IDS-iSYS 17-OH Progesterone Control Set, IDS-iSYS 17-OH Progesterone Calibration Verifiers
Device Facts
| Record ID | K161082 |
|---|---|
| Device Name | IDS-iSYS 17-OH Progesterone Control Set, IDS-iSYS 17-OH Progesterone Calibration Verifiers |
| Applicant | Immunodiagnostic Systems , Ltd. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | May 17, 2016 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Indications for Use
The IDS-iSYS 17-OH Progesterone Control Set is for in vitro diagnostic use, for the quality control of the IDS-iSYS 17-OH Progesterone on the IDS-iSYS Multi-Discipline Automated System. Rx Only. The IDS-iSYS 17-OH Progesterone Calibration Verifiers are an in vitro diagnostic device intended for medical purposes in the quantitative verification of assay calibration and measuring range of the IDS-iSYS 17-OH Progesterone assay, when performed on the IDS-iSYS Multi Discipline Automated System. Rx Only.
Device Story
IDS-iSYS 17-OH Progesterone Control Set and Calibration Verifiers are liquid human serum-based reagents containing 17-OH Progesterone and sodium azide preservative. Used on IDS-iSYS Multi-Discipline Automated System; operated by laboratory personnel. Controls monitor assay performance; calibration verifiers confirm assay calibration and measuring range. Value assignment performed gravimetrically and confirmed via immunologic analysis on the IDS-iSYS system. Healthcare providers use results to ensure analytical accuracy of patient sample testing. Benefits include reliable quality assurance and verification of assay measuring range, supporting accurate clinical diagnostic results.
Clinical Evidence
Bench testing only. Value assignment performed on three IDS-iSYS systems over 21 runs (controls) or five runs (verifiers). Stability validated via accelerated studies (CLSI EP25-A) supporting 9-month shelf life at 2-8°C; real-time studies ongoing. Open-vial stability confirmed for 49 days; on-board stability confirmed for 4 hours. Precision acceptance criteria: CV ≤ 10% for low concentration, ≤ 8% for middle/high concentrations.
Technological Characteristics
Liquid human serum matrix with <0.1% sodium azide preservative. Analyte: 17-OH Progesterone. System: IDS-iSYS Multi-Discipline Automated System. Storage: 2-8°C. Stability testing per CLSI EP25-A. No electronic or software components integral to the reagent itself; functions as a consumable for the automated analyzer.
Indications for Use
Indicated for in vitro diagnostic quality control and calibration verification of the IDS-iSYS 17-OH Progesterone assay on the IDS-iSYS Multi-Discipline Automated System.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
- K990393 — ARCHITECT PROGESTERONE MASTERCHECK MODEL #6C26-05 · Bio-Rad · Mar 22, 1999
- K143144 — ST AIA-PACK PROG II Calibrator Set · Tosoh Bioscience, Inc. · Dec 3, 2014
- K150955 — Elecsys Progesterone III Cal Check 5 · Roche Diagnostics · May 1, 2015
- K103683 — IMMULITE/IMMULITE 1000 PROGESTERONE CALIBRATION VERIFICATION (CVM) · Siemens Healthcare Diagnostics · Apr 28, 2011