ELECSYS CALCHECK T UPDATE

K963181 · Boehringer Mannheim Corp. · JJX · Sep 10, 1996 · Clinical Chemistry

Device Facts

Record IDK963181
Device NameELECSYS CALCHECK T UPDATE
ApplicantBoehringer Mannheim Corp.
Product CodeJJX · Clinical Chemistry
Decision DateSep 10, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

Elecsys CalCheck T Uptake is intended for use in the quantitative verification of the calibration curve established by the Elecsys T Uptake reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.

Device Story

Elecsys CalCheck T Uptake; three-level single-analyte set; lyophilized human-based material. Used for calibration verification of Elecsys T Uptake reagents/calibrators on Elecsys 1010/2010 immunoassay analyzers. User reconstitutes material; assays in triplicate; compares results to target values. Verifies calibration curve accuracy on automated immunoassay systems; supports clinical decision-making by ensuring analytical performance of T Uptake testing.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Three-level lyophilized human-based material; single analyte set; designed for use on Elecsys 1010/2010 immunoassay analyzers.

Indications for Use

Indicated for quantitative verification of calibration curves for T Uptake assays on Elecsys 1010 or 2010 immunoassay analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SEP 10 1996 K963181 # 510(k) Summary ## Elecsys® CalCheck™ T Uptake | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | --- | --- | | 1) Submitter name, address, contact | Boehringer Mannheim Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 845-2000 Contact Person: John D. Stevens Date Prepared: August 12, 1996 | | 2) Device name | Proprietary name: Elecsys CalCheck T Uptake Common name: Calibration Verification Material Classification name: Single (specified) analyte controls (assayed + unassayed) | | 3) Predicate device | We claim substantial equivalence to the Tosoh Medic AIA-Pack HCG Calibration Verification Test Set, K924862. The intended use for both products is to verify the calibration curve of automated immunoassay analyzers | | 4) Device Description | The Elecsys CalCheck T Uptake is a three level single analyte set of lyophilized, human based materials. After reconstitution they are assayed in triplicate and the results are compared to the target values. | | 5) Intended use | Elecsys CalCheck T Uptake is intended for use in the quantitative verification of the calibration curve established by the Elecsys T Uptake reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers. | {1} K963181 # 510(k) Summary, Elecsys® CalCheck™ T Uptake, Continued ## 6) Comparison to predicate device The Boehringer Mannheim Elecsys CalCheck T Uptake is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Tosoh Medics AIA-Pack HCG Calibration Verification Test Set, K924862. The intended use of both the Elecsys CalCheck T Uptake and the predicate device are the same in that they are intended to be used for the calibration verification for their labeled analytes on automated immunoassay analyzers. 12
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...