The Liquid Stable HbA1c Control is intended for use as a quality control material to monitor the performance of laboratory testing procedures for HbA1c quantitation. The control is designed for use with lon exchange HPLC assays, Immunoassay based assays and Boronate Affinity based assays.
Device Story
Liquid Stable HbA1c Control is a bi-level, ready-to-use quality control material derived from human red cells; used in clinical laboratories to monitor precision and bias of HbA1c testing procedures. Compatible with ion exchange HPLC, Immunoassay, and Boronate Affinity assays (e.g., Bayer DCA 2000, Bio-Rad Variant II, Roche Hitachi 917, Primus CLC 385). Controls provide objective assessment of method performance within clinical range. Human source material tested negative for HIV, HBV, HCV, and Syphilis. Healthcare providers use control results to verify assay accuracy and reliability as part of standard laboratory practice.
Clinical Evidence
No clinical data. Bench testing only. Stability studies confirmed open vial stability of 30 days and closed vial stability of 18 months at 2-8°C. Expected ranges for normal and abnormal levels were established via duplicate testing across three methodologies.
Technological Characteristics
Bi-level liquid control; matrix consists of human hemoglobins, broad-spectrum antibiotic, and stabilizers. Ready-to-use. Storage at 2-8°C. Compatible with ion exchange HPLC, Immunoassay, and Boronate Affinity platforms.
Indications for Use
Indicated for use as a quality control material to monitor the precision and bias of laboratory HbA1c quantitation procedures. Intended for in vitro diagnostic use by laboratory professionals to monitor performance within the clinical range. No specific patient population or contraindications listed.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Quantimetrix Corporation, GlycoHemosure HbA1c Control (k032791)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k043070
B. Purpose for Submission:
Clearance of New device
C. Measurand:
Quality Control Material (assayed) for Hemoglobin A1c (HbA1c)
D. Type of Test:
Not applicable
E. Applicant:
Canterbury Scientific Ltd
F. Proprietary and Established Names:
Liquid Stable HbA1c Control
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1660; Quality Control Material (assayed and unassayed)
2. Classification:
Class I
3. Product code:
JJX; single (specified) analyte controls (assayed and unassayed)
4. Panel:
75; Chemistry
{1}
H. Intended Use:
1. Intended use(s):
See indications for use.
2. Indication(s) for use:
The Liquid Stable HbA1c Control is intended for use as a quality control material to monitor the performance of laboratory testing procedures for HbA1c quantitation. The control is designed for use with ion exchange HPLC assays, Immunoassay based assays and Boronate Affinity based assays.
The use of quality control materials is indicated as an objective assessment of the precision and bias of methods and techniques in use and is an integral part of good laboratory practices. The two levels of controls allow performance monitoring within the clinical range.
The controls are for in vitro diagnostic use only and should not be used past the expiry date.
3. Special conditions for use statement(s):
For Prescription use only
The control is designed for use with ion exchange HPLC assays, Immunoassay based assays and Boronate Affinity based assays.
4. Special instrument requirements:
Bayer DCA 2000, Bio-Rad Variant II, Roche Hitachi 917 and Primus CLC 385
I. Device Description:
The Liquid Stable HbA1c Control is a bi-level liquid HbA1c control. The control materials are made from human red cells and are provided ready-to-use. Value assignment and stability testing information is below.
Human source material was tested and found negative for HIV 1 and 2, HIV-1 antigen, HBV, HCV and Syphilis (TPHA and RPR) using FDA approved methods.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Quantimetrix Corporation, GlycoHemosure HbA1c Control
{2}
2. Predicate 510(k) number(s):
k032791
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Quality Control Material | Quality Control Material |
| Form | Liquid ready to use (2 levels) | Liquid ready to use (2 levels) |
| Matrix | Human hemoglobins, broad spectrum antibiotic, stabilizers | Human hemoglobins, preservatives, stabilizers |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Volume | 0.5 mL – 1.0 mL | 2.0 mL |
| Instrumentation | Ion Exchange HPLC, Immunoassay and Boronate Affinity | Immunoassay and Boronate Affinity |
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Not applicable. This submission is for clearance of control material.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
{3}
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The Liquid Stable HbA1c Controls are prepared from normal non-diabetic adult human blood. A hemolysate is prepared from washed red cells. The abnormal HbA1c level control is prepared by in vitro glycation of non-diabetic hemolysate under controlled conditions.
Each lot is assayed by randomly selecting ten normal and abnormal liquid control samples. These samples are run in duplicate within one day on three different methodologies. The abnormal level expected range is derived from the mean of all the values ± 11.0%. The normal level expected range is derived from the mean of all the values ± 5.0%.
Stability:
Real time stabilities studies have been conducted. Protocols and acceptance criteria were described and found to be acceptable. The stability is listed below:
Open vial stability is 30 days at 2 to 8°C.
Closed vial stability is 18 months at 2 to 8°C.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
{4}
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.