OXICOM CONTROL

K053327 · Bionostics, Inc. · JJX · Jan 26, 2006 · Clinical Chemistry

Device Facts

Record IDK053327
Device NameOXICOM CONTROL
ApplicantBionostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateJan 26, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Oxicom Control is intended to be used to monitor and evaluate the analytical performance of the Waters Oxicom models 2000, 2100 and 3000 oxygen saturation meters for the measurement of oxygen saturation. The use of quality control materials is indicated as an objective assessment of the precision of methods and techniques in use and is an integral part of good laboratory practice. The three levels of oxygen saturation provided by the controls allow performance monitoring within the clinically important range.

Device Story

Oxicom Control is a liquid solution containing dyes and polystyrene beads; designed to mimic absorbance of whole blood at specific wavelengths used by Waters Oxicom oxygen saturation meters (models 2000, 2100, 3000). Used as an in vitro diagnostic quality control material; handled like patient samples within clinical laboratory workflows. Provides three levels of oxygen saturation to assess instrument precision and analytical performance. Healthcare providers use the output to verify that the analyzer is functioning correctly, ensuring reliable oxygen saturation measurements for patient care.

Clinical Evidence

No clinical data. Bench testing only: closed bottle stability, stability after opening, correlation to predicate devices, and test precision/range verification.

Technological Characteristics

Liquid solution containing suspended styrene beads and dyes. Designed to simulate whole blood absorbance at specific wavelengths for Waters Oxicom systems. In vitro diagnostic quality control material. No electronic components, software, or energy source.

Indications for Use

Indicated for use as a quality control material to monitor and evaluate the analytical performance of Waters Oxicom model 2000, 2100, and 3000 oxygen saturation meters in a laboratory setting.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k053327 B. Purpose for Submission: Clearance to market quality control material for oxygen saturation meters. C. Measurand: The product under submission serves as a control for a device that measures oxygen saturation in whole blood. D. Type of Test: Quality Control E. Applicant: Bionostics Inc. F. Proprietary and Established Names: Oxicom Controls G. Regulatory Information: 1. Regulation section: 21 CFR 862.1660: Quality control material (assayed and unassayed). 2. Classification: Class I (reserved) 3. Product code: JJX {1} 4. Panel: (75) Chemistry H. Intended Use: 1. Intended use(s): Please see indications for use. 2. Indication(s) for use: Oxicom Control is intended to be used to monitor and evaluate the analytical performance of the Waters Oxicom model 2000, 2100 and 3000 oxygen saturation meters for the measurement of oxygen saturation. The use of quality control materials is indicated as an objective assessment of the precision of methods and techniques in use and is an integral part of good laboratory practice. The three levels of oxygen saturation provided by the controls allow performance monitoring with the clinically important range. For In Vitro diagnostic Use. 3. Special conditions for use statement(s): For Prescription Use Only. 4. Special instrument requirements: Waters Oxicom oxygen saturation meters: models 2000, 2100, 3000. I. Device Description: The Oxicom Control is a liquid solution with dyes and polystyrene beads used to mimic the absorbance of whole blood at a range of oxygen saturation levels. The concentration of beads and dyes are designed to simulate the behavior of whole blood at the wavelengths used by Waters Oxicom Systems. These liquid controls can be handled as normal patient samples and can be incorporated into the clinical workflow. J. Substantial Equivalence Information: 1. Predicate device name(s): OPTI-Check Quality Control Oxicom 2100 QC Filters 2. Predicate 510(k) number(s): 2 {2} K974822: OPTI-Check Quality Control K921519: Oxicom 2100 QC Filters 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate(K974822) | | | | | | Intended Use | As a quality control material in the measurement of oxygen saturation | Same | | Number of levels | 3 | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate(K974822) | | Material | Aqueous solution containing suspended styrene beads and dyes | Aqueous solution containing suspended styrene beads. | | Supported Platform | Waters Oxicom systems | AVL OPTI | | Additional analytes | None | pH, PCO2, PO2, Na,K, Cl, iCa, tHb | K. Standard/Guidance Document Referenced (if applicable): ISO 14971:2000: Medical devices - Application of risk management to medical devices. ISO 15223: Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied. FDA Guidance: "Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use" Available at: http://www.fda.gov/cdrh/ocd/guidance/4444.pdf FDA Guidance: "Points to consider guidance document on assayed and unassayed quality control material". {3} L. Test Principle: The product under submission is used to verify the performance of Oxicom analyzers. The control is treated as a typical sample. Processing is done per instrument instructions. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The performance of this control was verified by comparing it to the existing Oxicom 2100 QC Filters used for quality control on the required Waters Oxicom Systems instruments. Samples were run once a day for 30 days. The results are presented in the table below. | | Oxicom Controls | | | | --- | --- | --- | --- | | Oxicom Model | High Control | Mid Control | Low Control | | 2000 n | 30 | 30 | 30 | | Mean | 83.4 | 72.7 | 48.7 | | SD | 0.40 | 0.69 | 1.39 | | CV% | 0.5% | 1.0% | 2.9% | | 2100 n | 30 | 30 | 30 | | Mean | 86.7 | 75.9 | 51.4 | | SD | 0.30 | 0.67 | 1.27 | | CV% | 0.3% | 0.9% | 2.5% | | 3000 n | 30 | 30 | 30 | | Mean | 83.0 | 72.0 | 47.9 | | SD | 0.55 | 0.69 | 1.03 | | CV% | 0.7% | 1.0% | 2.1% | b. Linearity/assay reportable range: Not applicable for a device of this type. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Value assignment is determined based upon replicate assays of representative samples of controls on multiple instruments and lots of measurement kits. {4} The proposed device will serve as a quality control material for Waters Oxicom oxygen saturation meters. The stability of the device is one of the primary criteria for acceptance. The company assessed the opened shelf life of the device through real-time studies designed to mimic the behavior of device in the hands of the users. The company evaluated the performance of open but capped material by comparing measurements on this opened material to material that had not been unsealed. Both the opened and unopened material were measured on 5 Oxicom instruments. Measurements were made weekly. All material demonstrated significantly less that a 10% change in behavior, the performance criteria set for the product. The data supplied by the company supports the 2 month opened shelf life claimed by the company. The company assessed the shelf life of the product through a series of accelerated aging studies. The data supplied by the company supports the 3 year shelf life claimed by the company. d. Detection limit: Not applicable for a device of this type. e. Analytical specificity: Not applicable for a device of this type. f. Assay cut-off: Not applicable for a device of this type. {5} 2. Comparison studies: a. Method comparison with predicate device: The company demonstrated their equivalence by direct comparison to the existing Oxicom 2100 QC Filters on the required Waters Oxicom Systems instruments.. Measurement comparisons were made across all 3 model lines and on multiple instruments within in each product line. Samples were run once a day for 30 days. The results are presented in the table below. | | Oxicom Controls | | | Existing QC Method | | | | --- | --- | --- | --- | --- | --- | --- | | Oxicom Model | High Control | Mid Control | Low Control | QC 1 | QC 2 | QC 3 | | 2000 n Mean SD CV% | 5 | 5 | 5 | 5 | 5 | 5 | | | 85.1 | 72.6 | 52.3 | 83.9 | 91.3 | 65.4 | | | 0.57 | 0.75 | 1.71 | 0.90 | 2.39 | 7.02 | | | 0.7% | 1.0% | 3.3% | 1.1% | 2.6% | 10.7% | | 2100 n Mean SD CV% | 22 | 22 | 22 | 22 | 22 | 22 | | | 84.9 | 72.4 | 54.1 | 77.3 | 85.7 | 69.6 | | | 0.81 | 1.05 | 2.78 | 10.31 | 2.74 | 11.41 | | | 1.0% | 1.5% | 5.1% | 13.3% | 3.2% | 16.4% | | 3000 n Mean SD CV% | 14 | 14 | 14 | 14 | 14 | 14 | | | 85.1 | 73.0 | 54.0 | 82.1 | 89.5 | 61.4 | | | 0.70 | 0.96 | 2.38 | 4.77 | 2.84 | 8.65 | | | 0.8% | 1.3% | 4.4% | 5.8% | 3.2% | 14.1% | b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable for a device of this type. b. Clinical specificity: Not applicable for a device of this type. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. {6} 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Oxygen saturation in blood, expressed as a percentage, can range from 0% - 100%. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...