human tissue or bodily fluids. Preservatives and stabilizers are added to maintain product integrity. Three levels of control are available and are provided ready to use. Each kit will contain 6 vials each: 2 vials of Level 1, 2 vials of Level 2, and 2 vials of Level 3. Each vial contains 3 mLs. The MAS and Moni-Trol products are the same. J. Substantial Equivalence Information: 1. Predicate device name(s): Liquichek™ Specialty Immunoassay Control 2. Predicate 510(k) number(s): k04310 3. Comparison with predicate: Similarities: | | Predicate Device, K043108 Bio-Rad Liquichek™ Controls | Proposed new device K092749 MAS® PTH Control and Moni-Trol® PTH Control | | --- | --- | --- | | Intended Use | For use as quality control serum to monitor the precision of laboratory testing procedures listed in the package insert. | MAS® PTH and Moni-Trol® PTH Control are intended for use in the clinical laboratory as a consistent test sample of known concentration for monitoring assay conditions in many Parathyroid hormone (PTH) determinations. Include PTH Control with patient serum specimens when assaying for Parathyroid hormones. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument. | {2} 3 | Description of device | This product is prepared from human serum with added constituents, chemicals, stabilizers, and preservatives. | This product is a liquid stable control material prepared from human serum. Analyte levels are adjusted with various pure chemicals and preparations from human tissue or bodily fluids. Preservatives and stabilizers are added to maintain product integrity. | | --- | --- | --- | | Matrix | Human Serum | Same | | Form | Liquid | Same | | Open Vial Stability | 30 days at 2-8°C with exceptions | 30 days when stored tightly capped at 2-8°C | | Analytes | Anti-Tg Anti-TPO C-Peptide Erythropoietin Intact PTH IGF-I Osteocalcin 25-OH Vitamin D | PTH | Differences: | | Predicate Device, K043108 Bio-Rad Liquichek™ Controls | Proposed new device, K092749 MAS® PTH Control and Moni-Trol® PTH Control | | --- | --- | --- | | Storage Condition | -20°C to -70°C until expiration date on the label | -25°C to -15°C until expiration date on the label | | Closed Vial Stability | 30 days at 2-8°C with exceptions | 90 days when stored tightly capped at 2-8°C | | Levels and targets | Level 1: ~30 pg/mL Level 2: ~200 pg/mL Level 3: ~600 pg/mL (estimated from value assignment data in BioRad PI) | Level 1: 15 pg/mL Level 2: 200 pg/mL Level 3: 400 pg/mL | K. Standard/Guidance Document Referenced (if applicable): None were referenced. L. Test Principle: Not applicable. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable {3} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The stability studies were performed using the Roche COBAS PTH reagents on the Roche Elecsys 2010 analyser. All studies that were performed were done so using three lots of the control material. Open vial stability: Three lots of PTH control, each lot at three different levels were used. For each lot and level vials were opened on day 0, stored at 5°C, and tested at different time points for up to 42 days. The recommended claim based on the data is 30 days at 2 - 8°C. Closed vial stability: Three lots of PTH control, each lot at three different levels were used. For each lot and level, unopened vials were stored at 5°C, and tested at multiple time points for up to 90 days. The recommended claim based on the data is 90 days at 2 - 8°C. Shelf-Life stability: The Arrhenius model was applied using stress data to establish the frozen - 20°C shelf life claim. Three lots, each lot at three different levels were used for the study. For each lot and level, product was stressed at elevated temperatures of 41°C and 37°C for 14 days, and 25°C for 28 days. Multiple time points were tested and point of failure was determined by linear regression. The recommended claim based on the data is 24 months frozen at -20°C. Value assignment: Value assignment ranges are established 20% around the mean. Control ranges are provided as guidelines until the laboratory has established its own statistical limits. The values are based on replicate assays of representative samples by participating laboratories in accordance with established protocol. Twenty replicates per level per analyte over a period of 10 days were run on 4 instruments. Means established by individual laboratories should fall within the corresponding acceptable range. It is recommended that each laboratory establish its own means and acceptable ranges and use those provided only as guides. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable {4} b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...