IMMULITE® 2000 Calcitonin Calibration Verification Material , IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material
K143373 · Siemens Healthcare Diagnostics, Inc. · JJX · Feb 26, 2015 · Clinical Chemistry
Device Facts
Record ID
K143373
Device Name
IMMULITE® 2000 Calcitonin Calibration Verification Material , IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JJX · Clinical Chemistry
Decision Date
Feb 26, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The IMMULITE® 2000 Calcitonin Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Calcitonin assay on the IMMULITE 2000 systems. The IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE PAP assay on the IMMULITE 2000 systems.
Device Story
IMMULITE 2000 CVMs are lyophilized, multi-level quality control materials used to verify calibration of specific assays on IMMULITE 2000 systems. Calcitonin CVM contains bovine protein buffer matrix; PAP CVM contains human PAP in bovine protein buffer matrix. Used by laboratory technicians in clinical settings. Operators reconstitute lyophilized material, run on IMMULITE 2000, and compare recovered values against assigned target ranges. Output allows verification that the instrument's calibration remains accurate within established clinical limits. Benefits include ensuring assay reliability and diagnostic accuracy for patient testing.
Clinical Evidence
No clinical data. Bench testing only. Value assignment performed on IMMULITE 2000 analyzers using 27 replicates for Calcitonin and 15 replicates for PAP across multiple systems and reagent lots. Stability studies (real-time) support 5-year shelf life at -20°C and 8-hour open-vial stability at 15-25°C, with recovery within 90-110%.
Technological Characteristics
Lyophilized, multi-level (4 levels) calibration verification materials. Matrix: bovine protein buffer with preservatives (sodium azide in PAP CVM). Storage: -20°C (Calcitonin) or 2-8°C (PAP). Connectivity: used on IMMULITE 2000 systems. Standards: CEN 13640 for stability testing.
Indications for Use
Indicated for in vitro diagnostic use to verify calibration of IMMULITE Calcitonin and PAP assays on IMMULITE 2000 systems. Prescription use only.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
IMMULITE® 2000 Intact PTH Calibration Verification Material (k140258)
IMMULITE® 2000 SHGB Calibration Verification Material (k140541)
Related Devices
K132739 — IMMULITE 2000 AFP, BR-MA, AND OM-MA CALIBRATION VERIFICATION MATERIAL · Siemens Healthcare Diagnostics, Inc. · Jan 14, 2014
K132391 — IMMULITE 2000 ANTI-TG AB CALIBRATION VERIFICATION MATERIAL, IMMULITE 2000 ANTI-TPO AB CALIBRATION VERIFICATIN MATERIAL · Siemens Helathcare Diagnostics, Inc. · Sep 20, 2013
K140818 — IMMULITE: 2000 FREE T4 CALIBRATION VERIFICATION MATERIAL, 2000 IGF-I CALIBRATION VERIFICATION MATERIAL, 2000 PROLACTIN C · Siemens Healthcare Diagnostics, Inc. · May 1, 2014
K142878 — IMMULITE 2000 Beta-2 Microglobulin Calibration Verification Material, IMMULITE 2000 High Sensitivity CRP Calibration Verification Material · Siemens Healthcare Diagnostics, Inc. · Oct 30, 2014
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k143373
B. Purpose for Submission:
New device
C. Measurand:
Calibration Verification Materials for Calcitonin and Prostatic Acid Phosphatase (PAP)
D. Type of Test:
Not applicable
E. Applicant:
Siemens Healthcare Diagnostics, Inc.
F. Proprietary and Established Names:
IMMULITE® 2000 Calcitonin Calibration Verification Material
IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material
G. Regulatory Information:
1. Regulation section:
862.1660, Quality control material
2. Classification:
Class I, reserved
3. Product code:
JJX
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use statements below.
2. Indication(s) for use:
IMMULITE® 2000 Calcitonin Calibration Verification Material:
The IMMULITE® 2000 Calcitonin Calibration Verification Material (CVM) is for in
{1}
vitro diagnostic use in the verification of calibration of the IMMULITE Calcitonin assay on the IMMULITE 2000 systems.
IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material: The IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE PAP assay on the IMMULITE 2000 systems.
3. Special conditions for use statement(s): For prescription use only
4. Special instrument requirements: IMMULITE® 2000
I. Device Description:
The IMMULITE® 2000 Calcitonin Calibration Verification Material (CVM) contains one set of four vials, 3mL each after reconstitution. CVM1 contains bovine protein buffer matrix with preservatives and CVM2, CVM3, and CVM4 contain calcitonin in bovine protein buffer matrix with preservative.
The IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material (CVM) contains one set of four vials, 2mL each after reconstitution. CVM1 contains a bovine protein/buffer with 0.27% sodium azide and preservative. CVM2, CVM3, and CVM4 contain human prostatic acid phosphatase in a bovine protein/buffer matrix with 0.27% sodium azide and preservative.
Sponsor has the following caution statement in their labeling: "Contains human source material. Each donation of human blood or blood component was tested by FDA-approved methods for the presence of antibodies to human immunodeficiency virus type 1 (HIV-1) and type 2 (HIV-2) as well as for hepatitis B surface antigen (HBsAg) and antibody to hepatitis C virus (HCV). The test results were negative (not repeatedly reactive.).
J. Substantial Equivalence Information:
1. Predicate device name(s): IMMULITE® 2000 Intact PTH Calibration Verification Material (CVM) IMMULITE® 2000 SHGB Calibration Verification Material (CVM)
2. Predicate 510(k) number(s): k140258 k140541
3. Comparison with predicate:
{2}
IMMULITE® 2000 Calcitonin Calibration Verification Material:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: IMMULITE® 2000 Calcitonin Calibration Verification Material: | Predicate: IMMULITE® 2000 Intact PTH Calibration Verification Material (k140258) |
| Intended Use | Calibration Verification Material (CVM) is for the verification of calibration of the chemistry assay. | Same |
| Storage | ≤-20°C | Same |
| Form | Lyophilized | Same |
| Levels | 4 | Same |
| Matrix | Bovine protein buffer with preservatives | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device (k140258) |
| Analyte | Calcitonin | PTH |
IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: IMMULITE® 2000 Prostatic Acid Phosphatase (PAP) Calibration Verification Material: | Predicate: IMMULITE® 2000 SHGB Calibration Verification Material (k140541) |
| Intended Use | Calibration Verification Material (CVM) is for the verification of calibration of the chemistry assay. | Same |
| Storage | 2 – 8 °C | Same |
| Form | Lyophilized | Same |
| Levels | 4 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device (k140541) |
| Analyte | Prostatic Acid Phosphatase | SHBG |
| Matrix | Bovine protein/buffer matrix with 0.27% Sodium Azide and preservatives | Buffered bovine/protein with preservatives |
{3}
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Value assignment and stability studies were performed on the IMMULITE 2000 analyzer.
Value assignment:
Value Assignment for the IMMULITE 2000 Calcitonin CVM was based on using assigned reference calibrators. The assigned reference calibrators were prepared using Calcitonin antigen stock traceable to WHO 2nd IRP 89/620. The CVM was tested on 27 replicates (in total comprised of 9 runs, 3 replicates per run) on 8 IMMULITE 2000 systems, using 3 reagent kit lots. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The target values are provided in the IMMULITE® 2000 CVM Calibration verification Material lot- specific value card. The expected assay range is 2 to 2000 pg/mL. The target values in table below can be considered as guidelines.
| CVM Level | Target Mean (ng/mL) | Standard Deviation (SD) | Guideline ±2SD Range (ng/mL) | |
| --- | --- | --- | --- | --- |
| CVM1 | 0.00 | - | 0.00 | 0.02 |
| CVM2 | 18.9 | 2.95 | 13 | 24.8 |
| CVM3 | 316 | 16 | 284 | 348 |
| CVM4 | 2082 | 125 | 1832 | 2332 |
{4}
Value Assignment for the IMMULITE 2000 Prostatic Acid Phosphatase (PAP) CVM was based on using PAP antigen stock traceable to an internal standard which was gravimetrically prepared. The CVM was tested on 15 replicates (in total comprised of 5 runs, 3 replicates per run) on 4 IMMULITE 2000 systems, using 3 reagent kit lots. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD), except CVM4 needs to be diluted with CVM1 before assaying. The expected assay range is 0.2 to 100 mg/L. The target values in table below can be considered as guidelines.
| CVM Level | Target Mean (ng/mL) | Standard Deviation (SD) | Guideline ±2SD Range (ng/mL) | |
| --- | --- | --- | --- | --- |
| CVM1 | 0.00 | - | 0.00 | 0.05 |
| CVM2 | 1.05 | 0.105 | 0.84 | 1.26 |
| CVM3 | 5.40 | 0.54 | 4.32 | 6.48 |
| CVM4 | 142 | - | - | - |
| 75% CVM4 + 25% CVM1 | 107 | 12.84 | 81.3 | 133 |
Stability:
Shelf-Life and Open Vial Stability testing protocols and acceptance criteria for the IMMULITE 2000 Calcitonin CVM were described and found to be adequate. The real time stability study shows study results up to 5 years when stored at -20°C, supporting the claim of 5 years shelf life and up to 9 hours at ambient or room temperature (15-25°C) supporting the claim of 8 hours after reconstitution. The study results were within 90% to 110% recovery.
Shelf-Life and Open Vial Stability testing protocols and acceptance criteria for the IMMULITE 2000 Prostatic Acid Phosphatase (PAP) CVM were described and found to be adequate. The real time stability study shows study results up to 5 years when stored at -20°C, supporting the claim of 5 years shelf life and up to 9 hours at ambient or room temperature (15-25°C) supporting the claim of 8 hours after reconstitution. The study results were within 90% to 110% recovery
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
{5}
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.