LOQTEQ SMALL FRAGMENT SET LOQTEQ CORTICAL SCREW3.5, SMALL HEAD, T15, SELF TAPPING LOQTEQ CORTICAL SCREW 3.5, T15, SELF T

K113652 · Aap Implantate AG · HRS · Aug 30, 2012 · Orthopedic

Device Facts

Record IDK113652
Device NameLOQTEQ SMALL FRAGMENT SET LOQTEQ CORTICAL SCREW3.5, SMALL HEAD, T15, SELF TAPPING LOQTEQ CORTICAL SCREW 3.5, T15, SELF T
ApplicantAap Implantate AG
Product CodeHRS · Orthopedic
Decision DateAug 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The aap LOQTEQ Small Fragment Set includes Straight Plates 3.5, Reconstruction Plates 3.5 and One Third Tubular Plates 3.5. The plates accept 3.5 mm locking screws and 3.5 mm cortical screws. They are intended for Fixation of fractures, osteotomies, and non-unions of the scapula, olecranon, humerus, radius, ulna, distal tibia, and fibula, particularly in osteopenic bone

Device Story

The LOQTEQ Small Fragment Set consists of bone plates (straight, reconstruction, 1/3 tubular) and cortical screws used for internal fixation of bone fractures, osteotomies, and non-unions. The system functions as an internal fixator when plates are used with locking screws. The device is intended for use by surgeons in clinical settings. The plates and screws provide mechanical stabilization of bone fragments to facilitate healing, with specific utility in osteopenic bone. The system includes a set of surgical instruments for implantation.

Clinical Evidence

No clinical data. Effectiveness and safety were demonstrated through non-clinical bench testing, including implant fatigue tests with progressive loading and static/dynamic 4-point bending tests in accordance with ASTM F382.

Technological Characteristics

Plates: cp Titanium (ASTM F67 or ISO 5832-2). Screws: Ti6Al4V (ASTM F136 or ISO 5832-3). System includes straight, reconstruction, and 1/3 tubular plates with 3.5 mm locking and cortical screws. Mechanical fixation via bone screws. Non-sterile or sterile (implied by surgical use).

Indications for Use

Indicated for fixation of fractures, osteotomies, and non-unions of the scapula, olecranon, humerus, radius, ulna, distal tibia, and fibula, particularly in patients with osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ oop Implantate AG aap LOQTEQ Small Fragment Set # ర్క్ ( ( ( ( ( ( ) ( ( ( ) ( ( ) ( ( ) ( ) Summary of Safety and Effectiveness KII 3.652 Page 1 of 2 | Sponsor: | aap Implantate AG<br>Lorenzweg 5<br>D-12099 Berlin<br>Germany | AUG 30 2012 | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Company Contact: | Dipl.-Ing. Marc Seegers<br>Phone: +49-30-750-19 -192<br>Fax: +49-30-750-19 - 111 | | | Date | December 5, 2011 | | | Trade Name: | aap LOQTEQ Small Fragment Set | | | Common Name: | Small Fragment Set | | | Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic<br>bone fixation appliances and accessories – Class II and<br>21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener -<br>Class II | | | Device Product Code and<br>Panel Code: | Orthopedics/87/ HRS: Plate, Fixation, Bone<br>Orthopedics/87/ HWC: Screw, Fixation, Bone | | | Predicate device: | Small Fragment Locking Compression Plate (LCP®) System,<br>Synthes (USA), K000684 (APR 28, 2000) and K011335 (JUL 27, 2001) and<br>Zimmer® Small Fragment Universal Locking System K082527 (DEC 18,<br>2008) | | | Device Description: | Bone plates and screws are used for fixation of bone fragments, i.e., for<br>treatment of bone fractures and other bone injuries. Bone plates are<br>fixed with bone screws. Bone plates and bone screws are implants. If<br>the plates are used in conjunction with locking screws, a so called inter-<br>nal fixator will be realized (internal fixation). | | | | The LOQTEQ® Small Fragment Set consists of:<br><ul><li>LOQTEQ® Straight Plate 3.5</li><li>LOQTEQ® Reconstruction Plate 3.5</li><li>LOQTEQ® 1/3 Tubular Plate 3.5</li><li>LOQTEQ® Cortical Screw 3.5, T15, self-tapping</li><li>LOQTEQ® Cortical Screw 3.5, small head T15, self-tapping</li><li>Cortical Screw 3.5, self-tapping</li><li>Cortical Screw 3.5, small head, self-tapping</li><li>Set of Instruments, Small Fragment Set</li></ul> | | | Material: | Plates are made of cp Titanium (ASTM F67 or ISO 5832-2).<br>Screws are made of Ti6Al4V (ASTM F136 or ISO 5832-3). | | {1}------------------------------------------------ aap Implantate AG aap LOQTEQ Small Fragment Set Indications: K113652 PAGE 2 of 2 Fixation of fractures, osteotomies, and non-unions of the so ranon, humerus, radius, ulna, distal tibia, and fibula, particularly in osteopenic bone Substantial Equivalence The Substantial Equivalence of the new device and the predicate device is based on similar intended use, design, functionality, components and materials in use. Documentation including mechanical testing to show the substantial equivalence and safety and effectiveness has been provided with this submission. Performance Data (Non-Clinical and / or Clinical): Non-clinical tests have been performed and show the effectiveness and safety of the device. #### Summary of Non-clinical tests: Type of test: Implant Fatigue tests with progressive loadings, representing worst case scenario with respect to clinical use and static and dynamic 4-Point bending tests acc. to ASTM F382 Assessment of test results: Substantial equivalence with respect to the mechanical performance of the aap system could be stated due to the test results gained. The subject device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses. Documentation regarding the mechanical testing to show the substantial equivalence and safety and effectiveness has been provided with this submission. {2}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 aap Implante AG % Mr. Marc Seegers, Dipl .- Ing., Director OA/RA Lorenzweig 5 D-12099 Berlin Germany Re: K113652 Trade/Device Name: aap LOOTEO Small Fragment Set Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS. HWC Dated: August 20, 2012 Received: August 22, 2012 Dear Mr. Seegers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set AUG 30 2012 Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. {3}------------------------------------------------ Page 2 - Mr. Marc Seegers forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Einell Keith As-Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use Statement #### 510(k) Number (if known): K113652 ## Device Name: LOQTEQ Small Fragment Set #### Indications for Use: The aap LOQTEQ Small Fragment Set includes Straight Plates 3.5, Reconstruction Plates 3.5 and One Third Tubular Plates 3.5. The plates accept 3.5 mm locking screws and 3.5 mm cortical screws. They are intended for Fixation of fractures, osteotomies, and non-unions of the scapula, olecranon, humerus, radius, ulna, distal tibia, and fibula, particularly in osteopenic bone Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aseb (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113652 Page _1_ of _1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...