Multichem IA Plus

K162530 · Techno-Path Manufacturing · JJY · Mar 31, 2017 · Clinical Chemistry

Device Facts

Record IDK162530
Device NameMultichem IA Plus
ApplicantTechno-Path Manufacturing
Product CodeJJY · Clinical Chemistry
Decision DateMar 31, 2017
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

For In Vitro Diagnostic use only. Multichem IA Plus is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

Device Story

Multichem IA Plus is a liquid, frozen quality control serum used in clinical laboratories to monitor the precision of diagnostic testing procedures. It is prepared from human serum supplemented with purified biochemical materials (human/animal origin), chemicals, drugs, stabilizers, and preservatives. The device is provided in three levels to allow performance monitoring across the clinical range. Laboratory technicians use the control by running it alongside patient samples; the resulting values are compared against expected ranges to verify the accuracy and precision of the testing system. This objective assessment helps ensure the reliability of clinical diagnostic results, ultimately supporting accurate patient diagnosis and treatment monitoring. Each donor unit is screened for HIV, HCV, and HBsAg.

Clinical Evidence

No clinical data. Bench testing only. Stability testing (shelf-life and open-vial) performed per CLSI EP25-A. Value assignment conducted using 15 replicates per analyte across multiple runs to establish target means and 3SD ranges.

Technological Characteristics

Quality control material composed of processed human serum with added biochemicals, drugs, stabilizers, and preservatives. Provided in liquid, frozen form. Shelf life: 36 months at -20°C to -80°C. Open-vial stability varies by analyte (1 hour to 10 days at 2-8°C). No electronic components, software, or energy sources.

Indications for Use

Indicated for use as a quality control serum to monitor the precision of laboratory testing procedures for specific analytes in clinical settings. Prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k162530 B. Purpose for Submission: Addition of the following 13 assayed analytes to a previously cleared quality control material (k132091): Acetaminophen, ACTH, Amikacin, Calcitonin, Human Growth Hormone, Lithium, Salicylate, Thyroglobulin, Tobramycin, Free Estriol, EPO, NT Pro-BNP and Troponin T. C. Measurand: Quality control materials for Acetaminophen, ACTH, Amikacin, Calcitonin, Human Growth Hormone, Lithium, Salicylate, Thyroglobulin, Tobramycin, Estriol Free, EPO, NT Pro BNP, Troponin T, Alpha Fetoprotein (AFP), Anti-thyroglobulin, Anti-thyroperoxidase, Human Chorionic Gonadotropin (BhCG), BNP (1-32), CA 125, CA 15-3, CA 19-9, Carbamazepine, Carcinogenic Embryonic Antigen (CEA), CK-MB, Cortisol, C-Peptide, DHEA-Sulfate, Digoxin, Estradiol, Ferritin, Folate, Prostate Specific Antigen Free, Follicle Stimulating Hormone (FSH), Triiodothyronine Free (FT3), Thyroxine Free (FT4), Gentamicin, Homocysteine, Immunoglobulin E, Insulin, Luteinizing Hormone, Myoglobin, Parathyroid Hormone (PTH) (1-84), Phenobarbital, Phenytoin, Progesterone, Prolactin, Prostate Specific Antigen Total, Sex Hormone Binding Globulin, T-Uptake, Testosterone, Theophylline, Troponin I, Thyroid Stimulating Hormone (TSH), Triiodothyronine Total (TT3), Thyroxine Total (TT4), Valproic Acid, Vancomycin, Vitamin B12, 25-OH Vitamin D. D. Type of Test: Not applicable. E. Applicant: Techno-Path Manufacturing Ltd. F. Proprietary and Established Names: Multichem IA Plus {1} G. Regulatory Information: 1. Regulation section: 21 CFR 862.1660 2. Classification: Class I, reserved 3. Product code: JJY 4. Panel: Clinical chemistry (75) H. Intended Use: 1. Intended use(s): See “Indication(s) for use” below 2. Indication(s) for use: For In Vitro Diagnostic use only. Multichem IA Plus is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: The package insert lists the following instruments for which the quality control material is intended: Abbott Architect platforms, Beckman Access II platforms, Roche Cobas platforms and Beckman DXI platforms. I. Device Description: The Multichem IA Plus controls are prepared from human serum with added constituents of human and animal origin, chemicals, therapeutic drugs, stabilizers and preservatives. The control is available in liquid form. Three levels of control are available to allow performance monitoring within the clinical range. {2} Each donor unit used in the preparation of the control material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV, and non-reactive for HBsAg. The following kit configurations are available: Multichem IA Plus Model IA310A (Trilevel) with Level 1, 2 & 3 control; 12 vials with 5 mL contents Model 05P56-10 (Trilevel) with Level 1, 2 & 3 control; 12 vials with 5 mL contents Model IA311A with Level 1 control; 12 vials with 5 mL contents Model IA312A with Level 2 control; 12 vials with 5 mL contents Model IA313A with Level 3 control; 12 vials with 5 mL contents ## J. Substantial Equivalence Information: 1. Predicate device name(s): Multichem IA Plus 2. Predicate 510(k) number(s): k132091 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Characteristics | Multichem IA Plus Assayed Control (k162530, Candidate Device) | Multichem IA Plus Assayed Control (k132091, Predicate Device) | | Intended Use | Multichem IA Plus is intended for use as a quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. | Same | | Form | Liquid, Frozen | Same | | Matrix | Processed Human Serum | Same | {3} | Differences | | | | --- | --- | --- | | Characteristics | Multichem IA Plus Assayed Control (k162530, Candidate Device) | Multichem IA Plus Assayed Control (k132091, Predicate Device) | | Assayed Analytes | Same plus 13 analytes: Acetaminophen, ACTH, Amikacin, Calcitonin, Human Growth Hormone, Lithium, Salicylate, Thyroglobulin, Tobramycin, Free Estriol, EPO, NT Pro-BNP and Troponin T | Alpha Fetoprotein (AFP), Anti-thyroglobulin, Anti-thyroperoxidase, Human Chorionic Gonadotropin (hCG), BNP (1-32), CA 125, CA 15-3, CA 19-9, Carbamazepine, Carcinogenic Embryonic Antigen (CEA), CK-MB, Cortisol, C-Peptide, DHEA-Sulfate, Digoxin, Estradiol, Ferritin, Folate, Prostate Specific Antigen Free, Follicle Stimulating Hormone (FSH), Triiodothyronine Free (FT3), Thyroxine Free (FT4), Gentamicin, Homocysteine, Immunoglobulin E, Insulin, Luteinizing Hormone, Myoglobin, Parathyroid Hormone (PTH) (1-84), Phenobarbital, Phenytoin, Progesterone, Prolactin, Prostate Specific Antigen Total, Sex Hormone Binding Globulin, T-Uptake, Testosterone, Theophylline, Troponin I, Thyroid Stimulating Hormone (TSH), Triiodothyronine Total (TT3), Thyroxine Total (TT4), Valproic Acid, Vancomycin, Vitamin B12, 25-OH Vitamin D. | | Storage Shelf Life (Unopened) | 36 Months @ -20°C to -80°C | 30 Months @ -20°C to -80°C | | Storage Open Vial | 10 days at 2 to 8°C, with the following exceptions: BNP = 1 hour, C-Peptide = 7 days, Folate = 2 days, Homocysteine = 7 days, PTH Intact = 4 days, Troponin I = 4 days, 25OH Vitamin D = 4 days, ACTH = 8 hours, Calcitonin and Troponin T = 7 days at 2 to 8°C. | 10 days at 2 to 8°C, with the following exceptions: BNP = 1 hour, C-Peptide = 7 days, Folate = 2 days, Homocysteine = 7 days, PTH Intact = 4 days, Troponin I = 4 days, and 25OH Vitamin D = 4 days at 2 to 8°C. | {4} K. Standard/Guidance Document Referenced (if applicable): CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents, 2009. L. Test Principle: Not applicable. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable. b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability - The base matrix of the three control levels is processed human serum. The traceability / source of materials used to prepare the three control levels for the 13 new assayed analytes are as listed in the table below: | Assigned Analytes | Source of Material | | --- | --- | | Acetaminophen | Chemical - Therapeutic Drug | | ACTH | Endogenous and spiked with purified human ACTH | | Amikacin | Chemical - Therapeutic Drug | | Calcitonin | Endogenous and spiked with purified human Calcitonin | | EPO | Endogenous and spiked with purified human EPO | | hGH | Endogenous and spiked with purified human or recombinant Growth Hormone | | Free Estriol | Endogenous and spiked with chemical | | Lithium | Chemical - Therapeutic Drug | | NT Pro-BNP | Endogenous and spiked with purified recombinant NT Pro-BNP | | Salicylate | Chemical - Therapeutic Drug | | Thyroglobulin | Endogenous and spiked with purified human Thyroglobulin | | Tobramycin | Chemical - Therapeutic Drug | | Troponin T | Endogenous and spiked with purified human Troponin T | {5} 6 # Stability - Shelf-Life (closed vial) and in-use (open vial) stability testing protocols and acceptance criteria were described and found to be adequate. Real-time stability testing supports a shelf-life stability claim of 36 months when stored frozen at -20 °C to -80°C for the Multichem IA Plus three level controls. Real-Time stability testing supports open vial stability claims as listed below: - Acetaminophen, Amikacin, EPO, hGH, Free Estriol, Lithium, NT Pro-BNP, Salicylate, Thyroglobulin and Tobramycin of 10 days at 2 to 8°C - ACTH of 8 hours at 2 to 8°C - Calcitonin and Troponin T of 7 days at 2 to 8°C # Value Assignment - The value assignment for all analytes was performed on analyte specific assay/analyzer systems using the proposed new value assignment protocol. Fifteen replicates of each analyte were tested over a minimum of three runs with a minimum of three replicates per run using and run over multiple days. The mean from the 15 data points was used as the target concentration. The assignment of value assignment ranges was established based on the 3SD value for each level of the analyte control. The target mean and range for each analyte level along with their respective acceptance criteria were provided for review and found acceptable. Representative values for target mean and range for each analyte for a lot are provided in tabular form in the value assignment sheets. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Not applicable. {6} b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Expected values are provided in the value assignment sheets provided with the package insert. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...