ELECSYS PRECICONTROL ANEMIA

K051517 · Roche Diagnostics Corp. · JJY · Jul 5, 2005 · Clinical Chemistry

Device Facts

Record IDK051517
Device NameELECSYS PRECICONTROL ANEMIA
ApplicantRoche Diagnostics Corp.
Product CodeJJY · Clinical Chemistry
Decision DateJul 5, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Elecsys® PreciControl Anemia is used for quality control of the Elecsys® Ferritin, Folate II, and Vitamin B12 immunoassays on the Elecsys® immunoassay systems.

Device Story

Elecsys® PreciControl Anemia is a lyophilized quality control product; consists of human serum matrix spiked with Ferritin and Aprotinine. Used in clinical laboratory settings to monitor performance of Elecsys® Ferritin, Folate II, and Vitamin B12 immunoassays. Product requires reconstitution with distilled water before use. Healthcare professionals use the control to verify assay accuracy and precision on Elecsys® immunoassay systems; results help ensure reliability of patient diagnostic testing for anemia-related markers.

Clinical Evidence

No clinical data. Performance evaluated via bench testing for value assignment and stability characteristics.

Technological Characteristics

Lyophilized human serum matrix containing ferritin (human origin) and aprotinine (bovine origin). Three levels of control. Reconstituted with distilled water. Stability verified for various storage conditions (refrigerated, frozen, and on-analyzer).

Indications for Use

Indicated for quality control of Elecsys® Ferritin, Folate II, and Vitamin B12 immunoassays on Elecsys® immunoassay systems.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k051517 B. Purpose for Submission: New device C. Measurand: Quality Control Material (assayed for Ferritin, Folate, and Vitamin B12) D. Type of Test: Quality Control Materials for Ferritin, Folate, and Vitamin B12 E. Applicant: Roche Diagnostics F. Proprietary and Established Names: Elecsys® PreciControl Anemia G. Regulatory Information: 1. Regulation section: 21 CFR §862.1660 Quality control material (assayed and unassayed) 2. Classification: Class I 3. Product code: JJX 4. Panel: 75 – Clinical Chemistry {1} 2 H. Intended Use: 1. Intended use: See Indications for use below. 2. Indications for use: Elecsys® PreciControl Anemia is used for quality control of the Elecsys® Ferritin, Folate II, and Vitamin B12 immunoassays on the Elecsys® immunoassay systems. 3. Special conditions for use statement: N/A 4. Special instrument requirements: Elecsys® immunoassay systems I. Device Description: The Elecsys® PreciControl Anemia is a lyophilized product consisting of added ferritin and aprotinine in human serum matrix. During manufacture, the analytes are spiked into the matrix at the desired concentrations. J. Substantial Equivalence Information: 1. Predicate device name: Elecsys® PreciControl MultiAnalyte 2. Predicate 510(k) number: k033937 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Characteristic | Elecsys PreciControl Anemia | Predicate device - Elecsys PreciControl MultiAnalyte | | Intended Use | Elecsys® PreciControl Anemia is used for quality control of the Elecsys® Ferritin, Folate II, and Vitamin B12 immunoassays on the Elecsys® immunoassay systems. | Elecsys® PreciControl MultiAnalyte is used for quality control of the Elecsys® C-Peptide and Elecsys® Insulin immunoassays on the Elecsys® immunoassay systems. | | Levels | Three | Two | | Format | Lyophilized | Same | | Handling | Reconstitute with exactly 2.0 mL distilled water and allow to stand closed for 30 minutes to reconstitute, and then mix gently. | Reconstitute with exactly 2.0 mL distilled water and allow to stand closed for 15 minutes to reconstitute, and then mix gently. | {2} | Similarities | | | | --- | --- | --- | | Characteristic | Elecsys PreciControl Anemia | Predicate device - Elecsys PreciControl MultiAnalyte | | Stability | Unopened: Store at 2 - 8°C until expiration date Reconstituted: • 20 - 25°C: up to 8 hours • on the analyzers at 20 - 25°C: up to 5 hours • at 2-8°C: 3 days • at -20°C: 1 month (freeze only once) • after thawing: use only once | Unopened: Store at 2 - 8°C until expiration date Reconstituted: • on the analyzers at 20 - 25°C: up to 3 hours • at -20°C: 1 month (freeze only once) • after thawing: use only once | | Differences | | | | --- | --- | --- | | Characteristic | Elecsys PreciControl Anemia | Predicate device Elecsys PreciControl MultiAnalyte | | Matrix | Human serum with added Ferritin (human origin) and Aprotinine (bovine origin) | Equine serum with added C-Peptide and insulin | # K. Standard/Guidance Document Referenced (if applicable): None referenced # L. Test Principle: Not applicable # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: N/A b. Linearity/assay reportable range: N/A c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: Roche Diagnostics maintains a set of master calibrators for each assay that have values assigned that are traceable to the various method "gold standards". These standards can either be WHO, NIST or in-house developed {3} standards. The Elecsys® PreciControl products are assayed and compared to these reference preparations and target values and ranges are assigned. Values are assigned using Elecsys® MODULAR ANALYTICS E170, Elecsys® 2010 and Elecsys® 1010 Immunoassay Analyzers. Two independent series of analyses are performed on each instrument for each analyte. Each sample is tested in duplicate. The target value is calculated as the median of the determined values. Stability: Five studies were performed to verify the stability claims for Elecsys® PreciControl Anemia. The acceptance criterion for Ferritin and Vitamin B12 is recovery of 90%-110% of the concentration based on unstressed material. The acceptance criterion for Folate II is for samples with concentrations of ≤5 ng/mL, a SD ≤0.5 ng/mL and for concentrations >5 ng/mL is recovery of +/- 10%. | Stability | Unopened | Reconstituted | | --- | --- | --- | | Elecsys PreciControl Anemia | Store at 2 - 8°C until expiration date | °20 - 25°C: up to 8 hours °on the analyzers at 20 - 25°C: up to 5 hours °at 2-8°C: 3 days °at -20°C: 1 month (freeze only once) °after thawing: use only once | d. Detection limit: N/A e. Analytical specificity: N/A f. Assay cut-off: N/A 2. Comparison studies: a. Method comparison with predicate device: N/A b. Matrix comparison: N/A {4} 5 3. Clinical studies: a. Clinical Sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: Lot specific target values for Elecsys® 2010 analyzer are presented in the following table: | | PC Anemia I Low | PC Anemia II Medium | PC Anemia III High | | --- | --- | --- | --- | | Folate II (ng/mL) | 2.89 | 7.74 | 16.2 | | Ferritin (ng/mL) | 11.9 | 431 | 1320 | | Vitamin B12 (pg/mL) | 287 | 773 | 1610 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...