K051928 · Radiometer Medical Aps · JJY · Oct 12, 2005 · Clinical Chemistry
Device Facts
Record ID
K051928
Device Name
AUTOCHECK6+
Applicant
Radiometer Medical Aps
Product Code
JJY · Clinical Chemistry
Decision Date
Oct 12, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The AutoCheck6+ is a liquid four ampoule quality control system for checking the precision and accuracy of Radiometer Medical ApS analyzers for pH/Blood gases, co-oximetry, electrolytes, bilirubin, glucose, lactate, and creatinine.
Device Story
AutoCheck6+ is a liquid quality control system consisting of four levels of aqueous solutions packaged in ampoules. It is used to verify the precision and accuracy of Radiometer Medical ApS analyzers. The system contains components for pH, blood gases, co-oximetry, electrolytes, bilirubin, glucose, lactate, and creatinine. It includes added creatine, creatinine, and the enzyme creatininase to stabilize the equilibrium between creatine and creatinine. The device is used in clinical laboratory settings where Radiometer analyzers are operated by trained personnel. By measuring these control solutions, clinicians assess the performance of the analyzer to ensure reliable patient test results.
Clinical Evidence
No clinical data. Bench testing only. Stability established via real-time and accelerated studies (storage at 6°C, 10°C, 25°C, 32°C, and 40°C). Traceability to NIST SRM 914 for creatinine and other chemistry standards for remaining parameters.
Technological Characteristics
Liquid quality control system consisting of four levels of aqueous solutions. Contains pH/blood gas, co-oximetry, electrolyte, bilirubin, glucose, lactate, and creatinine components. Includes creatininase (creatinine amidohydrolase) for stabilization. Analyzed using standard Radiometer analyzer measurement protocols.
Indications for Use
Indicated for use as a quality control material to verify the precision and accuracy of Radiometer Medical ApS analyzers measuring pH, blood gases, co-oximetry, electrolytes, bilirubin, glucose, lactate, and creatinine.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
AutoCheck5+ (part of K992859 - ABL7000 with AutoCheck Module)
Related Devices
K980135 — QUALICHECK5+ MODELS S7730, S7740, S7750, S7760 · Radiometer America, Inc. · Jan 28, 1998
K961354 — QUALICHECK 4 METABOLITE(S3630,S3640,S3650,S3660) · Radiometer America, Inc. · Jun 14, 1996
K972868 — TRADEMARK BLOOD GAS/ELECTROLYTE/GLUCOSE, LACTATE, BUN AND CREATININE CONTROL · Bionostics, Inc. · Aug 28, 1997
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k051928
B. Purpose for Submission:
New device (control)
C. Measurand:
Control for pH/Blood gases, co-oximetry, electrolytes, glucose, lactate and creatinine
D. Type of Test:
Control material
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
AutoCheck6+
G. Regulatory Information:
1. Regulation section:
21 CFR Section 862.1660, Quality control material (assayed and unassayed)
2. Classification:
Class I
3. Product code:
JJY, control, Multi-analyte mixture
4. Panel:
Chemistry (75)
{1}
H. Intended Use:
1. Intended use(s):
The AutoCheck6+ is a liquid four ampoule quality control system for checking the precision and accuracy of Radiometer Medical ApS analyzers for pH/Blood gases, co-oximetry, electrolytes, bilirubin, glucose, lactate, and creatinine.
2. Indication(s) for use:
The AutoCheck6+ is a liquid four ampoule quality control system for checking the precision and accuracy of Radiometer Medical ApS analyzers for pH/Blood gases, co-oximetry, electrolytes, bilirubin, glucose, lactate, and creatinine.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
For use with Radiometer analyzers
I. Device Description:
The AutoCheck6+ quality control system includes control material at four different levels (Levels 1-4). The controls are aqueous and consist of a buffer with salts, metabolites, enzymes, dyes and preservatives, equilibrated with carbon dioxide and oxygen. The controls do not contain any animal or biological components.
The controls are supplied in 0.7 mL ampoules, 30 ampoules per box of each control level. The ampoules are used only once after being opened.
J. Substantial Equivalence Information:
1. Predicate device name(s):
AutoCheck5+
2. Predicate 510(k) number(s):
k992859
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | AutoCheck6+ | AutoCheck5+ |
| Intended use | same | For use in checking the precision and accuracy of Radiometer analyzers. |
| Format | same | Liquid control contained in ampoules. |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test parameters | pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil and the addition of cCrea (creatinine) | pH, pCO2, pO2, ctHb, sO2, FO2Hb, FCOHb, FMetHb, FHbF, cK+, cNa+, cCa2+, cCl-, cGlu, cLac, ctBil |
K. Standard/Guidance Document Referenced (if applicable):
N/A
L. Test Principle:
N/A
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
N/A
b. Linearity/assay reportable range:
N/A
{3}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The different parameters contained within AutoCheck6+ are traceable to various chemistry standards, with most parameters traceable to a NIST SRM. A Certificate of Traceability was provided for each level of control. The added parameter, creatinine, is traceable to NIST SRM 914.
The control ranges are determined by using an established protocol that includes performing runs on six different analyzers and with several different operators over a number of days. The final ranges for each level of the control are lot specific and found on the Control Ranges data sheets provided with the product. Additionally, the controls can only be used on the Radiometer analyzers listed on the data sheets.
Stability: The stability of the closed ampoules was established using both real-time and accelerated stability studies. Only closed container stability was performed since the control is used only one time after the ampoule is opened. Since the stability in the labeling is stated to be 24 months at 2-8°C, 4 weeks at up to 25°C, or 15 days for up to 32°C, several different stability studies are performed. These included storage at 6°C, 10°C, and 25°C for 103 weeks as well as storage at 32°C and 40°C for 39 weeks and 26 weeks respectively. Upon testing at the various time points, the measured results must fall within the preset acceptable performance limits for each parameter.
d. Detection limit:
N/A
e. Analytical specificity:
N/A
f. Assay cut-off:
N/A
2. Comparison studies:
a. Method comparison with predicate device:
N/A
b. Matrix comparison:
N/A
{4}
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
N/A
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports substantial equivalence decision.
5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.