AUDIT MICROCV TUMOR MARKERS LINEARITY SET, MODEL K719M-5
Applicant
Aalto Scientific, Ltd.
Product Code
JJY · Clinical Chemistry
Decision Date
Oct 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The Tumor Markers Linearity Set is intended to simulate human patient serum samples for the purpose of verifying and validating the Analytical Measurement Range for non-waived turnor marker testing methods as identified in the package insert.
Device Story
Audit™ MicroCV™ Tumor Markers Linearity Set is a human-serum-based, lyophilized, five-level quality control material. It contains six analytes: AFP, CEA, CA 125, CA 15-3, free PSA, and total PSA. Levels A through E are prepared via linear dilution to establish a linear relationship across the analytical measurement range of clinical chemistry instruments. The device is used by laboratory professionals to verify calibration and reportable ranges of tumor marker assays. It serves as a surrogate for patient samples to ensure instrument accuracy and performance consistency. Benefits include improved confidence in clinical diagnostic results for tumor markers.
Clinical Evidence
No clinical data. Bench testing only. Linearity was determined using CLSI EP6-A guidelines. Stability studies (accelerated and real-time) confirmed a 2-year shelf life and 7-day open-vial stability at 2-8°C.
Technological Characteristics
Human-serum-based, lyophilized, five-level QC material. Contains 6 analytes: AFP, CEA, CA 125, CA 15-3, free PSA, total PSA. Levels are related by linear dilution. Storage at 2-8° C.
Indications for Use
Indicated for use as a quality control material to verify the linear operating range of tumor marker test systems (AFP, CEA, CA-125, CA 15-3, free PSA, total PSA) in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Audit™ MicroCVTM General Chemistry Linearity Set (K042318)
Related Devices
K143573 — Audit MicroControls Linearity FD Tumor Markers II · Aalto Scientific, Ltd. · Jan 16, 2015
K101809 — FUJIREBIO DIAGNOSTICS TUMOR MARKER CONTROL MODEL 108-20 · Fujirebio Diagnostics,Inc. · Oct 5, 2010
K163629 — Audit MicroControls Linearity FD Tumor Markers II · Aalto Scientific, Ltd. · Mar 10, 2017
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k082717
B. Purpose for Submission:
New Device
C. Measurand:
Control Materials for Alpha fetoprotein (AFP), Carcinoembryonic antigen (CEA), CA-125, CA 15-3, Free Prostate specific antigen (free PSA), and total prostate specific antigen (total PSA) test systems.
D. Type of Test:
Quality control material – linearity verifiers for 6 serum tumor markers
E. Applicant:
Aalto Scientific, Ltd.
F. Proprietary and Established Names:
Audit™ MicroCV™ Tumor Markers Linearity Set
G. Regulatory Information:
1. Regulation section:
21 CFR§ 862.1660, Quality Control Material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJY, Multi-analyte controls, all kinds (assayed)
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The Audit™ MicroCV™ Tumor Markers Linearity Set consists of five levels in Human based serum. Each level contains the following analytes: Alpha fetoprotein (AFP), Carcinoembryonic antigen (CEA), CA-125, CA 15-3, Prostate specific antigen-free (free PSA), total PSA. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents, and instruments. This product may be used for proficiency testing in interlaboratory surveys.
3. Special conditions for use statement(s):
Prescription use only.
4. Special instrument requirements:
Quality control material for the analyzer systems specified in the package insert.
I. Device Description:
The Audit™ MicroCV™ Tumor Markers Linearity Set is a human serum based, lyophilized, five level set of QC material, with each level containing 6 analytes. It is used to confirm the linear operating range of Tumor Marker methods for the analytes listed.
{1}
Level A is near the lower limit level and Level E has concentrations near the upper limit of instruments. Levels B - D are related by linear dilution of Level A and Level E. Each serum, plasma or whole blood donor unit used in the preparation of this material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV and nonreactive for HBSAg.
# J. Substantial Equivalence Information:
1. Predicate device name(s): Audit MicroCV General Chemistry Linearity Set
2. Predicate 510(k) number(s): k042318
3. Comparison with predicate:
| Characteristics | AuditTM MicroCVTM Tumor Markers Linearity Set (New Device) | AuditTM MicroCVTM General Chemistry Linearity Set (K042318) |
| --- | --- | --- |
| Intended Use | The AuditTM MicroCVTM Tumor Markers Linearity Set consists of five levels in Human based serum. Each level contains the following analytes: Alpha fetoprotein (AFP), Carcinoembryonic antigen (CEA), CA-125, CA 15-3, Prostate specific antigen-free (free PSA), total PSA. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents, and instruments. This product may be used for proficiency testing in interlaboratory surveys. | AuditTM MicroCVTM General Chemistry Linearity Set is assayed quality control material consisting of human based serum. It is intended to simulate human patient serum samples for the purpose of monitoring the precision and to detect systematic analytical deviations of laboratory testing procedures. This product may also be used as unassayed quality control material for these same analytes and may be used for proficiency testing in interlaboratory surveys. In addition, this product may also be used to perform CLIA directed calibration verification for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations. |
| Number of Analytes per vial | 6 | 30 |
| Number of levels per set | 5 | 5 |
| Contents | 5 x 1 mL | 5 x 5 mL |
| Matrix | Human based serum | Human Based Serum |
| Type of Analytes | Cancer Markers | General Chemistry |
| Form | Lyophilized | Lyophilized |
| Stabilizers | None | None |
{2}
| Preservatives | Sodium Azide | Sorbitol Sodium Azide |
| --- | --- | --- |
| Storage | 2 to 8°C Until expiration date | 2 to 8°C Until expiration date |
| Reconstituted Stability | 7 days at 2 to 8°C | 7 days at 2 to 8°C except for enzymes and bilirubin, which are 48 hours |
# K. Standard/Guidance Document Referenced (if applicable):
None referenced.
# L. Test Principle:
Not applicable.
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Linearity was determined using CLSI EP6-A. An example linearity graph is as follows:

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Materials are obtained from approved vendors and put through an internal quality control process.
Stability:
The stability protocols and acceptance criteria for this product were reviewed and found to be acceptable. The product has an open-vial stability of 7 days when stored at $2 - 8^{\circ}\mathrm{C}$ . An accelerated stability study determined a shelf life stability of 2 years when stored at $2 - 8^{\circ}\mathrm{C}$ , and real-time studies are ongoing.
Value Assignment:
{3}
Level A (low pool) all analytes are adjusted to be within 5-10% of the lowest limit of measurability for their corresponding assay. For Level E (high pool), all analytes are adjusted to be within 5% to 10% of the highest limit of measurability for their corresponding assay. Levels are then diluted to the following:
Level A = 100% Low Pool
Level B = 75% Low Pool, 25% High Pool
Level C = 50% Low Pool, 50% High Pool
Level D = 25% Low Pool, 75% High Pool
Level E = 100% High Pool
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.