VALIDATE (R) GC3 CALIBRATION VERIFICATION / LINEARITY TEST KIT
Applicant
Maine Standards Company, LLC
Product Code
JJY · Clinical Chemistry
Decision Date
Jul 5, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
VALIDATE® GC3 Calibration Verification / Linearity Test Kit is intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range. See the package insert for the list of all analytes claimed.
Device Story
VALIDATE® GC3 is a quality control test kit containing purified chemicals in a human serum albumin base; used to verify linearity, calibration, and reportable range of automated, semi-automated, and manual chemistry systems. Kit consists of five levels of material with known concentrations to establish the relationship between theoretical and actual analyzer performance. Used by laboratory professionals in clinical settings. Output is a comparison of observed versus target values for specific enzymes, allowing clinicians to confirm instrument accuracy and reportable ranges. Benefits include ensuring reliable diagnostic results for patient testing.
Clinical Evidence
No clinical data. Bench testing only. Sponsor provided stability data for up to 8 months and value assignment data for Cholinesterase (CHE) across five levels, demonstrating target recovery within acceptable limits (max 8.0% difference).
Indicated for in vitro diagnostic use to determine linearity, verify calibration, and verify reportable range for clinical laboratory testing of specified analytes.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
VALIDATE® GC3 Calibration Verification / Linearity Test Kit (k091225)
Related Devices
K023352 — VALIDATE CHEM 9 CALIBRATION VERIFICATION TEST SET, MODEL 109 · Maine Standards Co. · Nov 22, 2002
K091225 — VALIDATE GC1, GC2, GC3 AND GC4 CALIBRATION VERIFICATION/LINEARITY TEST SET · Maine Standards Co. · Jun 25, 2009
K012120 — VALIDATE CHEM 4 CALIBRATION VERIFICATION TEST SET, MODEL 10004 · Maine Standards Co. · Aug 20, 2001
K023410 — VALIDATE CHEM 10 CALIBRATION VERIFICATION TEST SET, MODEL 110 · Maine Standards Co. · Jan 6, 2003
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K110880
B. Purpose for Submission:
Modified Device – Adding Cholinesterase (CHE) to a previously cleared device (k091225)
C. Measurand:
Alkaline Phosphatase (ALP), Alanine Aminotransferase (ALT), Amylase (AMY), Aspartate Aminotransferase (AST), Creatine kinase (CK), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LD) and Lipase (LIP) and Cholinesterase (CHE)
D. Type of Test:
Quality control material
E. Applicant:
Maine Standards Company
F. Proprietary and Established Names:
VALIDATE® GC3 Calibration Verification / Linearity Test Kit
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JJY | Class I, reserved | 21 CFR 862.1660 | Chemistry 75 |
H. Intended Use:
1. Intended use(s):
See Indications for Use below
{1}
2. Indication(s) for use:
VALIDATE® GC3 Calibration Verification / Linearity Test Kit is intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range. See the package insert for the list of all analytes claimed.
3. Special conditions for use statement(s):
For In Vitro Diagnostic Use. For prescription use only.
These tests are not intended for use as routine quality materials or as calibration materials.
4. Special instrument requirements:
The validation and stability studies were conducted using Beckman Coulter AU 680 analyzer.
I. Device Description:
Each VALIDATE® GC 3 Calibration Verification / Linearity Test Kit contains purified chemicals in a human serum albumin base. Five levels are provided to establish the relationship between theoretical and actual performance of each of the included analytes.
Material of human origin used in the manufacture of this test set has been tested using FDA approved methods and found to be non-reactive for HBsAg and antibodies to HCV and HIV-1/2.
J. Substantial Equivalence Information:
1. Predicate device
Maine Standards Company, VALIDATE® GC3 Calibration Verification / Linearity Test Kit
2. Predicate K number
k091225
3. Comparison with predicate
{2}
| | New
VALIDATE® GC3
Calibration Verification /
Linearity Test Kit | Predicate
VALIDATE® GC3
Calibration Verification /
Linearity Test Kit |
| --- | --- | --- |
| Intended Use | For in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated and manual chemistry systems. | Same |
| Analytes | ALP, ALT, AMY, AST, CK, GGT, LD, LIP, CHE | ALP, ALT, AMY, AST, CK, GGT, LD, LIP |
| Matrix | Human serum albumin | Same |
| Number of Levels | 6 including a base matrix | Same |
| Preparation | Liquid, ready to use | Same |
| Packaging | 3.0 mL each level | Same |
| Stability | Until expiration | Same |
| Storage | -10 to -25°C | Same |
K. Standard/Guidance Document Referenced (if applicable):
NCCLS. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. NCCLS document EP6-A [ISBN 1-56238-498-8]. NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2003.
L. Test Principle:
Not Applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not Applicable
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability
{3}
The sponsor provided protocol, acceptance criteria and summary data for stability for up to 8 months that were reviewed and were accepted to be adequate. The real time stability study is ongoing.
## Traceability:
Same as k091225, there are no traceable materials or methods available for the enzymes in GC3.
## Value assignment
Value assignment for the following analytes in the predicate device have already been cleared under k091225
Alkaline Phosphatase (ALP), Alanine Aminotransferase (ALT), Amylase (AMY), Aspartate Aminotransferase (AST), Creatine kinase (CK), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LD) and Lipase (LIP) The data provided by the sponsor indicates that for these analytes, each level is within the claim range and the five levels are linear.
However, since Cholinesterase (CHE) had not been validated previously value assignment for this analyte was provided separately.
The following table lists the Target recovery, the Mean observed, as well as the absolute difference and % difference of the Mean / Target.
| X | Target | Mean | +/- Diff | % Diff |
| --- | --- | --- | --- | --- |
| 1 | 0.97 | 0.94 | 0.03 | 3.1% |
| 2 | 4.37 | 4.44 | 0.07 | 1.6% |
| 3 | 7.78 | 7.74 | 0.04 | 0.5% |
| 4 | 11.18 | 10.74 | 0.44 | 3.9% |
| 5 | 14.58 | 13.41 | 1.17 | 8.0% |
The results provided satisfied the acceptance criteria
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
{4}
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Calibration verification/linearity material is manufactured such that equal distance exists between each consecutive level. The expected values are provided in the labeling for each specific lot.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.