CONGRUENT BONE PLATE SYSTEM
K012655 · Acu Med, Inc. · HRS · Nov 7, 2001 · Orthopedic
Device Facts
| Record ID | K012655 |
| Device Name | CONGRUENT BONE PLATE SYSTEM |
| Applicant | Acu Med, Inc. |
| Product Code | HRS · Orthopedic |
| Decision Date | Nov 7, 2001 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Acumed Congruent Bone Plate System provides fixation for fractures, fusions, or osteotomies for the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
Device Story
The Acumed Congruent Bone Plate System consists of titanium bone plates and screws used for internal fixation of fractures, fusions, and osteotomies. Plates are pre-bent to reduce intraoperative bending requirements. The system includes specialized instrumentation to assist in plate and screw insertion. The device is intended for use by surgeons in a clinical/surgical setting. The system provides mechanical stabilization of bone segments to facilitate healing. Implants are provided non-sterile and require steam sterilization by the user prior to surgery.
Clinical Evidence
Bench testing only. No clinical data provided. Sterilization validation data confirms that instrumentation and implants can be successfully steam sterilized to a sterility assurance level (SAL) of 10^-6.
Technological Characteristics
System consists of titanium bone plates and screws. Provided non-sterile; validated for steam sterilization. Includes insertion instrumentation. Design features pre-bent plates to minimize intraoperative contouring.
Indications for Use
Indicated for patients requiring bone fixation for fractures, fusions, or osteotomies of the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Howmedica's Distal Humeral Plate (K890939)
- Luhr Fixation System (K951415)
- Synthes' Curved Reconstruction Plate (K912936)
- Modular Foot System (K001941)
- Link's May Tibia Bone Plates (K912936)
Related Devices
- K063460 — ACUMED CONGRUENT BONE PLATE SYSTEM · Acumed, LLC · Dec 7, 2006
- K071715 — ACUMED CONGRUENT BONE PLATE SYSTEM · Acumed, LLC · Jul 18, 2007
- K051410 — ACUMED RIB CONGRIENT BONE PLATE SYSTEM · Acumed, LLC · Jul 21, 2005
- K241599 — Signature Osteosynthesis Plate System · Signature Orthopaedic Pty , Ltd. · Feb 6, 2025
- K242995 — OIC Small / Mini Fragment Plate System · The Orthopaedic Implant Company · Oct 23, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
K012655
## NOV 0 7 2001
### Appendix V - 510(k) Summary
This summary regarding 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.93.
Classification Name: Plate, Fixation, Bone Common Name: Bone Plate Proprietary Name: Congruent Bone Plate System Proposed Regulatory Class: II Device Product Code: HRS Manufacturing Facility: Acumed, Inc. 10950 SW 5th Street, Suite 170 Beaverton, OR 97005 U.S.A. Establishment Registration No .: 3025141
Contact: Shari Jeffers Labeling/Promotional Materials: See Appendix III Substantial Equivalence:
Substantial Equitvatenee.
The Acumed Congruent Bone Plate System is similar in indication, intended use, material, design, and size The Howmedica's Distal Humeral Plate (K890939) and Luhr Fixation System (K951415), to Synthes' to Trommodiate o Disal Transval 1334), the Modular Foot System (K001941), and the One-Third Tubular Curred Reconstitution Fibia Bone Plates (K912936). Literature on these predicate devices is included in Appendix IV.
The Congruent Bone Plate System consists of bone plates and screws for fractures, fusions, and osteotomies. The bone plates are pre-bent to minimize bending which is done intraoperatively. Instruments are supplied with the implans to aid in the insertion of the plates and screws. Each of the plate styles utilizes the same screw types and screw instruments for insertion. All of the plates and screws are manufactured from titanium and are provided non-sterile. The screws were cleared for marketing and distribution under K942340 and K942341.
Congruent Bone plates are provided non-sterile and are individually packaged in a plastic bag. On file at Acumed is data which shows that the instrumentation and implants can be successfully steam sterilized under specific process parameters which will obtain a resulting SAL of 106. Information regarding labeling has been provided.
Predicate devices that are substantially equivalent to Acumed's Congruent Bone Plate System are Howmedica's Distal Humeral Plate and Luhr Fixation System; Synthes' Curved Reconstruction Plate, the Modular Foot System, and the One-Third Tubular Plate; and Link's May Tibia Bone Plates. All the devices mentioned above are manufactured from similar material, and have the same indication/intended use and similar design and size characteristics.
Based on the similarities between the Acumed Congruent Bone Plate System and the predicate devices studied, the safety and effectiveness of the Acumed Congruent Bone Plate System is expected to be similar to the predicate devices mentioned above.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 7 2001
Ms. Shari Jeffers Manager of Regulatory Affairs Acumed. Inc. 10950 SW 5th Street, Suite 170 Beaverton, Oregon 97005
Re: K012655
Trade/Device Name: Acumed Congruent Bone Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HRS Dated: August 6, 2001 Received: August 13, 2001
Dear Ms. Jeffers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{2}------------------------------------------------
## Page 2 - Ms. Shari Jeffers
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Signature
ek Celia M. Witten, Ph.D., M.D. Director Divison of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
NOV 0 7 2001
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Congruent Bone Plate System Device Name:
# Indications For Use:
The Acumed Congruent Bone Plate System provides fixation for fractures, fusions, or osteotomies for the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|---------------------------------------|----------------------------------------------------------|---------|--------------------------|
| | (Division Sign-Off) | | |
| | Division of General Restorative and Neurological Devices | | |
| | 510(k) Number | K012655 | |
| Prescription Use (Per 21 CFR 801.109) | OR | | Over-The-Counter Use |
| | | | (Optional Format 1-2-96) |
(Optional Format 1-2-96)