ELECSYS CALCHECK LH
Device Facts
| Record ID | K964830 |
|---|---|
| Device Name | ELECSYS CALCHECK LH |
| Applicant | Boehringer Mannheim Corp. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Dec 16, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Indications for Use
The Boehringer Mannheim Elecsys CalCheck LH is used to verify the calibration assignment for the Boehringer Mannheim Elecsys LH assay.
Device Story
Elecsys CalCheck LH serves as calibration verification material for Elecsys LH assay; manufactured using human serum albumin, luteinizing hormone, stabilizers, and preservatives. Analyte spiked into matrix to specific concentration levels. Quality controlled against Enzymun LH assay kit calibrators, value-assigned via comparison to 2nd International Standard (NIBSC) 80/552. Used by laboratory personnel to confirm assay calibration accuracy; ensures reliable patient test results.
Clinical Evidence
Bench testing only; evaluation focused on value assignment and stability of the CalCheck material.
Technological Characteristics
Calibration verification material; composition includes human serum albumin, luteinizing hormone, stabilizers, and preservatives. Value assignment traceable to NIBSC 80/552 standard.
Indications for Use
Indicated for use in verifying calibration assignment for the Boehringer Mannheim Elecsys LH assay in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Tosoh Medic AIA Pack for FSH (K924863)
Related Devices
- K121279 — ELECSYS LH CALCHECK 5 · Roche Diagnostics · May 10, 2012
- K964829 — ELECSYS CALCHECK FSH · Boehringer Mannheim Corp. · Dec 16, 1996
- K031299 — ELECSYS LH CALSET II · Roche Diagnostics Corp. · May 12, 2003
- K970147 — ELECSYS CALCHECK PROLACTIN · Boehringer Mannheim Corp. · Feb 10, 1997
- K163521 — Lumipulse G LH Calibrators · Fujirebio Diagnostics,Inc. · Jan 12, 2017