TRIAGE PROFILER S.O.B. CONTROLS AND S.O.B. CALIBRATION VERIFICATION CONTROLS

K040459 · Biosite Incorporated · JJY · Jul 6, 2004 · Clinical Chemistry

Device Facts

Record IDK040459
Device NameTRIAGE PROFILER S.O.B. CONTROLS AND S.O.B. CALIBRATION VERIFICATION CONTROLS
ApplicantBiosite Incorporated
Product CodeJJY · Clinical Chemistry
Decision DateJul 6, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Triage® Profiler S.O.B. (Shortness of Breath) Calibration Verification Controls are to be used with the Triage® Profiler S.O.B. Panel and Triage® Meter Plus to verify the calibration of the Triage® Profiler S.O.B. Panel throughout the measurable range. The Triage® Profiler S.O.B. (Shortness of Breath) Controls are to be used with the Triage® Profiler S.O.B. Panel and Triage® Meter Plus to assist the laboratory in monitoring test performance.

Device Story

Triage® Profiler S.O.B. Controls are liquid, EDTA plasma-based quality control materials; used with Triage® Profiler S.O.B. Panel and Triage® Meter Plus. Calibration Verification Controls verify panel calibration across measurable range; standard Controls monitor test performance. Used by laboratory personnel to ensure system accuracy for analytes CK-MB, Troponin I, Myoglobin, BNP, and D-dimer. Output provides verification of assay performance; assists clinicians in maintaining diagnostic reliability for patients presenting with shortness of breath.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Liquid quality control material; matrix: EDTA human plasma; contains <0.1% sodium azide; analytes: CK-MB, Troponin I, Myoglobin, BNP, D-dimer; storage: -20°C or colder; form factor: 0.5 mL vials; intended for use with Triage® Meter Plus.

Indications for Use

Indicated for use as quality control material to verify calibration and monitor test performance of the Triage® Profiler S.O.B. Panel on the Triage® Meter Plus in laboratory settings.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k040459 B. Purpose for Submission: New device C. Analyte: CK-MB, Troponin I, Myoglobin, B-type natriuretic peptide (BNP), D-dimer D. Type of Test: Quality Control Material E. Applicant: Biosite Incorporated F. Proprietary and Established Names: Triage® Profiler S.O.B. (Shortness of Breath) Calibration Verification Controls Triage® Profiler S.O.B. (Shortness of Breath) Controls G. Regulatory Information: 1. Regulation section: 21 CFR 862.1660 2. Classification: Class I 3. Product Code: JJY 4. Panel: 75 H. Intended Use: 1. Intended use(s): The Triage® Profiler S.O.B. (Shortness of Breath) Calibration Verification Controls are to be used with the Triage® Profiler S.O.B. Panel and Triage® Meter Plus to verify the calibration of the Triage® Profiler S.O.B. Panel throughout the measurable range. The Triage® Profiler S.O.B. (Shortness of Breath) Controls are to be used with the Triage® Profiler S.O.B. Panel and Triage® Meter Plus to assist the laboratory in monitoring test performance. 2. Indication(s) for use: See Intended Use above {1} Page 2 of 4 3. Special condition for use statement(s): Prescription Use 4. Special instrument Requirements: Triage® Meter Plus I. Device Description: The Triage® Profiler S.O.B. Calibration Verification Controls are supplied as five 0.5 mL vials at levels A, B, C, D, and E and are composed of EDTA human plasma containing &lt; 0.1% sodium azide and CK-MB, Troponin I, Myoglobin, B-type natriuretic peptide (BNP), and D-dimer. The vials should be stored frozen at -20 °C or colder in a non-defrosting freezer and are stable until the date stamped on the box. The Calibration Verification Controls should be thawed at room temperature and mixed by vortexing prior to testing. Once thawed, the Calibration Verification Controls should be tested as soon as possible and should not be re-frozen. It is recommended that each vial be used once and discarded. The concentrations and standard deviations are printed on the enclosed insert. The Triage® Profiler S.O.B. (Shortness of Breath) Controls are supplied as two 0.5 mL vials each of Control 1 and Control 2 and are composed of EDTA human plasma containing &lt; 0.1% sodium azide and CK-MB, Troponin I, Myoglobin, B-type natriuretic peptide (BNP), and D-dimer. The vials should be stored frozen at -20 °C or colder in a non-defrosting freezer and are stable until the date stamped on the box. The controls should be thawed at room temperature and mixed by vortexing prior to testing. Once thawed, the controls should be tested as soon as possible and should not be re-frozen. It is recommended that each vial be used once and discarded. The concentrations and standard deviations are printed on the enclosed insert. J. Substantial Equivalence Information: 1. Predicate device name(s): Triage® Cardio Profiler Calibration Verification Controls Triage® Cardio Profiler Controls 2. Predicate K number(s): k030088, k030089 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | S.O.B. | Cardio Profiler | | | Controls/Calibration | Controls/Calibration | | | Verification Controls | Verification Controls | | Intended Use | Assayed control for monitoring test performance | Same | | Matrix | EDTA plasma | Same | | Form | Liquid | Same | | Storage | -20°C or colder | Same | | Differences | | | | Item | S.O.B. | Cardio Profiler | {2} Page 3 of 4 | | Controls/Calibration Verification Controls | Controls/Calibration Verification Controls | | --- | --- | --- | | Analytes | CK-MB, Troponin I, Myoglobin, BNP, D-dimer | CK-MB, Troponin I, Myoglobin, BNP | K. Standard/Guidance Document Referenced (if applicable): None referenced L. Test Principle: NA M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: NA b. Linearity/assay reportable range: NA c. Traceability (controls, calibrators, or method): The Triage® Profiler S.O.B. (Shortness of Breath) Calibration Verification Controls and Triage® Profiler S.O.B. (Shortness of Breath) Controls have been standardized using purified protein preparations of D-dimer, CK-MB, Troponin I, myoglobin, and BNP based on mass (concentration) of analyte present in EDTA plasma. d. Detection limit: NA e. Analytical specificity: NA f. Assay cut-off: NA 2. Comparison studies: a. Method comparison with predicate device: NA b. Matrix comparison: NA 3. Clinical studies: a. Clinical sensitivity: NA b. Clinical specificity: NA c. Other clinical supportive data (when a and b are not applicable): 4. Clinical cut-off: NA 5. Expected values/Reference range: NA {3} Page 4 of 4 N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...