ABX PENTRA MICRO ALB CONTROL L/H
Device Facts
| Record ID | K133676 |
|---|---|
| Device Name | ABX PENTRA MICRO ALB CONTROL L/H |
| Applicant | HORIBA ABX SAS |
| Product Code | JJY · Clinical Chemistry |
| Decision Date | Jul 11, 2014 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Indications for Use
ABX Pentra Micro ALB Control L/H is a quality control used to monitor the performance of ABX Pentra Micro ALBUMIN CP determination by immunoturbidimetry.
Device Story
ABX Pentra Micro ALB Control L/H is a two-level (Low/High) liquid quality control material; composed of human urine, added chemicals, human-origin constituents, stabilizers, and preservatives. Used on ABX PENTRA 400 benchtop clinical chemistry analyzer; operated by laboratory personnel. Device provides known values and confidence intervals for monitoring immunoturbidimetric microalbumin assays. Healthcare providers compare analyzer results against provided control ranges to verify assay precision and accuracy; ensures reliability of patient test results. No modifications to the analyzer are required for use.
Clinical Evidence
Bench testing only. Real-time stability studies confirmed shelf life of 24 months at 2-8°C and open vial stability of 90 days at 2-8°C. Acceptance criteria met.
Technological Characteristics
Liquid solution; human urine matrix with stabilizers/preservatives. Two-level (Low/High) control. 10mL vial format. Used for immunoturbidimetric assay quality control. No electronic components or software algorithms; standalone control material.
Indications for Use
Indicated for use as a quality control material to monitor the performance of ABX Pentra Micro ALBUMIN CP immunoturbidimetric assays on the ABX PENTRA 400 clinical chemistry analyzer.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
- K052242 — ADVIA CHEMISTRY MICROALBUMIN CONTROLS · Bayer Healthcare, LLC · Oct 3, 2005
- K122177 — ELITECH CLINICAL SYSTEMS URINE TOTAL PROTEIN PLUS STANDARD 100 MG/DL, ELITECH CLINICAL SYSTEMS URINE CONTROL BI-LEVEL · Elitechgroup · Feb 15, 2013
- K121746 — RANDOX MICROALBUMIN CONTROL LEVEL 1 AND CONTROL LEVEL 2 (LIQUID) · Randox Laboratories, Ltd. · Jul 11, 2012