MASIMO RAINBOW SET PRONTO -7 PULSE CO- OXIMETER AND ACCESSORIES, MODEL PRONTO-7
Applicant
Masimo Corporation
Product Code
DQA · Cardiovascular
Decision Date
Jun 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO2), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
Device Story
The Masimo Rainbow SET Pronto-7 is a noninvasive pulse CO-oximeter for spot-checking SpO2, pulse rate, and total hemoglobin (SpHb). It utilizes optical sensors to measure light absorption, processing signals to derive physiological parameters. The device features a touch-screen interface, pleth waveform display, and wireless connectivity (Bluetooth/WiFi). It is intended for use by trained personnel in diverse settings including hospitals, clinics, and home environments. Output is displayed on the device screen, allowing clinicians to perform rapid assessments of patient oxygenation and hemoglobin status, facilitating immediate clinical decision-making and patient monitoring.
Clinical Evidence
Clinical validation performed per ISO 9919:2005. SpO2 accuracy validated in human blood studies on healthy adult volunteers (light to dark skin) during induced hypoxia (70-100% range) vs. laboratory CO-oximeter and ECG. SpHb accuracy validated in healthy adults and patients (6-18 g/dl range) vs. laboratory CO-oximeter. Pulse rate accuracy validated in healthy adults and patients (40-110 bpm). Results: SpO2 accuracy (rms) < 2%, pulse rate accuracy (rms) < 3 bpm, SpHb accuracy (rms) < 1 g/dl. No device-related adverse events reported.
Technological Characteristics
Noninvasive optical pulse CO-oximeter. Features: touch-screen, pleth waveform display, rechargeable lithium polymer battery, USB, Bluetooth 2.0, WiFi b/g, Micro SD slot. Standards: EN 60601-1-2 (EMC), IEC 60601-1/UL 60601-1 (Electrical Safety), Type BF-Applied Part, IPX1. Operates via spot-check mode. Connectivity: Wireless and wired data output.
Indications for Use
Indicated for noninvasive spot checking of SpO2, pulse rate, and total hemoglobin (SpHb) in adult and pediatric patients. For use by trained personnel in clinical and non-clinical settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET® Pronto Pulse CO-Oximeter and Accessories (K091057)
Related Devices
K110028 — MASIMO RADICAL Y PULSE CO-OXIMETER · Masimo Corporation · Mar 17, 2011
K091057 — MASIMO RAINBOW SET PRONTO PULSE CO-OXIMETER AND ACCESSORIES · Masimo Corporation · Jul 9, 2009
K080238 — MASIMO RAINBOW SET RADICAL 7 PULSE CO-OXIMETER, MASIMO RAINBOW SET RAD 87 PULSE CO-OXIMETER · Masimo Corporation · May 12, 2008
K182887 — Masimo Rad-67 Pulse CO-Oximeter and Accessories · Masimo Corporation · Mar 22, 2019
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten text that appears to be a combination of letters and numbers. The text reads 'Klo 403' in a bold, slightly irregular font. The characters are closely spaced, and the overall impression is that of a quick, informal note.
## 510(k) SUMMARY
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>949-297-7000<br>FAX 949-297-7592 |
|------------------------------------|------------------------------------------------------------------------------------------|
| Company Contact: | Marguerite Thomlinson, Manager of Regulatory Affairs |
| Date Summary Prepared: | April 23, 2010 |
| Trade Name | Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and<br>Accessories |
| Common Name | Oximeter |
| Classification Name/ Product Code: | Oximeter Section 870.2700/ Product Code DQA |
| Predicate Device: | Masimo Rainbow SET® Pronto Pulse CO-Oximeter and<br>Accessories, 510(k) Number - K091057 |
#### Device Description
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories (Pronto-7) include the noninvasive technology of measurement for functional saturation of arterial oxygen hemoglobin (SpO2), pulse rate and total hemoglobin concentration (SpHb), which is based on the same technology of the predicate device, the Masimo Rainbow SET® Pronto Pulse CO-Oximeter and Accessories (Pronto), 510(k) no. K091057.
### Intended Use/ Indications for Use
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO3), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
### Technology Comparison
Features in the Pronto-7, which are not in the Pronto, include touch-screen user interface, pleth waveform display, rechargeable batteries or AC power, USB connection, flash memory, earphone jack, and wireless connection. The Pronto-7 has the following specifications:
| FEATURES | SPECIFICATIONS |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Display Range | Saturation (SpO2): 0-100%; Pulse Rate (bpm): 30-250 bpm<br>(Total Hemoglobin (SpHb): 0-25 g/dl; Perfusion Index (PI): 0.02-20% |
| Accuracy:<br>Adults and Pediatrics > 30kg | SpO2: 70-100±2%; Pulse Rate: 30-250±3 bpm; SpHb: 6-18 g/dl ±1 g/dl |
| Resolution | SpO2: 1%; Pulse Rate: 1 bpm; SpHb: 0.1 g/dl |
| AC Power | Voltage Input: 100-240 VAC, 50-60 Hz; Max Power Consumption: 15 VA |
| Batteries | Rechargeable lithium polymer |
{1}------------------------------------------------
# 510(k) SUMMARY
| FEATURES | SPECIFICATIONS |
|--------------------------------------|-------------------------------------------------------------------------|
| Temperature | Operating/ Storage: 41 to 104°F (5 to 40°C)/ -40 to 158°F (-40 to 70°C) |
| Relative Humidity | 5 to 95% noncondensing |
| Operating Altitude | 500 to 1,060 mbar pressure; -1,000 to 18,000 ft (-304 to 5,486m) |
| Alarm | System Failure; Low Battery Alarm |
| Display and Indicators | SpO2 (%); Pulse Rate (bpm); SpHb (g/dl); Perfusion Index (%) |
| Display and Indicators | Pleth Waveform; Sensor Status; Status Messages; Battery Status |
| Connection/ Output | Wireless: Bluetooth 2.0; WiFi b/g; Flash Memory: Micro SD card slot |
| Earphone jack | Connection for standard 3.5 mm earphone jack |
| EMC/ Electrical Safety<br>Compliance | EN 60601-1-2, Class B |
| EMC/ Electrical Safety<br>Compliance | IEC 60601-1, UL 60601-1, Internally Powered, AC Power Class 2 |
| EMC/ Electrical Safety<br>Compliance | Type BF-Applied Part, IPX1, Class 2 |
| Mode of Operation | Spot check |
### Clinical Data Summary
Clinical Studies: SpO2 accuracy has been validated in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation. Pronto-7 measurements on volunteers induced hypoxia in the range of 70-100% SpO2 are compared against a laboratory CO-oximeter and ECG monitor. SpHb accuracy has been validated with (arterial/ venous) blood from healthy adult male and female volunteers and on patients with light to dark skin pigmentation in the range of 6-18 g/dl SpHb, with Pronto-7 measurements compared against a laboratory CO-oximeter. Pulse rate accuracy has been validated on healthy adult male volunteers and on patients with light to dark skin pigmentation in the range of 40-110 bpm. The variation in these studies equals plus or minus one standard deviation encompasses 68% of the population.
#### Clinical Results: No device-related adverse events.
The clinical studies were performed in accordance with ISO 9919:2005. The studies resulted in SpO3 accuracy (rms) < 2%, pulse rate accuracy (rms) < 3bpm, and SpHb accuracy (rms) < 1g/d.
#### Non-Clinical Data Summary
The Pronto 7 complies with the voluntary standards as detailed in this submission. Laboratory testing for biocompatibility, safety and environmental was conducted to verify that the Pronto 7 met all design specifications and was substantially equivalent to the predicate device.
#### Conclusions
The information in this 510(k) submission demonstrates that the Masimo Rainbow SET Pronto 7 Pulse CO-Oximeter and Accessories are substantially equivalent to the predicate device, with respect to safety. effectiveness, and performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618
JUN 2 3 2010
Re: K100403
Trade/Device Name: Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 11, 2010 Received: June 14, 2010
Dear Ms. Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Ms. Thomlinson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Rioner
Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{4}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known):
Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories Device Name:
#### Indications For Use:
The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SpO₂), pulse rate, and total hemoglobin concentration (SpHb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospital-type facilities, home, clinics, physician offices, blood donation facilities, and ambulatory surgery centers).
| Prescription Use _X_ | AND/OR | Over-The-Counter Use _ |
|--------------------------------|--------|--------------------------------|
| (Per 21 CFR 801.109 Subpart D) | | (Per 21 CFR 801.109 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
rrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100403
510(k) Traditional, Masimo Pronto-7
Page 10 of 1254. 2/5/2010
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.