PRCT2
K143721 · Fx Solutions · HRS · Aug 27, 2015 · Orthopedic
Device Facts
| Record ID | K143721 |
| Device Name | PRCT2 |
| Applicant | Fx Solutions |
| Product Code | HRS · Orthopedic |
| Decision Date | Aug 27, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
PRCT2 is indicated for use in fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus.
Device Story
PRCT2 is a bone fixation system comprising humeral plates, diaphyseal plates, locking/standard screws, trials, and surgical instruments. Used by orthopedic surgeons in clinical settings for internal fixation of proximal humerus fractures, osteotomies, and non-unions. System provides mechanical stabilization of bone segments to facilitate healing. Implants are surgically implanted to secure bone fragments; performance verified via mechanical bench testing against predicate devices.
Clinical Evidence
Bench testing only. Mechanical strength evaluated side-by-side against predicate devices per ASTM F382-99 (bone plates) and ASTM F543-13 (bone screws). Results demonstrate mechanical equivalence.
Technological Characteristics
System includes plates, screws, trials, and instruments. Materials: medical grade titanium alloy (TA6V ELI) per ISO 5832-3. Mechanical testing conducted per ASTM F382-99 and ASTM F543-13. No software or electronic components.
Indications for Use
Indicated for patients with fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Synthes (USA) 3.5mm LCP Periarticular Proximal Humerus Plates (K082625)
- Internal Fixation Systems, Inc. IFS Bone Plates, Screws, and Washers (K110086)
Related Devices
- K242436 — Proximal Humerus Fixation System · Skeletal Dynamics, Inc. · Nov 14, 2024
- K063095 — HUMERUS LOCKING PLATING SYSTEM · Newclip Technics · Dec 22, 2006
- K254249 — HKT Anatomical Locking Trauma System · Hankil Tech Medical Co., Ltd. · Feb 27, 2026
- K120108 — APTUS PROXIMAL HUMERUS SYSTEM · Medartis AG · Apr 27, 2012
- K100146 — EVOLVE EPS ORTHOLOC · Wrightmedicaltechnologyinc · Feb 2, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo are three stylized human profiles facing to the right, stacked on top of each other.
August 27, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Fx Solutions % Mr. Frank Ferguson CEO Ferguson Medical International Device Consultants LLC 332 Laskin Road, Suite 437 Virginia Beach, Virginia 23451
Re: K143721 Trade/Device Name: PRCT2 Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: July 29, 2015 Received: July 31, 2015
Dear Mr. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part
{1}------------------------------------------------
807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K143721
# INDICATIONS FOR USE
510(k) Number: K143721
Device Name: PRCT2
Indications For Use:
PRCT2 is indicated for use in fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus.
Prescription Use _______XX (21 CFR 801 Subpart D)
OR
Over-The- Counter Use __ (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the Ex Solutions PRCT2 device.
DATE PREPARED: 17 August 2015
#### APPLICANTS NAME AND ADDRESS:
Fx Solutions Attn: Jean-Jacques Martin 1663 rue de Majornas 01440 Viriat France Phone: +33 474 55 35 55; FAX: +33 474 52 44 01 jjmartin@fxsolutions.fr
#### APPLICANTS CONTACT PERSON IN THE USA:
Ferquson Medical International Device Consultants LLC Attn: Frank Ferguson, CEO 332 Laskin Road Suite 437 Virginia Beach, VA 23451 Phone: 757.286.4383; FAX: 801.749.0451 frankf@ferqusonmedical-idc.com
#### DEVICE NAME:
Trade Name: PRCT2 Common Name: Bone plates and screws Classification Name: 21CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories
{4}------------------------------------------------
21CFR 888.3040 Smooth or threaded metallic bone fixation fastener Product Code: HRS: Plate, Fixation, Bone HWC: Screw, Fixation, Bone
## LEGALLY MARKETED DEVICES TO WHICH FX SOLUTIONS IS CLAIMING SUBSTANTIAL EQUIVALENCE:
Device Names: Synthes (USA) 3.5mm LCP Periarticular Proximal Humerus Plates (K082625), Internal Fixation Systems, Inc. IFS Bone Plates, Screws, and Washers (K110086)
## DEVICE DESCRIPTION:
The PRCT2 is a system of plates and screws intended to be used in the fixation of fractures of the humerus. PRCT2 consists of 5 components:
- 1. Humeral Plates
- 2. Diaphyseal Plates
- 3. Screws (Locking and Standard)
- 4. Trials
- 5. Instruments
All implants and trials in the PRCT2 system are manufactured from medical grade titanium alloy (TA6V ELI) according to ISO 5832-3.
### INTENDED USE:
PRCT2 is indicated for use in fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus.
## SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICES:
The components of the Fx Solutions PRCT2 device are offered in the same general range of sizes and shapes of the predicate devices. The materials used for the Fx Solutions device are the same as that used to manufacture the predicates.
{5}------------------------------------------------
### PERFORMANCE DATA:
The mechanical properties of the Fx Solutions PRCT2 device were tested in accordance with ASTM F382-99, Standard Specification and Test Method for Metallic Bone Plates, and ASTM F543-13, Standard Specification and Test Method for Metallic Bone Screws.
The Fx Solutions device was mechanically tested side-by-side against the predicate device. The test results show equivalence in terms of mechanical strength, therefore substantially equivalent.
### CONCLUSION:
Based upon the testing and comparison to the predicate device, the Fx Solutions PRCT2 device has the same intended use and similar technological characteristics. The device performs as intended and does not raise any new safety or effectiveness issues, and is therefore substantially equivalent to the predicate devices.