Elecsys® PreciControl Varia 3 is used for quality control of specified Elecsys immunoassays on the Elecsys and cobas e immunoassay analyzers.
Device Story
Elecsys® PreciControl Varia 3 is a multi-analyte, lyophilized quality control product consisting of human serum matrix spiked with specific analytes (Ferritin, Folate, Vitamin B12, β-CTx, Osteocalcin, PTH). It is used in clinical laboratory settings by technicians to monitor the performance of Elecsys and cobas e immunoassay analyzers. The product requires reconstitution with distilled or deionized water before use. By measuring these control samples, healthcare providers verify the accuracy and precision of the immunoassay systems, ensuring reliable diagnostic results for patients. It combines analytes previously found in separate predicate controls (Anemia and Bone panels) into a single product.
Clinical Evidence
No clinical data. Bench testing only. Stability testing confirmed shelf life (15 months unopened) and reconstituted stability (5 hours at 20-25°C, 72 hours at 2-8°C, 31 days at -20°C). Value assignment was performed using median values from eight determinations across multiple analyzers.
Technological Characteristics
Lyophilized human serum matrix; multi-analyte control. Reconstituted with 3.0 mL distilled/deionized water. Analyte traceability via NIBSC standards, radio-binding assays, and in-house reference systems. Intended for use on Elecsys and cobas e immunoassay analyzers.
Indications for Use
Indicated for quality control of specified Elecsys immunoassays on Elecsys and cobas e immunoassay analyzers. No specific patient population or contraindications described.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
K051543 — ELECSYS PRECICONTROL BONE · Roche Diagnostics Corp. · Oct 17, 2005
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k111506
B. Purpose for Submission:
New Device – Combining two previously cleared devices (k082340 and k051543) into one new device
C. Measurand:
Quality Control Material
D. Type of Test:
Not applicable
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
Elecsys PreciControl Varia 3
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JJY
Multi-Analyte Controls, All
Kinds (Assayed) | Class I, reserved | 21 CFR § 862.1660 | Clinical Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
Elecsys PreciControl Varia 3 is used for quality control of specified Elecsys immunoassays on the Elecsys and cobas e immunoassay analyzers. See the package insert for the list of all analytes claimed.
3. Special conditions for use statement(s):
This device is intended for prescription use only.
4. Special instrument requirements:
Elecsys 2010 and cobas e 601 analyzers.
{1}
I. Device Description:
Elecsys PreciControl Varia 3 is a lyophilized product consisting of analytes spiked into human serum matrix at desired concentration levels. Specifically, the kit is comprised of control material for $\beta$ -CrossLaps, Ferritin, Folate, Osteocalcin, Parathyroid Hormone and Vitamin $\mathrm{B}_{12}$ . All products derived from human blood are prepared exclusively from the blood of donors tested individually and shown to be free from HBsAg and antibodies to HCV and HIV. The testing methods applied were FDA-approved or cleared in compliance with the European Directive 98/79/EC, Annex II, List A.
J. Substantial Equivalence Information:
Elecsys PreciControl Varia 3 is a multi-analyte control that combines the analytes in the Elecsys PreciControl Anemia (k082340) and Elecsys PreciControl Bone (k051543) multi-analyte controls.
| Predicate devices name | Predicate 510(k) number |
| --- | --- |
| Elecsys PreciControl Bone | k051543 |
| Elecsys PreciControl Anemia | k082340 |
Comparison with predicate:
| Similarities and Differences | | | |
| --- | --- | --- | --- |
| Item | Proposed Device | Predicate Device (k082340) | Predicate Device (k051543) |
| Intended Use | Same | Used for quality control of specified Elecsys immunoassays. | Used for quality control of specified Elecsys immunoassays. |
| Analyzer | Elecsys 2010, Modular Analytics E170 or Cobas e immunoassay analyzers | Elecsys 1010/2010, Modular Analytics E170 or Cobas e immunoassay analyzers | Elecsys 1010/2010, Modular Analytics E170 or Cobas e immunoassay analyzers |
| Analyte Concentration | β-CTx: ~320 pg/mL β-CTx: ~750 pg/mL | β-CTx: N/A β-CTx: N/A | β-CTx: ~315 pg/mL β-CTx: ~750 pg/mL β-CTx: ~3000 pg/mL |
| | Ferritin: ~14 ng/mL Ferritin: ~150 ng/mL Ferritin: ~1000 ng/mL | Ferritin: ~15 ng/mL Ferritin: ~500 ng/mL Ferritin: ~1500 ng/mL | Ferritin: N/A Ferritin: N/A Ferritin: N/A |
| | Folate: ~3.9 ng/mL Folate: ~12 ng/mL | Folate: ~3 ng/mL Folate: ~8 ng/mL Folate: ~14 ng/mL | Folate: N/A Folate: N/A |
| | Osteocalcin: ~20 ng/mL Osteocalcin: ~100 ng/mL | Osteocalcin: N/A Osteocalcin: N/A | Osteocalcin: ~20 ng/mL Osteocalcin: ~100 ng/mL Osteocalcin: ~205 ng/mL |
| | PTH: ~25 pg/mL PTH: ~60 pg/mL PTH: ~200 pg/mL | PTH: N/A PTH: N/A PTH: N/A | PTH: ~60 pg/mL PTH: ~205 pg/mL PTH: ~850 pg/mL |
{2}
3
| | Vitamin B_{12}: ~230 pg/mL
Vitamin B_{12}: ~500 pg/mL
Vitamin B_{12}: ~1000 pg/mL | Vitamin B_{12}: ~350 pg/mL
Vitamin B_{12}: ~700 pg/mL
Vitamin B_{12}: ~1500 pg/mL | Vitamin B_{12}: N/A
Vitamin B_{12}: N/A
Vitamin B_{12}: N/A |
| --- | --- | --- | --- |
| Format | Same | Same | Lyophilized |
| Volume after Reconstitution | 3.0 mL | 2.0 mL | 2.0 mL |
| Stability | -20°C: 31 days
2–8°C: 72 hours
20-25°C: 5 hours | -20°C: 1 month
2–8°C: 3 days
20-25°C: 5 hours | -20°C: 1 month
2–8°C: 5 days
20-25°C: 8 hours |
K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Staff – Assayed and Unassayed Quality Control Material
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability and Value Assignment:
Elecsys PreciControl Varia 3 assigned values are determined with respective Elecsys assays on the cobas e 601 and Elecsys 2010 analyzers. For each assay, a master calibrator set is available, which is traceable to external or internal reference material. The assigned values for Elecsys PreciControl Varia 3 are read from the master calibration curve. The traceability of the external or internal material used is described in the table below.
| Analyte | Reference Material or Reference Method |
| --- | --- |
| β-CrossLaps [pg/mL] | Gravimetry |
| Ferritin [ng/mL] | 1^{st} International Standard-NIBSC 80/062 |
| Folate [ng/mL] | Elecsys Folate II Assay |
| Osteocalcin [ng/mL] | In-house reference standards:
Commercially available osteocalcin enzyme immuno/radio-binding assay |
{3}
| Parathyroid Hormone [pg/mL] | In-house reference standards: Commercially available PTH radio-binding assay |
| --- | --- |
| Vitamin B12 [pg/mL] | In-house reference standards: Commercially available radio-binding assay |
Value assignment for Elecsys PreciControl Varia on the cobas e 601 analyzer or Elecsys 2010 analyzer was performed in the following manner. The assigned value of each control level was determined by the median value obtained from eight determinations (duplicate runs on four cobas e 601 analyzers or four Elecsys 2010 analyzers). The ranges are determined based upon a specified $\pm$ percentage of the assigned value for each analyte.
**Stability:**
Stability testing protocols (carried out on the cobas e 601 analyzer) and acceptance criteria were described and found to be acceptable. Elecsys PreciControl Varia 3 is stable for 15 months when stored unopened at $2 - 8^{\circ}$ C. Reconstituted Elecsys PreciControl Varia 3 vials are stable for 5 hours on the analyzers at $20 - 25^{\circ}\mathrm{C}$ , 72 hours at $2 - 8^{\circ}\mathrm{C}$ , and 31 days at $-20^{\circ}\mathrm{C}$ .
d. Detection limit: Not applicable
e. Analytical specificity: Not applicable
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device: Not applicable
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
{4}
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: The expected values are provided in the labeling for each analyte for each specific lot.
N. Proposed Labeling: The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.