MODIFICATION TO ROCHE DIAGNOSTICS PRECINORM PROTEINS IN URINE/CSF (PUC) AND PRECIPATH PROTEINS IN URINE/CSF (PUC)
Applicant
Roche Diagnostics Corp.
Product Code
JJY · Clinical Chemistry
Decision Date
Jul 27, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
Precinorm ® PUC/ Precipath ® PUC is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed value sheet
Device Story
Precinorm® PUC and Precipath® PUC are liquid, ready-to-use quality control materials based on buffered aqueous solutions. They are used in clinical laboratory settings to monitor the accuracy and precision of quantitative analytical methods on Roche/Hitachi MODULAR and COBAS INTEGRA analyzer systems. The modification involved adding Immunoglobulin A (IgA) and Immunoglobulin M (IgM) constituents to the Precipath® PUC formulation and updating the associated value sheets. The device functions as a control sample; healthcare providers compare analyzer results against assigned values to verify system performance and ensure reliable clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Liquid, ready-to-use control based on buffered aqueous solution. Analyte concentrations adjusted for normal and pathological ranges. Stability: Unopened at 2-8 °C until expiration; opened at 2-8 °C for 4 weeks. Multi-analyte control (Class I, 21 CFR 862.1660).
Indications for Use
Indicated for use as a quality control material to monitor the accuracy and precision of quantitative laboratory methods for proteins in urine and cerebrospinal fluid (CSF).
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Precinorm® Universal and Precipath® Universal Control Sera (K040280)
Related Devices
K040280 — PRECINORM PROTEINS IN URINE/CSF PRECIPATH PROTEINS IN URINE/CSR CONTROLS · Roche Diagnostics Corp. · Mar 11, 2004
K050026 — PRECINORM PROTEINS IN URINE/CSF (PUC) AND PRECIPATH PUC · Roche Diagnostics · Jan 31, 2005
K042389 — MODIFICATION TO PRECINORM UNIVERSAL PLUS AND PRECIPATH UNIVERSAL PLUS CONTROL SERA · Roche Diagnostics Corp. · Sep 14, 2004
Submission Summary (Full Text)
{0}
1
# SPECIAL 510(k): Device Modification
## ODE Review Memorandum
To: THE FILE
Re: K 041812 Roche Precinorm® PUC and Precipath® PUC
Date: July 21,2004
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was to: Add Immunoglobulin A and Immunoglobulin M constituents to Precipath® PUC. Modify corresponding labeling changes by the (addition of IgA and IgM values to value sheet), and edit labeling by removal of asterisks and footnotes from value sheet.
Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics.
## Device description and comparison
| Characteristic | Precinorm® PUC / Precipath® PUC (Predicate Device, K040280) | Precinorm® PUC / Precipath® PUC (Modified Device, K041812) |
| --- | --- | --- |
| Intended Use | For use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed value sheet. | Same |
| Format | Liquid ready-for-use control based on a buffered aqueous solution. Concentrations fo control components have been adjusted to represent normal and pathological ranges. | Same |
| Stability | Unopened | Same |
{1}
2
| | Stable at 2-8 °C until expiration date
**Opened**
Stable at 2-8 °C for 4 weeks. | |
| --- | --- | --- |
| Constituent analytes with Assigned Values | **Precinorm**
• Albumin
• Creatinine
• Total Protein | **Precinorm**
• Albumin
• Creatinine
• Total Protein |
| | **Precipath**
• Albumin
• Creatinine
• Total Protein | **Precipath**
• Albumin
• Creatinine
• Total Protein
• Immunoglobulin A
• Immunoglobulin M |
4. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
5. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Reviewer Comments:
{2}
3
The Roche Precinorm® PUC and Precipath® PUC Controls have been modified to Apply these 510(k) cleared reagents to Roche / Hitachi MODULAR analyzers and COBAS INTEGRA analyzer systems in accordance with Roche Diagnostics replacement reagent policy.
Modification of the Precinorm® PUC and Precipath® PUC Controls does not affect the intended use of indications for use as described in the labeling, nor does it alter the fundamental scientific technology of the device.
## Regulatory Classification
MULTI-ANALYTE CONTROLS, ALL KINDS (ASSAYED AND UNASSAYED)
862.1660
JJY
I
Panel 75
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.