The Flower Small and Medium Implants set is intended for use for internal fixation of fractures and reconstruction of bones, including the scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, hand and foot in adults and for use in long bones in adolescents (12-21) in whom the growth plates have fused. Examples of these internal fixations and reconstructions include compression fractures, intra-articular and extraaticular fractures, displaced fractures, osteotomies, non-unions and mal-unions. This system can be used for palmar, ventral, dorsal or orthogonal application.
Device Story
Bone plating system for internal fixation and bone reconstruction. Components include pure titanium (ISO 5832-2) plates of various geometries and titanium alloy screws. Surgeon selects implant based on anatomical site; plate contoured to bone; locking screws inserted into locking holes to provide fixed-angle stability. Used in clinical/surgical settings by orthopedic surgeons. Provides mechanical stabilization of fractures to facilitate healing. System includes general purpose instruments.
Clinical Evidence
Bench testing only. Biocompatibility established per ISO 10993-1 and ISO 10993-5. Sterilization validation performed for gamma sterilization. Packaging validation and shelf-life testing confirmed sterile barrier integrity.
Technological Characteristics
Plates: Pure titanium (ISO 5832-2). Screws: Titanium alloy. Fixed-angle locking mechanism. Various plate geometries (straight, reconstruction, T-shaped, H-shaped, etc.). Gamma sterilization. No software or electronic components.
Indications for Use
Indicated for internal fixation and reconstruction of fractures in adults and adolescents (12-21) with fused growth plates. Applicable to scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, hand, and foot. Includes compression, intra-articular, extra-articular, displaced fractures, osteotomies, non-unions, and mal-unions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes USA's 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications (K082807)
Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) (K092609)
Stryker's VariAx Distal Radius Locked Plating System Line Extension for Addition of Aiming Blocks (K112455)
Synthes 2.4mm VA-LCP Intercarpal Fusion System (K103243)
Submission Summary (Full Text)
{0}------------------------------------------------
# K123562 page 1/3
# 510(k) SUMMARY
# MAR 2 9 2013
# Flower Orthopedics Corporation's Flower Small and Medium Implants
# Submitter's Name, Address, Telephone Number, Contact Person, and Date Prepared
Flower Orthopedics Corporation 7715 Crittenden Street, #413 Philadelphia, PA 19118
Phone: +1 267 437 3063 Facsimile: +1 267 437 3072
Contact Person: Oliver B. Burckhardt
Date Prepared: March 29, 2013
### Name of Device and Name/Address of Sponsor
Flower Small and Medium Implants
#### Common or Usual Name
Bone plating system
# Classification Name/ Product Code
Classification Name: 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliance and accessories
Product Codes: HRS (Plate, Fixation, Bone), HWC (Screw, Fixation, Bone)
#### Predicate Devices
- Synthes USA's 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications (K082807)
- Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) (K092609)
- Stryker's VariAx Distal Radius Locked Plating System Line Extension for Addition of Aiming Blocks (K112455)
KLS-Martin Hand Plating System (K040598)
Synthes 2.4mm VA-LCP Intercarpal Fusion System (K103243)
{1}------------------------------------------------
#### Intended Use / Indications for Use
The Flower Small and Medium Implants set is intended for use for internal fixation of fractures and reconstruction of bones, including the scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, hand and foot in adults and for use in long bones in adolescents (12-21) in whom the growth plates have fused. Examples of these internal fixations and reconstructions include compression fractures, intra-articular and extraaticular fractures, displaced fractures, osteotomies, non-unions and mal-unions.
This system can be used for palmar, ventral, dorsal or orthogonal application.
## Technological Characteristics
The Flower Small and Medium Implants set consists of the following components and accessories: pure titanium small straight plates, small and medium reconstruction plates, medium osteosynthesis plates, proximal humerus plates, distal radius plates, T-plates, angular T-shaped plates, H-shaped plates, mediocarpal plate; and titanium alloy screws. The device is also provided with general purpose instruments. All plates are made of pure titanium (ISO 5832-2).
The Flower Small and Medium Implants set provides fixed-angle lockable screws and plates to assist with internal fixation of fractures and reconstruction of bones. The principles of operation of the device are similar to other bone plating systems. The plates are comprised of various shapes. alignments, thicknesses, widths, and lengths designed to contour to different bones and locations on the body for internal fixation or reconstruction following fracture. .Each of the plates contains several locking holes that allow for the insertion of Flower Small and Medium Implants locking screws. To use the Flower Small and Medium Implants set, the surgeon first selects an implant of the appropriate size and shape based on the intended site of use. The plate should be placed in an appropriate location on the given bone or anatomical location in need of repair.
#### Performance Data
In support of this 510(k) Premarket Notification, Flower Orthopedics has conducted the following testing. In all instances, the Flower Small and Medium Implants set functioned as intended.
- Biocompatibility in accordance with ISO 10993-1, ISO 10993-5 was established, demonstrating . that the materials are non-cytotoxic and biocompatible.
- . Sterilization validation of implants and instruments demonstrated assurance level of 10° for this method of sterilization using the specified gamma sterilization cycle.
- Packaging validation and shelf life testing ensured that the packaging can maintain its physical . integrity and maintain a sterile barrier over the stated period.
# Substantial Equivalence
The Flower Small and Medium Implants system is very similar to Synthes's 3.5 mm and 4.5 mm Locking Compression Plate (LCP) System with Expanded Indications (K082807), the Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) (K092609), Stryker's VariAx Distal Radius Locked Plating System Line Extension for Addition of Aiming Blocks (K112455), KLS-Martin's Hand Plating System (K040598), and Synthes 2.4mm VA-LCP Intercarpal Fusion System (K103243). The Flower Small and Medium Implants system has the same intended
{2}------------------------------------------------
uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The overall surgical procedure for the Flower Small and Medium Implants set and the predicate devices are very similar and there are no new types of safety or effectiveness concerns. The minor difference in the locking feature between the Flower system and the predicate systems do not significantly alter the surgical technique. The minor technological differences between the Flower Small and Medium Implants and its predicate devices, e.g., minor differences in the range of available geometries and dimensions, raise no new types of safety or effectiveness questions because these size differences are very minor and are largely encompassed within the range of similar parameters in the predicate devices. Engineering analysis has been performed to demonstrate that the Flower Small and Medium Implants system provides appropriate mechanical strength for its intended use. Thus, the Flower Small and Medium Implants system is substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
## Letter dated: March 29, 2013
Flower Orthopedics Corporation % Hogan Lovells US LLP Ms. Janice M. Hogan, Partner 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K123562
Trade/Device Name: Flower Small and Medium Implants Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 1, 2013 Received: March 1, 2013
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
# Page 2 -Ms. Janice M. Hogan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
K123562 510(k) Number (if known):
Device Name: Flower Small and Medium Implants
Indications for Use:
The Flower Small and Medium Implants set is intended for use for internal fixation of fractures and reconstruction of bones, including the scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula, hand and foot in adults and for use in long bones in adolescents (12-21) in whom the growth plates have fused. Examples of these internal fixations and reconstructions include compression fractures, intra-articular and extraaticular fractures, displaced fractures, osteotomies, non-unions and mal-unions.
This system can be used for palmar, ventral, dorsal or orthogonal application.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Krishna R 2013.03.2
> Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.