AUDIT MICROCV THERAPEUTIC DRUG (TDM) LINEARITY SET
Applicant
Aalto Scientific, Ltd.
Product Code
JJY · Clinical Chemistry
Decision Date
Apr 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is an assayed quality control material consisting of five levels of human based serum. Each level contains: Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin. These five levels demonstrate a linear relationship to each other for their respective analytes. It is intended to simulate human patient serum samples for purpose of determining linearity, calibration verification of reportable range for Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin. The product is intended for use with quantitative assays on the indicated analyzer provided in the labeling and may be used as quality control material for these analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set should not be used for calibration or standardization of the Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin assays. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is "For In Vitro Diagnostic Use Only".
Device Story
Audit® MicroCV™ TDM Linearity Set is an in-vitro diagnostic quality control material; consists of five levels of lyophilized human-based serum containing 15 therapeutic drug analytes. Used in clinical laboratories to verify linearity and reportable range of quantitative assays. Device does not perform analysis; rather, it serves as a reference material for laboratory instrumentation. Laboratory personnel process the five levels through their existing clinical chemistry analyzers; results are plotted to confirm linear response across the assay range. Benefits include ensuring instrument accuracy and reliability for therapeutic drug monitoring, helping clinicians maintain confidence in patient test results. Materials are tested for HbsAg, HCV, and HIV-1/2.
Clinical Evidence
No clinical data. Bench testing only. Stability studies confirmed 24-month shelf life and 7-day open-vial stability at 2-8°C. Linearity established via 5-point regression analysis on Abbott Axsym and Roche Hitachi 911 instruments, requiring a linear regression value > 0.90.
Technological Characteristics
Lyophilized human-based serum; contains 15 analytes; preserved with sodium azide. 5-level configuration (5 x 5 mL). Storage 2-8°C. Non-sterile. Analyte value assignment performed via quantitative clinical chemistry analyzers (Abbott Axsym, Roche Hitachi 911).
Indications for Use
Indicated for use as an assayed quality control material to simulate human serum for determining linearity, calibration verification, and verification of reportable range for 15 specific therapeutic drugs (Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Phenytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valproic Acid, Vancomycin) on quantitative clinical chemistry analyzers. Not for use as a calibrator or standard.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Related Devices
K082714 — AUDIT MICROCV THERAPEUTIC DRUG (TDM) LINEARITY SET, MODEL K707M-5 · Aalto Scientific, Ltd. · Nov 19, 2008
K033879 — VALIDATE TDM CALIBRATION VERIFICATION TEST SET, MODEL 126 · Maine Standards Co. · Feb 27, 2004
K042318 — AUDIT MICROCV GENERAL CHEMISTRY LINEARITY SET · Aalto Scientific, Ltd. · Dec 9, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summarv
## A. Submitter
Aalto Scientific, Ltd. 1959 Kellogg Ave. Carlsbad, CA 92008 Telephone: (760) 431-7922 Fax: (760) 431-6824
### Contact Person
Dessi Lyakov Regulatory Affairs Manager Telephone: (760) 431-7922 Ext. 118 E-mail: dlyakov@aaltoscientific.com
### Date of Summary Preparation
April 19, 2013
### B. Device Identification
| Product Trade Name: | Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set |
|------------------------|------------------------------------------------------|
| Common Name: | Audit® MicroCV™ Therapeutic Drug (TDM) Linearity |
| Review Panel: | Chemistry 75 |
| Classification Names: | Assay QC Material |
| Device Classification: | Class I, reserved |
| Product Code: | JJY |
| Regulation Number: | 21 CFR 862.1660 |
## C. Device to Which Substantial Equivalence is Claimed
Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set Aalto Scientific, Ltd. Carlsbad, California
510(k) Number: K082714
### D. Device Description:
The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is an in-vitro diagnostic device consisting of five levels of Lyophilized linearity material containing Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin and additives in human serum. There are five vials labeled A, B, C, D, and E, and contain 5 mL for each level.
Materials of human origin used in the manufacture of this linearity set has been tested using FDA approved methods and found to be non-reactive for HbsAg and antibodies to HCV and HIV-1/2.
| Levels | Catalog Number | Configuration<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|----------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| י מחירות מש<br>1. U. V. V. | .<br>K707M-F | 5 x 5 mL |
APR 2 9 2013
k130762
{1}------------------------------------------------
## E. Value Assignment:
All analytes but Lithium value assignment for Levels A through Level E is performed on one Abbatt Axsym instrument by using Abbott reagents. Lithium value assignment for Levels A through Level E is performed on one Roche Hitachi 911 instrument by using Pointe Scientific reagent. Each analyte was measured 5 times (5 separate vials) and the mean value of each analyte was used to establish target concentration values at each level. The target ranges were calculated as ± 15% of the target value. The mean concentration values of each level were plotted (concentration value vs. assigned level) and a linear regression value was obtained. If the five-point linear regression value is greater than 0.90 and if the plots are linear then the products demonstrate linearity.
## F. Intended Use:
The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is an assayed quality control material consisting of five levels of human based serum. Each level contains: Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin. These five levels demonstrate a linear relationship to each other for their respective analytes. It is intended to simulate human patient serum samples for purpose of determining linearity, calibration verification of reportable range for Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin.
The product is intended for use with quantitative assays on the indicated analyzer provided in the labeling and may be used as quality control material for these analytes. When used for quality control purposes, it is recommended that each laboratory establish its own
means and acceptable ranges and use the values provided only as guides. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set should not be used for calibration or standardization of the Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline,
Tobramycin, Valporic Acid and Vancomycin assays. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is "For In Vitro Diagnostic Use Only".
## G. Comparison with predicate
| Characteristics | Audit® MicroCV™ Therapeutic Drug<br>(TDM) Linearity Set<br>(Modified) | Audit® MicroCV™ Therapeutic Drug<br>(TDM) Linearity Set<br>(Unmodified-K082714) |
|-----------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Similarities | | |
| Intended Use | Linear, calibration verification quality<br>control material | Linear, calibration verification quality<br>control material |
| Number of levels<br>per set | 5 | 5 |
| Contents | 5 x 5 mL | 5 x 5 mL |
Similarities and differences between new and predicate devices
{2}------------------------------------------------
| Matrix | Human Based Serum | Human Based Serum |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of Analytes | Clinical Chemistry | Clinical Chemistry |
| Form | Lyophilized | Lyophilized |
| Preservatives | Sodium Azide | Sodium Azide |
| Storage | 2 to 8º C<br>for 24 months | 2 to 8º C<br>for 24 months |
| Open Vial<br>Stability | 7 days at 2 to 8º C | 7 days at 2 to 8º C |
| Sterile | No | No |
| | Differences | |
| Number of<br>Analytes per vial | 15 | 14 |
| Analytes | Acetaminophen, Amikacin,<br>Carbamazepine, Digoxin,<br>Gentamicin, Lithium, Methotrexate,<br>Phenobarbitol, Pheytoin, Quinidine,<br>Salicylate, Theophylline,<br>Tobramycin, Valporic Acid and<br>Vancomycin | Acetaminophen, Amikacin,<br>Carbamazepine, Digoxin,<br>Gentamicin, Methotrexate,<br>Phenobarbitol, Pheytoin,<br>Quinidine, Salicylate,<br>Theophylline, Tobramycin, Valporic<br>Acid and Vancomycin |
# H. Statement of Supporting data:
Stability studies have been performed to determine the open vial stability and shelf life for the Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set. All supporting data is retained on file at Aalto Scientific, Ltd. Product claims are as follows: Open Vial Stability: 7 days at 2 - 8º C. Shelf-life Stability: 24 months at 2 - 8º C.
## l. Conclusion:
Based upon the performance characteristics indicated above, Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is substantially equivalent to the predicate device K082714.
All supporting data is retained on file at Aalto Scientific, Ltd.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three curved lines that resemble a human figure in motion or a flowing ribbon-like design.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2013
Aalto Scientific, Ltd. C/O Dessi Lyakov 1959 Kellogg Ave. CARLSBAD CA 92008
Re: K130762_
Trade/Device Name: Audit™ MicroCV™ Therapeutic Drug (TDM) Linearity Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: I, reserved Product Code: JJY Dated: March 19, 2013 Received: March 29, 2013
Dear Dessi Lyakov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for
{4}------------------------------------------------
Page 2- Dessi Lyakov
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. .
Sincerely yours,
#### Carol C. Benson -S for
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K130762
## Device Name: Audit® MicroCV™TherapeuticDrug (TDM)LinearitySet
#### Indications For Use:
The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is an assayed quality control material consisting of five levels of human based serum. Each level contains: Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Cheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin. These five levels demonstrate a linear relationship to each other for their respective analytes. It is intended to simulate human patient serum samples for purpose of determining linearity, calibration verification and verification of reportable range for Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin.
The product is intended for use with quantitative assays on the indicated analyzer provided in the labeling and may be used as quality control material for these analytes. When used for quality control purposes, it is recommended that each laboratory establish its own means and
acceptable ranges and use the values provided only as guides. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set should not be used for calibration or standardization of the Acetaminophen, Amikacin, Carbamazepine, Digoxin, Gentamicin, Lithium, Methotrexate, Phenobarbitol, Pheytoin, Quinidine, Salicylate, Theophylline, Tobramycin, Valporic Acid and Vancomycin assays. The Audit® MicroCV™ Therapeutic Drug (TDM) Linearity Set is "For In Vitro Diagnostic Use Only".
Prescription Use X AND/OR
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)
Yung W. Chan -S
Division Sign - Off Office of In Vitro Devices and Radiologic Health
510(k): K130762
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.