LIQUICHEK MICROALBUMIN CONTROL LEVEL 1, LEVEL 2, CONTROL MINIPAK
Applicant
Bio-Rad, Diagnostics Grp.
Product Code
JJY · Clinical Chemistry
Decision Date
Dec 11, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Liquichek Microalbumin Control is intended for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for the analytes listed in this package insert. For Prescription Use Only
Device Story
Liquichek Microalbumin Control is a liquid, human urine-based quality control material; used in clinical laboratories to monitor precision of testing procedures for microalbumin and creatinine. Device is provided in liquid form; stored at 2-8°C; stable for 30 days once opened. Healthcare providers use the control to verify the accuracy and reliability of laboratory assays; results help ensure consistent performance of diagnostic tests for patient samples. No automated processing or software is involved.
Clinical Evidence
No clinical data. Stability studies performed to validate open vial stability (30 days at 2-8°C) and shelf life (two years at 2-8°C).
Technological Characteristics
Liquid control material derived from human urine with added human-origin constituents, chemicals, preservatives, and stabilizers. Provided in 10 mL vials. Storage at 2-8°C. No specific sensing or energy source; intended for use on existing laboratory analyzers.
Indications for Use
Indicated for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for microalbumin and creatinine in urine.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
K070848 — LIQUICHEK URINALYSIS CONTROL (BI-LEVEL), CONTROL LEVEL1, LEVEL 2,MINIPAK, 435,436,437,435X · Bio-Rad, Diagnostics Grp. · Apr 25, 2007
K971954 — LIQUICHEK URINE CHEMISTRY CONTROL (CAT.NO. 397,398) · Bio-Rad · Jun 11, 1997
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k072835
B. Purpose for Submission:
Marketing of laboratory control materials
C. Measurand:
Microalbumin and creatinine in urine
D. Type of Test:
This product is used as a quality control material to monitor the precision of laboratory testing procedures for microalbumin and creatinine in urine.
E. Applicant:
Bio-Rad Laboratories
F. Proprietary and Established Names:
Liquichek™ Microalbumin Control Level 1
Liquichek™ Microalbumin Control Level 2
Liquichek™ Microalbumin Control Level MiniPak
G. Regulatory Information:
1. Regulation section:
21CFR862.1660 Quality control material (assayed and unassayed)
2. Classification:
Class I, reserved
3. Product code:
JJY
4. Panel:
(75) Chemistry
H. Intended Use:
1. Intended use(s):
Liquichek Microalbumin Control is intended for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.
2. Indication(s) for use:
See intended use (above).
3. Special conditions for use statement(s):
For Prescription Use Only
{1}
4. Special instrument requirements:
Values are listed in the package insert for several analyzers.
I. Device Description:
Liquichek™ Microalbumin Control is prepared from human urine with added constituents of human origin, chemicals, preservatives and stabilizers. The control is provided in liquid form in 10 mL vials for both level 1 and level 2.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Liquichek™ Urine Chemistry Control
2. Predicate K number(s):
k020817
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item or Characteristic | Device | Predicate |
| Form | Liquid, level 1 and 2 | Liquid, level 1 and 2 |
| Matrix | Human urine based | Human urine based |
| Shelf Storage Claim (Unopened) | 2 to 8°C
Until expiration date | 2 to 8°C
Until expiration date |
| Open Vial Claim | 30 days at 2 to 8°C | 30 days at 2 to 8°C |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analytes | Contains:
Microalbumin and creatinine only | Contains:
- Amylase
- Calcium
- Chloride
- Cortisol
- Glucose
- Magnesium
- Osmolality
- pH
- Phosphorus
- Potassium
-Pregnancy
-Protein, Total
-Sodium
- Specific Gravity
- Urea
- Urea Nitrogen
- Uric Acid |
K. Standard/Guidance Document Referenced (if applicable):
None were referenced in the submission.
{2}
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
No information on traceability was provided. The mean values were derived from replicate analysis. The tests listed in the labeling were performed by the reagent manufacturer and/or independent laboratories using manufacturer supported reagents. The sponsor recommends in the labeling that each laboratory establish its own means and acceptable ranges and use those provided only as guides.
Stability studies assayed for both microalbumin and creatinine. Accelerated stability studies were performed to support shelf life of the product. Real time studies are on-going for the life of the product. The sponsor's acceptance criterion is defined as $T_{\text{final}}$ being $\pm 10\%$ $T_{\text{zero}}$.
Shelf life stability: 2 years at 2 to $8^{\circ}$ C.
Open vial stability: 30 days at 2 to $8^{\circ}$ C.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
{3}
Not applicable.
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected mean and assay range for specific analyzers are provided in the labeling. The sponsor recommends in the labeling that each laboratory establish its own means and acceptable ranges and use those provided only as guides.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10 including warning statements for biological source material and to be treated as infectious material with the added caution that it is for professional user.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.