SC9000/SC9015 MEDICAL INFORMATION BUS (MIS) PROTOCOL CONVERTER
K970368 · Siemens Medical Solutions USA, Inc. · DQA · May 6, 1997 · Cardiovascular
Device Facts
Record ID
K970368
Device Name
SC9000/SC9015 MEDICAL INFORMATION BUS (MIS) PROTOCOL CONVERTER
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 6, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display.
Device Story
Device acts as interface/protocol converter; transforms data from third-party medical devices (ventilators, infusion pumps, SvO2 monitors) into IEEE 1073 Medical Information Bus (MIB) format. Inputs include serial RS-232 or analog signals from third-party equipment. Output is IEEE 1073-compliant data transmitted to Siemens SC9000/SC9015 bedside monitors for display. Used in OR or ICU environments by clinicians. Facilitates centralized monitoring of disparate medical equipment data on a single display, aiding clinical decision-making by integrating patient data from multiple sources.
Indicated for connecting third-party medical devices (e.g., ventilators, SvO2 devices, infusion pumps) to Siemens SC9000/SC9015 bedside monitors. No specific patient population or contraindications defined; device is not connected to patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Hewlett-Packard Model M1032A Vuelink Interface Plug In Module (K923682)
Related Devices
K973222 — SC9000/SC9015 MEDICAL INFORMATION BUS (MIB) PROTOCOL CONVERTER · Siemens Medical Solutions USA, Inc. · Sep 4, 1997
K012461 — SIEMENS INFINITY MIB II DUO PROTOCOL CONVERTER · Siemens Medical Solutions USA, Inc. · Aug 17, 2001
K022766 — SIEMENS MEDICAL INFORMATION BUS (MIB/MIB II/MIB DUO) PROTOCOL COVERTER · Siemens Medical Solutions USA, Inc. · Sep 5, 2002
K020277 — SIEMENS MEDICAL INFROMATION BUS (MIB II) PROTOCOL CONVERTER · Siemens Medical Solutions USA, Inc. · Feb 19, 2002
K050974 — DRAEGER MEDICAL INFORMATION BUS (MIB, MIB II, MIB DUO) PROTOCOL CONVERTERS · Draeger Medical Systems, Inc. · Apr 29, 2005
Submission Summary (Full Text)
{0}
510 (K) Notification
Siemens SC9000/SC9015 Medical Information Bus (MIB) Protocol Converter
K970368
# 510(k) SUMMARY
as required per 807.92(c)
MAY - 6 1997
## 2: Submitters Name, Address:
Siemens Medical Systems, Inc.
16 Electronics Avenue
Danvers, MA 01923
Tel: (508) 750-7500
Fax: (508) 777-3398
Official Correspondent: David Simard, Director Quality Assurance and Regulatory Affairs
Contact person for this submission: Jacqueline E. M. Emery
Date submission was prepared: January 7, 1997
## 3: Trade Name, Common Name and Classification Name:
A. Trade Name: Siemens SC9000/ SC9015 Series Medical Information Bus (MIB) Protocol Converter
B. Common Name, Classification Number, Class and Regulation Number:
| Common Name | Classification Number | Class | Regulation Number |
| --- | --- | --- | --- |
| Transducer Signal amplifier and conditioner | 73 DQA | II | 21CFR870.2060 |
## 4: Predicate Device Identification: Hewlett-Packard Model M1032A Vuelink Interface Plug In Module cleared under 510(K) number: K923682
## 5. Device Description: The Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter transforms data generated by third party devices (e.g. ventilators, SvO2 devices and infusion pumps) into a common format per the IEEE 1073 Medical Information Protocol and transmits this data to the Siemens SC9000/SC9015 Bedside Monitor. Attachment cables specific to third party device are offered as accessories.
## 6. Intended Use: The Medical Information Bus (MIB) Protocol Converter is intended to connect third party medical devices that do not provide data per the IEEE 1073 Medical Information Protocol Standard to the Siemens SC9000/SC9015 Bedside Monitor for display.
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
{1}
510 (K) Notification
Siemens SC9000/SC9015 Medical Information Bus (MIB) Protocol Converter
7. Table of device similarities and differences to predicate device
| | Substantial Equivalent Device
Hewlett Packard Co
Model M1032A VueLink Module | Applicant
Siemens Medical Systems
SC 9000/9015 Series Medical Information Bus (MIB)
Protocol Converter |
| --- | --- | --- |
| Manufacturer | Hewlett Packard | Siemens Medical Systems -
Electromedical Group |
| 510K Number | K923682 | To be assigned |
| Intended Use | Provides an external device interface
capability to third party devices that
have a serial RS-232 and/or analog
output | Same |
| Intended Population | NA
Not connected to patients | Same |
| Intended Environment | OR or ICU | Same |
| Input port | RS-232, Analog, Analog/Digital
combination | Same |
| Output Port | HP specific | IEEE 1073 (MIB Connector) |
8. Assessment of non-clinical performance data for equivalence:
Substantial equivalence is claimed to the Hewlett Packard Model M1032A Vuelink Interface Plug-In Module cleared under 510(K) number K923682.
9. Assessment of clinical performance data for equivalence: Not applicable
10. Biocompatibility: Not applicable
11. Sterilization: Not applicable
12. Standards and Guidelines:
Currently there are no FDA standards for this device. However, the Siemens SC9000/SC9015 Series Medical Information Bus (MIB) Protocol Converter complies with:
- 1073.3.1-1994 IEEE Standard for Medical Device Communications
- Transport Profile- Connector Mode
- 1073.4.1-1994 IEEE Standard for Medical Device Communications
- Physical Layer Interface - Cable Connected
Siemens Medical Systems, Inc.
Electromedical Group
16 Electronics Avenue
Danvers, MA 01923
USA
Tel: (508) 750-7500
Fax: (508) 777-3398
Telex: 511958 (Siemensm SD)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches