ABC LINEARITY VERIFIERS

K973790 · Sc Calibrators & Controls, LLC · JJY · Oct 14, 1997 · Clinical Chemistry

Device Facts

Record IDK973790
Device NameABC LINEARITY VERIFIERS
ApplicantSc Calibrators & Controls, LLC
Product CodeJJY · Clinical Chemistry
Decision DateOct 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1
Attributes3rd-Party Reviewed

Indications for Use

ABC Linearity Verifiers is intended to be used to verify the calibration, linear operating range and reportable range (linearity) of methods used to determine the concentration of selected therapeutic drugs.

Device Story

ABC Linearity Verifiers are a five-level set of human serum-based controls containing 12 therapeutic drug analytes. Levels 1-4 are related by linear dilution; level 1 is a zero-level; level 5 is near the instrument's upper limit. Used in clinical laboratories by technicians or laboratory staff to confirm calibration and linear operating range of automated and manual TDM methods. The verifiers are assayed like patient samples; results assist in verifying instrument performance and reportable ranges. This process ensures accurate therapeutic drug concentration measurements, supporting clinical decision-making for drug dosing and monitoring.

Clinical Evidence

Bench testing only. Performance validated on two commonly used automated therapeutic drug monitoring systems (Abbott TDx Assay System and Behring EMIT system) and HPLC methods for selected analytes. Testing confirmed the device's ability to verify calibration and linear operating range across the intended analytical range.

Technological Characteristics

Five-level set of human serum-based verifiers containing 12 therapeutic drug analytes. Gravimetric and traceable calibration. Non-azide preservative. Stability: 60 days at 2-8°C or until expiration if frozen. Compatible with automated and manual immunoassay methods.

Indications for Use

Indicated for use in clinical laboratories to verify calibration, linear operating range, and reportable range of therapeutic drug monitoring (TDM) methods and instruments. Required for CLIA-regulated moderately complex, highly complex, and in-house methods at least twice annually, or upon reagent lot changes, major instrument maintenance, or when control values exhibit trends/shifts.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SC Calibrators & Controls LLC, 1701 Berkeley Street, Santa Monica CA 90404 Tel: (310) 828-7423 Fax: (310) 453-1666 Image /page/0/Picture/1 description: The image shows a black and white photograph of a person's face. The person's face is partially obscured by a white circle in the upper left corner of the image. The person's facial features are not clearly visible, but it appears that they have a light complexion. The background of the image is black. OCT 1 4 1997 ### 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ Contact Name: Rov F. Schall, Jr., Ph.D. Date: August 31. 1997 Product Name: ABCtm Linearity Verifiers Common Names:Calibration Verifiers, Linearity Verifiers Document TDM | Linearity Set. Casco Standards. Inc. Predicate Devices: LiniCAL Protein I Calibration Verifiers, International Enzymes, Inc. ABC Linearity Verifiers is an in vitro diagnostic med-Description of the Device: ical device intended for use with automated and manual methods monitoring selected therapeutic drugs (TDM). ABC Linearity Verifiers is a five level set of verifiers, with each level containing 12 analytes, is made with human serum and is used to confirm the proper callbration, linear operating range and reportable range (linearity) of TDM methods. The "ABC" designation derives from Antibiotic and Cardiac drugs. ABC Linearity Verifiers contains concentrations of analytes extending over a wide analytical range. Level 1 is a zero level, and level 5 has concentrations near the upper limit of instruments. Levels 1, 2, 3, and 4 are related by linear dilution. When assayed like patient samples, the verifiers assist in determination of calibration and linear operating range (linearity) of methods for the analytes included. (A second product, ABC Calibration Verifiers, is identical in every respect to this product except that the concentrations of levels 2, 3, and 4 aretargeted to regions of clinical interest. Both products will perform linear range and calibration verification, and their composition, stabilities and performance are identical except for concentration of middle levels.) ABC Linearity Verifiers is intended to be used to verify the calibra-Intended Use: tion, linear operating range and reportable range (linearity) of methods used to determine the concentration of selected therapeutic drugs. {1}------------------------------------------------ # Technical Characteristics Compared to Predicate Devices: ## Comparison of Features of ABC Linearity Verifiers and the two predicate devices: DOCUMENT TDM I Linearity Test Set and LiniCAL Protein I Callbration Verifiers. | Attribute | ABC Linearity Verifiers | DOCUMENT TDM I | LiniCAL Protein I | |--------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------| | No Analytes/bottle | 12 | 19 | 6 | | No of Levels/set | 5 | 8 | 5 | | Vol per level (mL) | 5 | 5 | 1 | | Type of Analytes | Therapeutic Drugs | Therapeutic Drugs | Serum Proteins | | Method(s) | Automated, Manual<br>Immunoassay, other | Automated<br>Immunoassay | Beckman Array<br>Nephelometer | | Base Matrix | Human Serum | Human Serum | Human Serum | | Preservative | Non-Azide | Azide | Azide | | Calibration | Gravimetric + Some<br>Traceable | Assayed<br>(Two Methods) | Assayed + Traceable | | Use | Verification of Calibration,<br>Linear Operating Range (OR),<br>Reportable Range<br>All Methods | Verif. of Calib,<br>Linear OR,<br>Reportable<br>Range for<br>Two Methods | Verif. of Calib,<br>Linear OR<br>Beckman Array | | Stablilty | 60 days at 2-8° C; | To Expiration | To Expiration Date | | | To Expir. Date if frozen | at 2-8° C | unopened at 2-8°C | | | (Vancomycin 30 days 2-8°C) | | 14 days opened | Technical Characteristics by Assessment of Performance: The performance of ABC Verifiers has been tested on two commonly used automated therapeutic drug monitoring systems (Abbott TDx Assay System and Behring (formerly Syva) EMIT system reagents) and by HPLC methods for selected analytes to validate their use. All methods tested demonstrated the desired functionality of the product. Based upon the purpose of the device, the descriptions and label-Conclusions: ing of the predicate devices, and upon the safety and efficacy using multiple instruments and methods, and stability data generated for the ABC Verifiers, the product is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle. In the center of the logo is a stylized symbol that resembles an abstract caduceus or a person embracing another, with flowing lines representing movement or connection. ## OCT 14 1997 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SC Calibrators & Controls LLC Ms. Carole Stamp c/o TUV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891 Re : K973790 Trade Name: ABC™ Linearity Verifiers Requlatory Class: I Product Code: JJY September 19, 1997 Dated: Received: October 6, 1997 Dear Ms. Stamp: We have reviewed your Section 510(k) notification of intent wo market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ Page 2 - Ms. Stamp Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655 . This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours, Steven Toutman Steven I. Gutman, M.D. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 12.0 The Office of Device Evaluation (ODE) has developed the attached optional form to assist them with instituting the requirement for all original 510(k)s received by ODE on or after 1/2/96. The requirement is for all 510(k) submissions to have clearly defined "Indications for Use". These indications will be attached by ODE to any substantial equivalence (SE) letter to define what the device is cleared for. No 510(k) submitted on or after 1/2/96 will be cleared for marketing by ODE without the inclusion of the indications for use information, which will be attached to an SE letter. INDICATIONS FOR USE STATEMENT 510(k) Number (if known): ___________ Linearity Verifiers Device Name: _A B C "" Indications For Use: #### Indications For Use ABC Linearity Verifiers should be used any time it is necessary to confirm the proper calibration ind linear operating range of TDM methods and instruments. The Clinical Laboratory Improvenent Act and its regulations (CLIA) require verification of timearity and calibration twice a year for modified moderately complex methods, highly complex methods and in-house methods. Verification should be performed more often if any of the following occurs: introduction of procedures for which control values change with a new reagent lot, major preventative maintenance or replacement of critical parts on the instrument, when control values exhibit an unusual trend or shift or are outside the acceptable limits of the laboratory, and if the laboratory's schedule of verification is more frequent. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IN NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K973790 | OR Over-The-Counter Use Description Use (Per 21 CFR 801.109)(Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...