ALPHALOK Plating System
K221558 · Vilex, LLC · HRS · Nov 15, 2022 · Orthopedic
Device Facts
| Record ID | K221558 |
| Device Name | ALPHALOK Plating System |
| Applicant | Vilex, LLC |
| Product Code | HRS · Orthopedic |
| Decision Date | Nov 15, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The ALPHALOK™ Plating System bone plates, screws, and washers are intended for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, ankle, and digits. Specific examples include: - Forefoot, Midfoot, and Hindfoot Osteotomies - Metatarsals and Metacarpals Corrections and Osteotomies - Stabilization and Fixation of Metatarsal and Metacarpal Fractures - Stabilization and Fixation of Ankle Fractures - Syndesmosis Joint Stabilization - Arthrodesis of Metatarsophalangeal (MTP) and Metacarpophalangeal (MCP) joints - Flatfoot and Cavus Foot Corrections - Charcot Fixation
Device Story
ALPHALOK Plating System provides bone fixation via titanium plates, screws, and washers; features polyaxial locking technology and low-profile design. System includes Met Fx (straight, T, Y, Jones, mini plates) and Recon (fusion, Lapidus, peanut, tab, MTP, Cotton & Evans, utility, ankle fracture plates) components. Used by orthopedic surgeons in clinical settings for bone fracture stabilization and osteotomy fixation. Implants are titanium (Ti-6Al-4V); instruments are stainless steel. Surgeon uses manual instrumentation (drills, drivers, guides) to secure plates to bone fragments. Provides mechanical stability to facilitate bone healing.
Clinical Evidence
No clinical data. Evidence consists of bench testing: static four-point bend testing (ASTM F382-17), torsional properties testing (ASTM F543-17), axial pullout calculations (per FDA guidance), and driving torque testing (ASTM F543-17). Cleaning and sterilization validations performed per ANSI/AAMI/ISO 17665-1.
Technological Characteristics
Implant material: Titanium alloy (Ti-6Al-4V, ASTM F136). Instrument material: Stainless steel. Design: Polyaxial locking plates, screws, and washers. Mechanical testing standards: ASTM F382-17 (bend), ASTM F543-17 (torsion/torque). Sterilization: Validated per ANSI/AAMI/ISO 17665-1.
Indications for Use
Indicated for patients requiring bone fixation, osteotomies, or fracture stabilization in upper/lower extremities, specifically hand, wrist, foot, ankle, and digits. Includes forefoot/midfoot/hindfoot osteotomies, metatarsal/metacarpal corrections/fractures, ankle fractures, syndesmosis stabilization, MTP/MCP arthrodesis, flatfoot/cavus foot corrections, and Charcot fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Vilex ALPHALOK™ Plating System (K212348)
- Paragon28 Baby Gorilla/Gorilla Plating System (K190365)
- Wright Medical Ortholoc 3Di Plating System (K163044)
- OrthoPro TC Plating System (K094037)
Related Devices
- K230204 — ALPHALOK Plating System · Vilex, LLC · Apr 25, 2023
- K243761 — A.L.P.S. Small Fragment Plating System · Biomet Orthopedics · Feb 19, 2025
- K222665 — Medline UNITE® Ankle Fusion Plating System · Medline Industries, Inc. · Jan 25, 2023
- K180808 — Omni Foot Plating System · Extremity Medical, LLC · May 22, 2018
- K123562 — FLOWER SMALL AND MEDIUM BONE IMPLANTS · Flower Orthopedics Corporation · Mar 29, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
November 15, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vilex LLC Brock Johnson President 111 Moffitt Street McMinnville, Tennessee 37110
Re: K221558
Trade/Device Name: ALPHALOK Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: May 27, 2022 Received: May 31, 2022
Dear Brock Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221558
Device Name ALPHALOKTM Plating System
### Indications for Use (Describe)
The ALPHALOK™ Plating System bone plates, screws, and washers are intended for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, ankle, and digits. Specific examples include:
- · Forefoot, Midfoot, and Hindfoot Osteotomies
- · Metatarsals and Metacarpals Corrections and Osteotomies
- · Stabilization and Fixation of Metatarsal and Metacarpal Fractures
- · Stabilization and Fixation of Ankle Fractures
- · Syndesmosis Joint Stabilization
- · Arthrodesis of Metatarsophalangeal (MTP) and Metacarpophalangeal (MCP) joints
- · Flatfoot and Cavus Foot Corrections
- · Charcot Fixation
| | Type of Use (Select one or both, as applicable) | | | |
|--|-------------------------------------------------|--|--|--|
| | | | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Vilex LLC
111 Moffitt Street
McMinnville, TN 37110
Contact Person: Brock Johnson, President of Vilex
Phone: (801) 916-4157
brock.johnson@yilex.com
Date Prepared: May 27, 2022
#### II. Device
| Device Proprietary Name: | ALPHALOK TM Plating System |
|--------------------------|--------------------------------------------------------------------|
| Common or Usual Name: | Bone Fixation Plates, Screws, and Washers |
| Classification Name: | Plate, Fixation, Bone<br>Screw, Fixation, Bone<br>Washer, Bolt Nut |
| Regulation Number: | 21 CFR 888.3030, Primary<br>21 CFR 888.3040 |
| Product Code: | HRS, HWC, HTN |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary Predicate: Vilex ALPHALOK™ Plating System (K212348)
{4}------------------------------------------------
Additional Predicates:
- Paragon28 Baby Gorilla/Gorilla Plating System (K190365) ●
- Wright Medical Ortholoc 3Di Plating System (K163044) ●
- . OrthoPro TC Plating System (K094037)
#### IV. Device Description
The ALPHALOK™ Plating System, consisting of the ALPHALOK™ Met Fx and ALPHALOK™ Recon, is a multi-indication reconstruction solution providing polyaxial locking technology and lowprofile design.
The ALPHALOK™ Met Fx contains various sizes of non-sterile straight plates, T plates, Y plates, Jones plates, mini plates, and locking and non-locking screws.
The ALPHALOK™ Recon contains various sizes of non-sterile fusion plates, Lapidus plates, peanut plates, straight plates, tab plates, T-plates, MTP plates, Cotton & Evans plates, utility plates, and Ankle Fracture plates along with locking and non-locking screws, and washers.
All implant components are manufactured from titanium (Ti-6A1-4V, ASTM F136).
Specific instrumentation including wires, drills, torx drivers, and drill guides are required for use with the system. The ALPHALOK™ instruments are manufactured from stainless steel.
#### V. Indications for Use
The ALPHALOK™ Plating System bone plates, screws, and washers are intended for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, ankle, and digits. Specific examples include:
- . Forefoot, Midfoot, and Hindfoot Osteotomies
- . Metatarsals and Metacarpals Corrections and Osteotomies
- Stabilization and Fixation of Metatarsal and Metacarpal Fractures .
- . Stabilization and Fixation of Ankle Fractures
- Syndesmosis Joint Stabilization
- Arthrodesis of Metatarsophalangeal (MTP) and Metacarpophalangeal (MCP) joints ●
- Flatfoot and Cavus Foot Corrections ●
- Charcot Fixation
#### VI. Comparison of Technological Characteristics
The subject and predicate devices have similar intended uses and share identical core characteristics.
The systems are intended to be used in the hand, foot, and ankle along with other upper and lower extremities small bones. The subject and predicate systems include bone plates along with locking and non-locking screws with similar implant designs made from titanium alloy material. Similar instrumentation is included in all the system.
{5}------------------------------------------------
The technological differences between the subject device and predicate devices do not raise different questions of safety or effectiveness and substantial equivalence is demonstrated through the testing described below.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
- Static four point bend testing per ASTM F382-17 ●
- Torsional properties testing per ASTM F543-17
- Axial pullout calculations per FDA Guidance
- . Driving torque testing per ASTM F543-17
- Engineering Calculations
In addition, cleaning and sterilization validations, performed in accordance with ANSI/AAMI/ISO 17665-1, from the applicant's own predicate device were leveraged. Biocompatibility, cleaning and sterilization are identical to K212348 and no changes have been made with respect to material, manufacturing, cleaning, or sterilization.
#### VIII. Conclusion
The information provided above supports that the ALPHALOK™ Plating System is as safe and effective as the predicate device. Although minor differences in design exist between the subject and predicate devices, the testing supports that these differences do not raise any new questions of safety and effectiveness. Therefore, it is concluded that the ALPHALOK™ Plating System is substantially equivalent to the predicate device.
-----------------------------This space intentionally left blank--------------------------------------------------------------------------------------------------------------