SYNTHES UNIVERSAL LOCKING TROCHANTER STABILIZATION PLATE (ULTSP)

K052677 · Synthes (Usa) · HRS · Oct 12, 2005 · Orthopedic

Device Facts

Record IDK052677
Device NameSYNTHES UNIVERSAL LOCKING TROCHANTER STABILIZATION PLATE (ULTSP)
ApplicantSynthes (Usa)
Product CodeHRS · Orthopedic
Decision DateOct 12, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Synthes Universal Locking Trochanter Stabilization Plate for DHS® or LCP® DHHS is intended to treat stable and unstable intertrochanteric, subtrochanteric, pertrochanteric and basilar neck fractures when used in conjunction with the Synthes Dynamic Hip Screw (DHS®) or the LCP® Dynamic Helical Hip System (DHHS) side plates with four or more holes.

Device Story

Metallic bone fixation plate; attaches to Synthes Dynamic Hip Screw (DHS) or LCP Dynamic Helical Hip System (DHHS) sideplates. Features proximal extension for greater trochanter stabilization; includes screw holes for 4.5 mm cortex screws, 6.5 mm cancellous screws, or 6.5/7.0/7.3 mm cannulated screws; proximal locking holes accept 3.5 mm locking screws, k-wires, cables, or sutures. Used by orthopedic surgeons in clinical/surgical settings to provide supplemental fixation for femoral fractures. Enhances construct stability; aids in fracture healing.

Clinical Evidence

No clinical data provided; substantial equivalence based on design and intended use comparison to predicate device.

Technological Characteristics

Metallic bone fixation plate; dimensions/form factor designed for attachment to DHS/LCP DHHS sideplates. Features locking and non-locking screw holes. Materials and sterilization methods are consistent with standard orthopedic metallic fixation appliances (21 CFR 888.3030).

Indications for Use

Indicated for patients with stable and unstable intertrochanteric, subtrochanteric, pertrochanteric, and basilar neck fractures requiring fixation with DHS or LCP DHHS side plates.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # OCT 1 2 2005 ### Page 1 __ of __ 1 3.0 510(k) Summary Synthes (USA) Sponsor: 1302 Wrights Lane East West Chester, PA 19380 (610) 719-5000 Synthes Universal Locking Trochanter Stabilization Plate (ULTSP) Device Name: Class II, §888.3030 - Single/multiple component metallic bone Classification: fixation appliances and accessories Synthes Trochanter Stabilization Plate (TSP) for DHS Predicate Devices: The Synthes ULTSP for DHS® and LCP DHHS fits over the Device Description: sideplate of the Synthes DHS and LCP® DHHS. The ULTSP screw holes line up with the most proximal screw holes of the DHS and LCP DHHS sideplate and accept 4.5 mm cortex screws used for femoral shaft fixation. Slots accommodate both the DHS lag screw or the LCP DHHS Helix Blade and a parallel anti-rotational screw. The screw slot for the anti-rotational screw accepts a 6.5 mm cancellous bone screw or 6.5 mm, 7.0 mm or 7.3 mm cannulated screw. The proximal extension of the plate resides along the lateral side of the greater trochanter. The proximal locking holes of the ULTSP accept 3.5 mm locking screws and/or k-wires, cables, or sutures up to 2.0 mm in diameter. The Synthes Universal Locking Trochanter Stabilization Plate for Intended Use: DHS® or LCP DHHS is intended to treat stable and unstable intertrochanteric, subtrochanteric, pertrochanteric and basilar neck fractures when used in conjunction with the Synthes Dynamic Hip Screw (DHS®) or the LCP Dynamic Helical Hip System (DHHS) side plates with four or more holes. Substantial Information presented supports substantial equivalence. Equivalence: {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is a stylized graphic of a human figure, represented by three curved lines, symbolizing health and well-being. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 2 2005 Sheri L. Musgnung Sr. Regulatory Specialist Synthes (USA) 1302 Wrights Lane East West Chester, Pennsylvania 19380 Re: K052677 Trade/Device Name: Synthes (USA) Universal Locking Trochanter Stabilization Plate (ULTSP) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: September 27, 2005 Received: September 29, 2005 Dear Ms. Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreative) it cegandy to together that of the Medical Device Amendments, or to conninered price to May 20, 1978, in accordance with the provisions of the Federal Food, Drug, de riccs that have been rocksonical require approval of a premarket approval application (PMA). alla Costiete Act (71ct) that do not required to the general controls provisions of the Act. The r ou may, dicierore, mance the Act include requirements for annual registration, listing of general controls provisions of use ice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 above) als. Existing major regulations affecting your device can may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Outs neements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease oc advised that I Dri of issuaires over device complies with other requirements of the Act that I DA has made a develminations administered by other Federal agencies. You must or any I cuchar statutes and regulations and and limited to: registration and listing (21 comply with an the Act 31equirements) (1) ; good manufacturing practice requirements as set CI K I at 607), adomig (21 OF R Part 820), and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Sheri L. Musgnung This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w you'ls cognifinding of substantial equivalence of your device to a legally premailed notifications on "certain" for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you a sent office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, 2 Sor Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use KOSZ677 510(k) Number (if known): Device Name: 2.0 Synthes (USA) Universal Locking Trochanter Stabilization Plate (ULTSP) INDICATIONS FOR USE: The Synthes Universal Locking Trochanter Stabilization Plate for DHS® or LCP® DHHS is intended to treat stable and unstable intertrochanteric, subtrochanteric, pertrochanteric and basilar neck fractures when used in conjunction with the Synthes Dynamic Hip Screw (DHS®) or the LCP® Dynamic Helical Hip System (DHHS) sideplates with four or more holes. Prescription Use X (Per 21 CFR 801.109) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...