ELECSYS CALCHECK FT3
Device Facts
| Record ID | K963163 |
|---|---|
| Device Name | ELECSYS CALCHECK FT3 |
| Applicant | Boehringer Mannheim Corp. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Aug 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Indications for Use
Elecsys CalCheck FT 3 is intended for use in the quantitative verification of the calibration curve established by the Elecsys FT 3 reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.
Device Story
Elecsys CalCheck FT 3 consists of three-level, lyophilized, human-based materials; used for calibration verification of FT 3 assays on Elecsys 1010/2010 immunoassay analyzers. User reconstitutes material, performs triplicate assay, and compares results against target values to verify calibration curve integrity. Used in clinical laboratory settings by laboratory personnel. Output provides verification of analytical system performance; assists in ensuring accurate patient test results.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Lyophilized, human-based material; three-level set; single analyte; intended for use on Elecsys 1010/2010 immunoassay analyzers.
Indications for Use
Indicated for quantitative verification of calibration curves for FT 3 reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Tosoh Medic AIA-Pack HCG Calibration Verification Test Set (K924862)
Related Devices
- K963167 — ELECYS CALCHECK T 3 · Boehringer Mannheim Corp. · Aug 27, 1996
- K963164 — ELECSYS CALCHECK FT4 · Boehringer Mannheim Corp. · Aug 27, 1996
- K963147 — ELECSYS CALCHECK TSH · Boehringer Mannheim Corp. · Aug 23, 1996
- K963178 — ELECSYS CALCHECK T4 · Boehringer Mannheim Corp. · Aug 27, 1996
- K963181 — ELECSYS CALCHECK T UPDATE · Boehringer Mannheim Corp. · Sep 10, 1996